Overview
The ORBISPHERE 6110 is a total package analyser designed for final product quality control, specifically measuring oxygen and carbon dioxide in both headspace and liquid phases of bottles or cans. It is tailored for at-line environments, offering a user-friendly interface and automatic operation.

Key Features
  • Productivity & Performance: At-line operation saves time and allows rapid response. It can switch between bottles and cans without additional accessories and offers fast measurement times.
  • Ease of Use: No sample preparation is needed, featuring a piercing location spot and one-button operation.
  • Safety: Includes a protection shield, automatic rim detection, and an emergency stop button.
  • Comfort: Features an optimal touch screen position and accessible sample position from both sides, with a Menu Wizard for setup and service operations.
  • Design: Single enclosure with easy data transfer options.

Data Confidence
The analyser ensures data confidence through self-calibration and validation functions, traceable data with an independent audit trail, and multilevel access with password protection.

Cost Efficiency
The ORBISPHERE 6110 offers a low cost of ownership with immediate process response, low analysis costs, minimal maintenance, and maximum uptime due to self-checks.

Measurement Process
The analysis is fully automated in a five-step process: package piercing, headspace O2 measurement, liquid O2 and CO2 measurement, headspace volume determination, and system standby mode.

Technical Specifications
  • Operating Range: O2: 1 ppb – 20 ppm; CO2: 0.75–5 v/v or 1.5–10 g/kg.
  • Repeatability: TPO ±5 µg/l or ±10%; CO2 ±0.05 v/v or 0.10 g/kg or ±2%.
  • Analysis Time: Approximately 4 minutes with a throughput of 12 samples per hour.
  • Display Units: Various units for O2, CO2, total package quantities, pressure, and temperature.
  • Operating Limits: Package temperature: -2 to 30 °C; pressure: 1.4 to 6.8 bar; ambient temperature: 0 to 40 °C.
  • Package Settings: Supports metallic and PET closures, with specific height and volume limits.
  • Enclosure: Dimensions: 537 × 540 × 942 mm; Weight: 55 kg; Protection: IP 20.
  • Power Requirements: 100/240 VAC, 50/60 Hz, max 250 VA.
  • Purge Gas: CO2 with >99.9% purity at 4–7 bar.
  • Compliance: Meets European directives and safety standards.

Additional Services
HACH LANGE offers seminars, workshops, on-site technical support, and quality assurance services.
See more

Catalog excerpts

ORBISPHERE 6110 total package analyser-1

Final product quality control ORBISPHERE 6110 total package analyser INFORMATION FINAL PACKAGE ANALYSIS TOTAL PACKAGE ANALYSER ORBISPHERE 6110

 Open the catalog to page 1
ORBISPHERE 6110 total package analyser-2

Total package analyser for Oxygen & Carbon Dioxide measurements The ORBISPHERE 6110 total package analyser is the latest technology in final package analysis for gas measurements. It has been designed to meet the practical needs of the at-line environment and to deliver a simple, high-quality user interface. This automatic analyser provides oxygen and carbon dioxide values for both headspace and liquid phase in bottles or cans. Automatic operation, self-validation and at-line ergonomics result in an exceptional tool for final product quality control. At-line ergonomics – simple, effective, time...

 Open the catalog to page 2
ORBISPHERE 6110 total package analyser-3

Data confi dence Data confi dence is driven in two ways: - The OBISPHERE 6110 incorporates self-calibration and self-validation functions to ensure continuous reliability of measurement. - Stored data or data retrieved through LIMS/OPC are traceable with an independent audit trail. - Multilevel operations access is provided with password protection. Low cost of ownership The ORBISPHERE 6110 delivers ROI from day one: - At-line operation means immediate response to process changes or out of spec results. - Low cost per analysis ensures operating costs are low. - Minimum maintenance, due to gas...

 Open the catalog to page 3
ORBISPHERE 6110 total package analyser-4

Top 3 Germany´s most sustainable Products and Services 2009 HACH LANGE GMBH Willstätterstraße 11 D-40549 Düsseldorf Tel. +49 (0)2 11 52 88-0 Fax +49 (0)2 11 52 88-143 [email protected] www.hach-lange.com Technical data TOTAL PACKAGE ANALYSER_ORBISPHERE 6110 DOC062.52.20051.Jul09 Analysis Operating range O2 1 ppb – 20 ppm CO2 0.75–5 v/v or 1.5–10 g/kg Repeatability, r95 (at 20 ±5 °C) TPO* ±5 ìg/l ±10 % whichever the greater CO2 ±0.05 v/v or 0.10 g/kg ±2 % whichever the greater Typical analysis time About 4 minutes Typical sample throughput 12 samples / hour Display units O2 concentration: ppb,...

 Open the catalog to page 4
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.