Overview: The Grace Instrument M9190 Nanodarcy Permeameter is engineered to measure the low and ultra-low permeability of tight core samples from shale and other tight gas reservoirs. It employs a pulse decay procedure with high-pressure nitrogen and includes a confining pressure system to simulate overburden pressure at reservoir conditions.
Specifications:
  • Working Pressure: Pore pressure up to 2,500 psi
  • Confining Pressure: Up to 10,000 psi
  • Temperature Range: Ambient to 175 °C, ± 0.5 °C
  • Permeability Range: 10 nanodarcy to 0.5 millidarcy
  • Core Length: 0.125'' to 3.0''
  • Core OD: 1'', 1.5''
  • Pulse Range: 5 - 200 psi
Operational Features:
  • Mercury-free design
  • Compatibility with sour gas
  • Digital display and accurate measurement of gas volume
  • Three-volume function achieved by a two-chamber design
Procedure: The device measures core permeability at a set pore pressure, then propagates a differential pressure pulse through the sample. The M9190 software records upstream pressure, downstream pressure, differential pressure across the sample, and time. Data can be exported in .CSV format for further analysis.
Contact Information: Grace Instrument, 10770 Moss Ridge Rd., Bldg. B, Houston, TX USA 77043. Phone: 713-783-1560, Fax: 713-974-7144.
See more

Catalog excerpts

M9190-1

10770 Moss Ridge Rd., Bldg. B • Houston, TX USA 77043 • Phone: 713-783-1560 • Fax: 713-974-7144 Low Permeability Measurement Apparatus The Grace Instrument M9190 Nanodarcy Permeameter is designed to measure the low and ultra low permeability of tight plug sized core samples from shale and other tight gas reservoirs. It uses a pulse decay procedure and high pressure nitrogen to measure permeability. Confining pressure system is included for simulating overburden pressure at reservoir condition. The M9190 Nanodarcy Permeameter can operate at pore pressure up to 2,500 psi and confining pressure up to 10,000 psi, pulse range between 5 to 200 psi. The convection oven is used to provide a constant temperature during testing. Simple Operation and Dependable Measurement The M9190 Nano-Darcy Permeameter measure the core permeability at a set pore pressure, then propagates a differential pressure pulse through the sample, M9190 software will record upstream pressure, downstream pressure, differential pressure across the sample, and time. The recorded data can then be exported in .CSV format into a spreadsheet file. Operational Features • Mercury-free design • Compatibility with Sour Gas • Digital Display and Accurate Measurement of Gas Volume • Three-volume Function Accomplished by Two-chamber Design Specifications: Working Pressure: Confining Pressure: Temperature Range: Permeability Range: Core Length: Core OD: Pulse Range: [email protected] • www.gra

 Open the catalog to page 1

All Grace Instrument Co. catalogs and technical brochures

  1. M3900

    1  Page

  2. M9700

    1  Page

  3. M9250

    1  Page

  4. M9103

    1  Page

  5. M9106

    1  Page

  6. M7540

    1  Page

  7. M1722

    1  Page

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About DirectIndustry
DirectIndustry is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. DirectIndustry is an entity of VirtualExpo Group.