video corpo

GB-BF20010

GB-BF20010

Product catalog summary
Application
The GB-BF20010 Mask Synthetic Blood Penetration Tester is designed to evaluate the resistance of masks to synthetic blood penetration under various test pressures.
Specifications
  • Standard Compliance: GB 19083-2010, YY0469-2011
  • Features: The instrument utilizes a gas source providing (20 ± 1) kPa pressure for continuous pressurization, independent of test site space limitations. It includes a pressure gauge for adjustable pressure display and uses compressed air as the pressurized medium. A special stainless steel test slot ensures firm sample clamping and prevents synthetic blood splashing. It features a square metal barrier net with open space ≥50% and bending ≤5mm at 20kPa, a digital display timer with ±1 second accuracy, and a clamp generating 13.5Nm torque.
Principle
The mask material is tested with synthetic blood under continuous pressure, and penetration is visually assessed.
Technical Parameters
  • Pressure Range: 0~50KPa
  • Nozzle Diameter: 0.84mm
  • Jet Pressure: 10.6 KPa (80 mmHg), 16 KPa (120 mmHg), 21.3 KPa (160 mmHg) with ±0.5%FS accuracy
  • Jet Velocity: 450cm/s, 550cm/s, 635cm/s
  • Spray Distance: 300mm±10mm Adjustable
  • Dimensions: 825mm×440mm×335mm
  • Power Supply: AC 220V, 50Hz
Contact Information
Address: NO.1 Minghua 3 Street, Jinxiu Road, Economic Development Zone, Guangzhou, China
Tel: 86-20-86153790, 86-20-86153742
Fax: 86-20-82087405
Website: https://www.gbpitester.com
Email: [email protected]
See more

Catalog excerpts

GB-BF20010-1

Application GB-BF20010 Mask Synthetic Blood Penetration Tester is suitable for the resistance of masks to the penetration of synthetic blood under different levels of test pressure. Principle The mask material was tested with synthetic blood under continuous applied pressure, and the penetration of synthetic blood on the material was visually checked. Specifications The instrument uses a gas source that can provide (20 ± 1) kPa pressure to continuously pressurize the sample without being limited by the space of the test site. The instrument has a pressure gauge to display the pressure, the pressure can be adjusted. Use pressurized medium: compressed air. The special stainless steel penetrating test slot ensures that the sample is firmly clamped, and the synthetic blood is prevented from splashing around. Square metal barrier net: open space >50%; bending <5mm at 20kPa. Digital display timer, accuracy ± 1 second. The instrument has a clamp that can generate 13.5Nm torque. Item Pressure Range measurement Accuracy Spray Distance Nozzle Diameter Jet Pressure JetVelocity Dimensions PowerSupply https://www.gbpitester.com E-mail: [email protected]

 Open the catalog to page 1

All Guangzhou Biaoji Packaging Equipment Co., Ltd. catalogs and technical brochures

  1. GB-QL-720A

    7  Pages

  2. 9802-K

    1  Page

  3. GBN702

    1  Page

  4. GBN701

    1  Page

  5. GB-XF1000

    1  Page

  6. GB-PF400H

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.