OneFerm pH VP 120 NTC

OneFerm pH VP 120 NTC

OneFerm pH VP 120 NTC

Product catalog summary
Product Overview
The OneFerm pH VP 120 NTC sensor is part of the OneFerm Arc Compatible pH Sensors family, designed to integrate with Arc Intelligent sensor technology for enhanced data collection in single-use applications.
Specifications
  • Length: 120 mm
  • Electrical Connector: VP6
  • Sensor Output: Traditional mV
  • Measurement Principle: Combination electrode; pH potential measured against reference
  • Measuring Range: pH 3 to 10
  • Sensitivity: 57 to 60 mV/pH at 25 °C before gamma irradiation; shift of ± 1.5 mV/pH after irradiation
  • Zero Point: ± 20 mV before gamma irradiation; shift of ± 10 mV after irradiation
  • Calibration Data: Provided on sensor label
  • pH Accuracy: ± 0.15 pH after gamma irradiation and dry storage; ± 0.1 pH after product calibration
  • Drift at Room Temperature: < 0.1 pH per week after wet-in time, in buffer pH 7
  • Membrane Shape: Spherical
  • Diaphragm: HP-Coatramic
  • Temperature Sensor: NTC 22 kOhm
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Glass, VMQ (Silicone elastomer)
  • O-ring Material: Compliant with FDA 21 CFR 177.2600; EG 1935/2004; USP <87>; USP <88> Class VI (121 °C)
  • Biocompatibility: Tested according to ISO 10993-5 and USP <87>
  • Wet-in Time: 30 min after dry storage
  • Shelf Life: 12 months prior to gamma irradiation; 18 months after gamma irradiation
  • Autoclavable: No
  • CIP: No
  • Steam Sterilizable: No
  • Gamma Irradiatable: Yes, up to 45 kGy
  • Operating Temperature Range: 4 to 50 °C
  • Sample Minimum Conductivity: 100 µS/cm
  • Pressure Range: 0 to 2 bar
Key Considerations
  • The sensor is not autoclavable, CIP, or steam sterilizable, but it is gamma irradiatable up to 45 kGy.
  • It has a shelf life of 12 months before and 18 months after gamma irradiation.
  • Biocompatibility is ensured as per ISO and USP standards.
See more

Catalog excerpts

OneFerm pH VP 120 NTC-1

OneFerm pH VP 120 NTC Specification Sheet (Part/REF # 243236) OneFerm Arc Compatible pH Sensors fully integrate with Arc Intelligent sensor technology for data-rich sensor information in single use applications. Product Specifications Sensor Family Electrical Connector Sensor Output Measurement Principle Combination electrode; pH potential measured against reference Measuring Range Before gamma irradiation: 57 to 60 mV/pH at 25 °C (97 % -101 %); Shift after gamma irradiation: ± 1.5 mV/pH Zero Point Before gamma irradiation: ± 20 mV; Shift after gamma irradiation: ± 10 mV Calibration Data Yes, provided on sensor label After gamma irradiation and dry storage: ± 0.15 pH with provided calibration values. ± 0.1 pH after product calibration at the measurement temperature and within 2 pH units from the product calibration point Drift at Room Temperature < 0.1 pH per week after wet-in time, in buffer pH 7 Membrane Shape Temperature Sensor Process Connection Wetted Parts Glass VMQ (Silicone elastomer) O-ring Material Wet-in Time Shelf Life 12 months prior gamma irradiation. 18 months after gamma irradiation Steam Sterilizable Gamma Irradiatable Operating Temperature Range 4 to 50 °C Sample min Conductivity Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/243236

 Open the catalog to page 1

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold G 120

    2  Pages

  2. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.