1. Catalogs
  2. halstrup-walcher GmbH
  3. Data sheet digital manometer EMA 84

Data sheet digital manometer EMA 84

Data sheet digital manometer EMA 84

Product catalog summary
Specifications:
Measurement ranges are available from 0 to 100 Pa up to 0 to 100 kPa. Measurement accuracy varies by range, with ±0.2% FS for 1 to 10 kPa, ±0.5% FS for 1 to 100 kPa, and ±1% FS for other ranges. Overload capacity is 10 times for ranges ≤ 10 kPa and 2 times for ranges > 10 kPa. The device supports zero-point correction via a potentiometer.

Analog Output:
Optional analog output is available at 0 to 1 V with a BNC connector, requiring a load resistance of ≥ 2 kΩ.

Display and Controls:
The device features a 3 ½ digit LCD with a character height of 13 mm. Time constants can be toggled between 0.02 s, 0.2 s, and 1 s. It includes switches for mbar/Pa, on/off, zero-point correction, and time constant.

Environmental Conditions:
Operating temperature ranges from 10 to 60 °C, while storage temperature ranges from -10 to 70 °C. The preferred operating position is horizontal.

Power Supply and Physical Characteristics:
The device is powered by a 9 V battery and weighs approximately 0.8 kg. Pressure ports are designed for tubing NW 6 mm.

Certifications:
The device is certified with CE and UKCA.

Features:
The EMA 84 is a robust digital pressure gauge ideal for service technicians, featuring an easy-to-read display, high accuracy, and manual zero-point correction. It offers an optional analog output for data logging.

Calibration Options:
Calibration options include none, factory calibration, or calibration according to DKD-R 6-1 standards.
See more

Catalog excerpts

Data sheet digital manometer EMA 84-1

or ± 1 % FS Overload capacity Zero-point correction Operating temperature Storage temperature Operating position preferably horizontal Power supply Pressure ports x High level of accuracy 3 ½ digit LCD, character height = 13 mm Time constants x Ideal for service technicians, easy-to-read display x Very robust digital pressure gauge Analog output via potentiometer on front face 10 x for measurement ranges ≤ 10 kPa 2 x for measurement ranges > 10 kPa M easurement accuracy of the reference 0.3 Pa, precision of the reference 0.12 Pa, for measuring ranges ≤ ±1 kPa Measurement range x Manual zero-point correction x ith optional analog output for writer or W power / voltage logger analog output Measurement accuracy Analog output Calibration certificate Factory calibration Calibration according to DKD-R 6-1 pressure port for pos. pressure pressure port for neg. pressure time constant switch on / off switch Zero-point mbar / Pa correction switch Data sheet EMA 84 – Date: 12 / 

 Open the catalog to page 1

All Halstrup-walcher GmbH catalogs and technical brochures

Archived catalogs