Conducell 4UHF Arc 120

Conducell 4UHF Arc 120

Conducell 4UHF Arc 120

Product catalog summary
Product Overview: The Conducell 4UxF Arc conductivity sensors are designed for storing calibrations within the sensor and direct connection to control systems. They are part of the Conducell 4UxF sensor family and measure conductivity using a 4-pole contacting principle.
Specifications:
  • Parameter: Conductivity (4-Pole)
  • Sensor Output: Arc: Modbus, 4 to 20 mA
  • Length: 120 mm
  • Electrical Connector: VP8
  • Measuring Range: 1 µS/cm to 300 mS/cm
  • Accuracy at 25 °C: ± 3% at 1 µS/cm to 100 mS/cm, ± 5% at 100 to 300 mS/cm
  • Cell Constant: 0.36/cm
  • Temperature Sensor: NTC 22 kOhm
  • Configurable Values: Conductivity: µS/cm, mS/cm; Temperature: °C, °F, K
  • Diameter: 12 mm
  • Process Connection: PG13,5
  • Wetted Parts: Stainless Steel 2.4602, PEEK, EPDM
  • Surface Quality of Steel: Ra < 0.4 µm (N5, electropolished)
Interfaces:
  • Analog Interface 1: 4 to 20 mA for conductivity, programmable
  • Analog Interface 2: 4 to 20 mA for Temperature, programmable
  • Galvanic Isolation: Not isolated; pulse width modulation 3.5 kHz
  • Digital Interface: RS485 Modbus (max. 31 addresses)
  • Baud Rate: 4800, 9600, 19200, 38400, 57600, 115200 Bd
Operating Conditions:
  • Operating Voltage: 7 to 30 VDC max. 150 mW
  • Operating Temperature Range: Analog interface: 0 to 110 °C; Digital interface: 0 to 140 °C
  • Pressure Range: 0 to 20 bar (130 °C)
Certifications and Approvals:
  • Serial Number: Yes
  • Certificate: Yes, with calibrated cell constant
  • ATEX Approval: No
  • IECEx Approval: No
  • Autoclavable: Yes, max. Temperature 130 °C
  • CIP: Yes
  • Steam Sterilizable: Yes, max. Temperature 140 °C
Additional Information: Specifications are subject to change without notice. For more details, refer to the Material Specification document.
See more

Catalog excerpts

Conducell 4UHF Arc 120-1

Conducell 4UHF Arc 120 Specification Sheet (Part/REF # 243590-3111) Conducell 4UxF Arc conductivity sensors allow calibrations to be stored in the sensor and direct connection to the control system. Product Specifications Sensor Family Sensor Output Electrical Connector Measurement Principle Measuring Range ± 3 % at 1 µS/cm to 100 mS/cm, ± 5 % at 100 to 300 mS/cm Cell Constant Temperature Sensor Configurable Values Conductivity: µS/cm, mS/cm; Temperature: °C, °F, K Process Connection Wetted Parts Stainless Steel 2.4602 PEEK (Polyetheretherketone) EPDM (Ethylene propylene elastomer) See compliance details in Material Specification document Surface Quality of Steel Galvanically not isolated; pulse width modulation 3.5 kHz Digital Interface Baud Rate Operating Voltage Serial Number Yes, with calibrated cell constant ATEX Approval IECEx Approval Steam Sterilizable

 Open the catalog to page 1
Conducell 4UHF Arc 120-2

Operating Temperature Range Analog interface: 0 to 110 °C; Digital interface: 0 to 140 °C Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/243590-3111

 Open the catalog to page 2