Densitometers

Densitometers

Densitometers

Product catalog summary
Overview: The document provides detailed specifications and features of two densitometer models, DEN-1 and DEN-1B, designed for measuring the turbidity of cell suspensions in life-science applications. These devices are compact, efficient, and factory calibrated for specific measurement ranges.
Specifications:
  • Measurement Range: DEN-1 measures from 0.5 to 4.0 McFarland units, while DEN-1B measures from 0.00 to >7.5 McFarland units. Both can extend up to 15.00 McFarland units with increased standard deviation.
  • Precision: ±3% with a standard deviation of ±0.1 McF at 3.00 McF units.
  • Measurement Time: 1 second.
  • Light Source: LED with a wavelength of 565 ± 15 nm.
  • Tube Compatibility: Designed for 18 mm tubes, with an optional adaptor for 16 mm tubes.
  • Power Supply: DEN-1B can be powered by 3 x AA batteries or an external 12V power supply.
  • Display: DEN-1 uses an LED display with 0.1 McF resolution, while DEN-1B uses an LCD display with 0.01 McF resolution.
  • Dimensions and Weight: Both models have dimensions of 75 mm (h) x 115 mm (d) x 165 mm (w) and weigh 0.7 kg.
Features:
  • Both models are factory calibrated and can retain calibration without a power supply. User calibration is possible with commercial standards or laboratory-prepared cell suspensions.
  • Compact design suitable for small workspaces.
  • Applications include determining cell concentration in fermentation, testing microorganism susceptibility to antibiotics, and measuring optical density.
Accessories:
  • DEN MCF STDS: Set of liquid McFarland standards in 16 mm glass tubes.
  • DEN MCF 18 STDS: Set of gel McFarland standards in 18 mm glass tubes.
  • D16 Spare Tube Adaptor: For 16 mm outer diameter tubes, included as standard.
Additional Information: The document notes that 1 McF unit is approximately equal to 3 x 108 CFU/ml. Both models come with a 2-year warranty.
See more

Catalog excerpts

Densitometers-1

Densitometers DEN-1 for 0.5 - 4.0 McFarland units DEN-1B for 0.00 - >7.5 McFarland units, battery powered

 Open the catalog to page 1
Densitometers-2

Densitometers DEN-1 and DEN-1B Compact and efficient benchtop densitometers for measuring turbidity of cell suspensions in a variety of life-science applications. The densitometers are designed and factory calibrated to measure turbidity in the range of 0.5 to 4.0 McFarland units (DEN-1) 0.00 to >7.5 McFarland units (DEN-1B) with a small standard deviation. If required, they can deliver a wider measurement range (up to 15.00 McFarland units), but with a greater standard deviation. n Measurement range: 0.5 to 4.0 McFarland units (DEN-1), 0.00 to >7.5 McFarland units (DEN-1B) Measurement time:...

 Open the catalog to page 2
Densitometers-3

Densitometers – specification Densitometers Light source Wavelength RangeMcFarland units Measurement time Tube diameter, external diameter Sample volume Display / display resoluton External power supply Independant power supply Ambient temperature range Power supply Accessories DEN MCF STDS set of liquid McFarland standards in 16mm ø glass tubes. 0.5 / 1.0 / 2.0 / 3.0 / 4.0 can be decanted into an alternative tube (shelf-life no longer valid). Requires D16 adaptor, supplied with DEN-1 / DEN-1B. DEN MCF 18 STDS set of gel McFarland standards in 18mm ø glass tubes. 0.0 / 0.5 / 3.0 / 6.0 / >7.5....

 Open the catalog to page 3

All Grant Instruments catalogs and technical brochures

  1. LSB Aqua Pro

    4  Pages

  2. OLS Aqua Pro

    4  Pages

  3. JB Academy

    2  Pages

  4. JB Nova

    2  Pages

  5. SUB Aqua Pro

    2  Pages

  6. Inspissator

    3  Pages

  7. Aspirator

    3  Pages

  8. Centrifuges

    10  Pages

  9. Flow Heater

    3  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.