Applications:The BP-Rumore Noise
Test Bench is designed for measuring noise produced by taps and hydraulic appliances in water distribution systems. It is primarily used for research and development of sanitary taps, valves, and hydraulic components, as well as for product comparison. The test bench complies with EN ISO 3822 and NF31-014 annex C standards, allowing noise measurement without the need for an anechoic room. It is suitable for laboratory use as ambient noise does not affect measurements. However, it is not intended for official measurements.
Technical Data:- Dimensions: 1500 mm (Length) x 1100 mm (Depth) x 1800 mm (Height)
- Weight: Approximately 250 kg
- Electrical Supply: 400 V 3Ph+N+G - 50 Hz
- Power: 2.0 kW
- Flow Rate: 0.1 to 47 L/min
- Water Temperature: Ambient
Main Technical Specifications:- Aluminium and laminated panel structural frame
- Moving carriage with test rig and transducer
- Pipes compliant with standards
- Internal pump with inverter control and water recovery tank
- Adjustable measurement tube on sliding supports
- Stainless steel test inlets with pressure transducer and hydrophone connections
- Hydrophone for fluid noise measurements
- Analogical acoustic noise meter with A weighting, slow & fast functions, and a range of 0 to 117 dB
- Acquisition card for supply and signal conditioning via dedicated software
Accessories:- Hydraulic resistance adaptor kit
- Class A/B/C/D tubular hydraulic resistances in a wooden box
- Sample noise generator (45 dB) compliant with IEN ISO 3822-1
Installed Measuring Instruments:- Temperature: Accuracy ±0.3 °C, Resolution 0.01 °C, Pt100 probe
- Pressure: Accuracy ±0.05 bar, Resolution 0.01 bar, Range 0 to 10 bar
- Flow Rate: Accuracy ±0.25%, Resolution 0.01 L/min, Range 0.1 to 47 L/min
- Noise: Accuracy ±0.5 dB at 45 dB, Resolution 0.1 dB, Range 0 to 117 dB
Certification:
All instruments are certified for final testing, stability, and accuracy, traceable to Accredia standards.
See more