1. Catalogs
  2. halstrup-walcher GmbH
  3. Data sheet high precision differential pressure transmitter P 26

Data sheet high precision differential pressure transmitter P 26

Data sheet high precision differential pressure transmitter P 26

Data sheet high precision differential pressure transmitter P 26

Product catalog summary
Measurement Ranges: The document specifies various measurement ranges for differential pressure, including 10/50/100/250/500 Pa and 1/2.5/5/10/20/50/100 kPa, with options for customization. Measurement accuracy is ±0.2% FS for ranges ≤50 kPa or ±0.5% FS.
Temperature Coefficients: The span temperature coefficient is 0.03% FS/K within 10-50°C, and the zero-point temperature coefficient is ±0% due to cyclical zero-point correction.
System Pressure and Overload Capacity: Maximum system pressure is 600 kPa for ranges ≥2.5 kPa, with an overload capacity of 200x for ranges <2.5 kPa.
Medium and Response Time: Suitable for air and non-aggressive gases, with an adjustable step response time of 25 ms to 40 s.
Temperature and Power Specifications: Rated temperature range is 10-50°C, storage temperature is -10 to 70°C, and power consumption is approximately 6 VA.
Physical and Protection Specifications: The device weighs approximately 750 g, has IP65 protection (IP40 with USB), and features 3 x M16 cable glands and NW 6 mm pressure ports.
Output and Power Supply: Configurable output signals include 0-10 V, 0-20 mA, and 4-20 mA. Power supply options include 24 V AC/DC and 230/115 V AC.
Features: The P26 is a high-precision differential pressure transmitter with options for top-hat rail or wall mounting, scalable measurement ranges, zero-point correction, and multilingual menu. Optional features include contact points, USB interface, and air meter function.
Applications: The P26 is suitable for air-conditioning, cleanroom, and process applications, with a wide range of units for pressure and volume flow measurement.
Accessories and Software: Accessories include silicone and Norprene tubing, and a Y-piece for tubing. User software is available for parameter setting and monitoring via USB or RS232 interface, compatible with P26, P34, and P29 models.
See more

Catalog excerpts

Data sheet high precision differential pressure transmitter P 26-1

P 26 Measurement ranges (also ± measurement ranges) 10 / 50 / 100 / 250 / 500 Pa 1 / 2.5 / 5 / 10 / 20 / 50 / 100 kPa freely scalable from 10 .. 100 % within a measurement range others available upon request Measurement accuracy 1) Temperature coefficient span Temperature coefficient zero point ± 0 % (cyclical zero-point correction) Max. system pressure / Overload capacity 600 kPa for measurement ranges ≥ 2.5 kPa 200 x for measurement ranges < 2.5 kPa Step response time (T63) (Time constant) Rated temperature range Storage temperature Power consumption Cable glands Pressure ports Protection class x igh precision differential pressure transmitter H for air-conditioning, cleanroom and process Picture: Version with display others available on request Measurement accuracy for the reference 0.3 Pa, for measuring ranges ≤ ±1.5 kPa Contact points x Wide range of units for pressure and volume flow x Also ± measurement ranges x Top-hat rail or wall mounting x Scalable measurement ranges and units (changeover contacts) x Zero-point correction prevents zero-point drift o utput signals can be configured freely Data interface x uilt-in valve provides a high level of overpressure B protection Power supply x Multilingual menu (English / French / German / Italian) (data cable supplied) ± 10 % (with galvanic separation) x xtensive configuration options using free E parametrization software via internal RS232 interfaces Optional x Contact points with adjustable switching outputs (data cable supplied) x Set the zero-point via an external interface Supply voltage of 24 V DC required x Display and operating keys Pressure connections laboratory tube tail Calibration certificate Calibration according to DKD-R 6-1 Factory calibration Can be pre-set on request: Time constant, relay parameter, analogue output root-extracted / linear, deactivation of the cyclic zeroing Measurement accuracy

 Open the catalog to page 1
Data sheet high precision differential pressure transmitter P 26-2

MEASUREMENT OF DIFFERENTIAL PRESSURE Measurement of differential pressure is useful in a broad range of applications. It is used in ventilation and air-conditioning technology but also in many areas of air handling process technology. halstrup-walcher offers a wide range of products for stationary measurement of differential pressure: Product High precision, freely scalable pressure transmitter for critical applications Measuring transmitter with very small dimensions – ideal for the control cabinet High precision, freely scalable pressure transmitter for natural gas Housing installation For...

 Open the catalog to page 2

All Halstrup-walcher GmbH catalogs and technical brochures

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.