Chemotrode P PHI 120

Chemotrode P PHI 120

Chemotrode P PHI 120

Product catalog summary
Product Overview: The ChemoTrode is designed for use in aggressive industrial environments, specifically targeting pH measurement.
Specifications:
  • Sensor Family: ChemoTrode
  • Parameter: pH
  • Sensor Output: Traditional mV
  • Length: 120 mm
  • Electrical Connector: S7
  • Measurement Principle: Combination electrode with pH potential measured against reference
  • Measuring Range: pH 0 to 14
  • Sensitivity: 57 to 59 mV/pH at 25°C
  • Zero Point: 0 ± 20 mV
  • pH Glass: Hamilton Type PHI Glass
  • pH Glass Resistance: Moderate
  • Membrane Shape: Spherical
  • Diaphragm: HP-Ceramic with 3 diaphragms
  • Electrolyte: Protelyte
  • Reference System: Hamilton Everef-F with Silver ion barrier
  • Liquid Earth: No
  • Temperature Sensor: No
  • Diameter: 30 mm
  • Wetted Parts: Glass, Ceramic
  • Serial Number: Yes
  • Recommended Storage: Hamilton Storage Solution or 3 M KCl
  • Certificate: Yes, with measured values in Buffer pH 4.01 and pH 7.00
  • ATEX Approval: II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
  • IECEx Approval: II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db
  • Autoclavable: No
  • CIP: Yes
  • Steam Sterilizable: Yes
  • Operating Temperature Range: 0 to 130°C
  • Sample Minimum Conductivity: 50 µS/cm
  • Pressure Range: 0 to 6 bar
  • Shaft Diameter Below: 12 mm
Additional Information: Specifications are subject to change without notice. For more details, visit the Hamilton website.
See more

Catalog excerpts

Chemotrode P PHI 120-1

Chemotrode P PHI 120 Specification Sheet (Part/REF # 238761) The ChemoTrode targets aggressive media found in industrial applications. Product Specifications Sensor Family Sensor Output Electrical Connector Measurement Principle Combination electrode; pH potential measured against reference Measuring Range Zero Point Hamilton Type PHI Glass Membrane Shape Reference System Hamilton Everef-F, Silver ion barrier included Liquid Earth Temperature Sensor Wetted Parts Glass Ceramic Serial Number Recommended Storage Hamilton Storage Solution or 3 M KCl Yes, with measured values in Buffer pH 4.01 and pH 7.00 ATEX Approval II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db IECEx Approval II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db Steam Sterilizable Operating Temperature Range

 Open the catalog to page 1
Chemotrode P PHI 120-2

Shaft Diameter Below Specifications are subject to change without notice https://www.hamiltoncompany.com/sensors/238761

 Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold G 120

    2  Pages

  2. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.