1. Catalogs
  2. Greenlight Innovation Corp.
  3. Cell Voltage Monitoring System

Cell Voltage Monitoring System
1 /2Pages

Cell Voltage Monitoring System

Cell Voltage Monitoring System
1 /2Pages

Catalog excerpts

Cell Voltage Monitoring System-1

CELL VOLTAGE MONITORING DC VOLTAGE MONITORING Battery and Fuel Cell Testing » Each module contains 32 channels: 8 modules per 4U rack (as shown) » Each slot supports input types: ±100mV, ±5V, ±10V, ±75V, ±150V, ±300V » 16-bit A/D converter on each channel » Modules can be configured in daisy-chained mode (up to 32 channels), differential mode (up to 16 channels), or mixed configurations » Maximum channel-to-channel common mode voltage isolation: 1,000 VDC » Channel input impedance (voltage channel): ≥1MΩ » Channel accuracy: ≤ 0.05% full scale » Custom form factors available: 4U 19” box is shown » Standard and Pro software packages available

 Open the catalog to page 1
Cell Voltage Monitoring System-2

Each module can be connected in a star Ethernet network configuration for 1,000+ channels Simultaneous data sampling on each board; not multiplexed 1kHz burst sampling mode possible via external trigger input Up to 100Hz continuous data sampling Communication interface: 10/100 Mbps Ethernet (optional CAN Bus) Thermal drift compensation on each channel maintains accuracy over a wide temperature range Software Specifications SOFTWARE FEATURE Configurable data logging/real-time charting Configurable on-screen alarm notification Configurable signal scaling Historical graph/xy graph/trigger-based...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.