ChemoTrode Bridge 120

ChemoTrode Bridge 120

ChemoTrode Bridge 120

Product catalog summary
Product Overview: The ChemoTrode is designed for use in aggressive media typically found in industrial applications. It is part of the ChemoTrode sensor family and is used to measure pH levels.
Specifications:
  • Parameter: pH
  • Sensor Output: Traditional mV
  • Length: 120 mm
  • Electrical Connector: S7
  • Measurement Principle: Combination electrode; pH potential measured against reference
  • Measuring Range: pH 0 to 14
  • Sensitivity: 57 to 59 mV/pH at 25 °C
  • Zero Point: 0 ± 20 mV
  • pH Glass: Hamilton Type PHI Glass
  • pH Glass Resistance: Moderate
  • Membrane Shape: Spherical
  • Diaphragm: Platinum fibers
  • Number of Diaphragms: 1
  • Electrolyte: Skylyte
  • Reference System: Hamilton Everef-B
  • Liquid Earth: No
  • Temperature Sensor: No
  • Diameter: 30 mm
  • Wetted Parts: Platinum, Glass
  • Serial Number: Yes
  • Recommended Storage: Hamilton Storage Solution or 3 M KCl
  • Certificate: No
  • ATEX Approval: No
  • IECEx Approval: No
  • Autoclavable: No
  • CIP: Yes
  • Steam Sterilizable: Yes
  • Operating Temperature Range: 0 to 130 °C
  • Sample Minimum Conductivity: 50 µS/cm
  • Pressure Range: 0 to 6 bar
  • Shaft Diameter Below: 12 mm
Additional Information: Specifications are subject to change without notice. For more details, visit the Hamilton Company website.
See more

Catalog excerpts

ChemoTrode Bridge 120-1

ChemoTrode Bridge 120 Specification Sheet (Part/REF # 238770) The ChemoTrode targets aggressive media found in industrial applications. Product Specifications Sensor Family Sensor Output Electrical Connector Measurement Principle Combination electrode; pH potential measured against reference Measuring Range Zero Point Hamilton Type PHI Glass Membrane Shape Platinum fibers Reference System Hamilton Everef-B Liquid Earth Temperature Sensor Wetted Parts Platinum Glass Serial Number Recommended Storage Hamilton Storage Solution or 3 M KCl ATEX Approval IECEx Approval Steam Sterilizable Operating Temperature Range

 Open the catalog to page 1
ChemoTrode Bridge 120-2

Shaft Diameter Below Specifications are subject to change without notice Spec. Version B

 Open the catalog to page 2

All Hamilton Process Analytics catalogs and technical brochures

  1. OxyGold G 120

    2  Pages

  2. OxyGold B 120

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.