1. Catalogs
  2. Hamilton Process Analytics
  3. Incyte Arc Expert - 120

Incyte Arc Expert - 120

Incyte Arc Expert - 120

Incyte Arc Expert - 120

Product catalog summary
Product Overview: The Incyte Arc Expert sensors are designed for real-time, in-situ measurement of viable cell density, allowing for immediate detection and response to events without increased off-line sampling.
Specifications:
  • Sensor Family: Incyte Arc Expert
  • Length: 120 mm
  • Parameter Measured: Viable Cell Density
  • Electrical Connector: VP8
  • Sensor Output: Arc: Modbus
  • Measurement Principle: Permittivity
  • Measuring Range: 0 to 700 pF/cm, equivalent to 5 x 105 to 8 x 109 cells/mL (mammalian)
  • Accuracy at 25°C: Conductivity (at 0 pF): ± 25 µS or ± 1%, whichever is greater
  • Precision at 25°C: Permittivity (at 1 MHz): ± 1 pF or ± 1%, whichever is greater
  • Temperature Sensor: Yes
  • Configurable Values: VCD, Conductivity, Temperature
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Platinum, Stainless Steel 1.4435, PEEK, EPDM
  • O-ring Material: EPDM
  • Surface Quality of Steel: Ra < 0.4 µm (N5)
  • Digital Interface: RS485 Modbus (max. 31 addresses)
  • Baud Rate: 19200, 38400, 57600, 115200 Bd
  • Operating Voltage: 24 V (± 10%)
  • Serial Number: Yes
  • Certificate: Yes
  • Autoclavable: Yes, max. Temperature 130°C (Revision 00), 140°C (Revision 01 and higher)
  • CIP: Yes
  • Steam Sterilizable: Yes, max. Temperature 140°C
  • Operating Temperature Range: 0 to 60°C
  • Conductivity Range: 1 to 80 mS/cm (firmware below CDCUM003), 0.5 to 80 mS/cm (firmware CDCUM003 and above with Hamilton calibration)
  • Pressure Range: 0 to 12 bar
  • Resolution at 25°C: 0.1 pF/cm
  • Max. Accuracy: Conductivity (at 0 pF): ± 100 µS (for whole operating temperature range)
  • Max. Precision: Permittivity (over all frequencies): ± 1 pF (for whole operating temperature range)
Additional Information: Specifications are subject to change without notice. For more details, visit the Hamilton Company website.
See more

Catalog excerpts

Incyte Arc Expert - 120-1

Incyte Arc Expert - 120 Specification Sheet (Part/REF # 243950-0211) Incyte Arc sensors enable real-time, in-situ measurement of viable cells, making it possible to detect events and respond in real time without increased off-line sampling. Product Specifications Sensor Family Viable Cell Density Electrical Connector Sensor Output Measurement Principle Measuring Range 0 to 700 pF/cm, equivalent to 5 x 10^5 to 8 x 10^9 cells/mL (mammalian) Conductivity (at 0 pF): ± 25 µS or ± 1 % Whichever value is greater over the entire measuring range Permittivity (at 1 MHz): ± 1 pF or ± 1 % Whichever value is greater over the entire measuring range Temperature Sensor Configurable Values Process Connection Wetted Parts Platinum Stainless Steel 1.4435 PEEK (Polyetheretherketone) EPDM (Ethylene propylene elastomer) O-ring Material EPDM (Ethylene propylene elastomer) Surface Quality of Steel Digital Interface Baud Rate Operating Voltage Serial Number Yes, max. Temperature 130 °C (Incyte Arc Revision 00) Yes, max. Temperature 140 °C (Incyte Arc Revision 01 and higher) Steam Sterilizable Operating Temperature Range Conductivity Range 1 to 80 mS/cm for firmware below CDCUM003 0.5 to 80 mS/cm for firmware CDCUM003 and above (with Hamilton ca

 Open the catalog to page 1
Incyte Arc Expert - 120-2

Conductivity (at 0 pF): ± 100 µS (for whole operating temperature range) Permittivity (over all frequencies): ± 1 pF (for whole operating temperature range) Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/243950-0211 Spec. Version

 Open the catalog to page 2