EasyFerm Bio PHI Arc 120
2Pages

{{requestButtons}}

Catalog excerpts

EasyFerm Bio PHI Arc 120 - 1

EasyFerm Bio PHI Arc 120 Specification Sheet (Part/REF # 243632-1313) Arc intelligent sensor technology securely stores your sensor data and outputs your measurement information directly to the control system. Product Specifications Sensor Family Electrical Connector Sensor Output Measurement Principle Combination electrode; pH potential measured against reference Measuring Range Zero Point Hamilton Type PHI Glass Membrane Shape Pressurized Foodlyte Reference System Hamilton Everef-F, Silver ion barrier included Temperature Sensor Configurable Values Process Connection Wetted Parts Glass VMQ (Silicone elastomer) O-ring Material Galvanically not isolated; pulse width modulation 3.5 kHz Baud Rate Operating Voltage Serial Number Recommended Storage Hamilton Storage Solution or 3 M K

Open the catalog to page 1
EasyFerm Bio PHI Arc 120 - 2

Certificate Yes, with measured values in Buffer pH 4.01 and pH 7.00 Steam Sterilizable Yes, max. Temperature 140 °C Operating Temperature Range Analog interface: 0 to 110 °C; Digital interface: 0 to 140 °C Specifications are subject to change without notice Spec. Version C

Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures