MecoTrode VP 120

MecoTrode VP 120

MecoTrode VP 120

Product catalog summary
Product Overview
MecoTrode VP 120 is a traditional pH sensor designed for industrial applications, including those involving hydrofluoric acid (HF). It provides a traditional millivolt (mV) output and features a combination electrode design.

Specifications
  • Sensor Family: MecoTrode
  • Parameter: pH
  • Output: Traditional mV
  • Length: 120 mm
  • Connector: VP6
  • Measurement Principle: Combination electrode; pH potential measured against reference
  • Measuring Range: pH 0 to 14
  • Sensitivity: 57 to 59 mV/pH at 25 °C
  • Zero Point: 0 ± 20 mV
  • pH Glass: Hamilton Type H Glass
  • Glass Resistance: Moderate
  • Membrane Shape: Spherical
  • Diaphragm: HP-Ceramic, 3 diaphragms
  • Electrolyte: Viscous 3 M KCl-Pharma, blue
  • Reference System: Hamilton Everef
  • Temperature Sensor: Pt100
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Glass, Ceramic, EPDM
  • O-ring Material: FDA 21 CFR 177.2600; EG 1935/2004; USP <87>; USP <88> Class VI (121 °C)
  • Serial Number: Yes
  • Recommended Storage: Hamilton Storage Solution or 3 M KCl
  • Certificate: Yes, with measured values in Buffer pH 4.01 and pH 7.00
  • ATEX Approval: II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
  • IECEx Approval: II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
  • Autoclavable: No
  • CIP: No
  • Steam Sterilizable: No
  • Operating Temperature Range: 0 to 130 °C
  • Sample Min Conductivity: 50 µS/cm
  • Pressure Range: 0 to 16 bar at 25 °C, 0 to 6 bar at 130 °C

Additional Information
Specifications are subject to change without notice. For more details, visit the Hamilton Company website.
See more

Catalog excerpts

MecoTrode VP 120-1

MecoTrode VP 120 Specification Sheet (Part/REF # 238437) MecoTrode traditional pH sensors offer a traditional mV output combined with glass and reference design for industrial applications including HF. Product Specifications Sensor Family Sensor Output Electrical Connector Measurement Principle Combination electrode; pH potential measured against reference Measuring Range Zero Point Hamilton Type H Glass Membrane Shape Viscous 3 M KCl-Pharma, blue Reference System Hamilton Everef Liquid Earth Temperature Sensor Process Connection Wetted Parts Glass Ceramic EPDM (Ethylene propylene elastomer) O-ring Material Serial Number Recommended Storage Hamilton Storage Solution or 3 M KCl Yes, with measured values in Buffer pH 4.01 and pH 7.00 ATEX Approval II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db IECEx Approval II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db

 Open the catalog to page 1
MecoTrode VP 120-2

Steam Sterilizable Operating Temperature Range Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/238437 Spec. Version

 Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold G 120

    2  Pages

  2. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.