Microwave Signal Generator 2500B Series
21Pages

{{requestButtons}}

Catalog excerpts

Microwave Signal Generator 2500B Series - 1

Technical Datasheet Microwave Signal Generator Ultra-Low Phase Noise and Fast-Switching Speed in a Single Unit <ZlA Giga tronics

Open the catalog to page 1
Microwave Signal Generator 2500B Series - 5

2500B Series Microwave Signal Generator Specifications formally describe product performance. A specification is a numerical value, or range of values, that bounds the performance of a product parameter. The product warranty covers the performance of parameters described by specifications. Products meet all specifications when shipped from the factory. Typical and Nominal describe product performance that is useful in the application of the product, but is not covered by the product warranty. They describe performance that is expected of a given product at room temperature after 30 minutes...

Open the catalog to page 5
Microwave Signal Generator 2500B Series - 6

2500B Series Technical Specifications Frequency 2502B 2508B Range (with option 18) Frequency Accuracy Same as time base Frequency Resolution Power Slope Phase Adjust Phase Adjust Resolution Frequency Stability Internal Reference Output Aging Rate > +5 dBm square wave into 50 Ω External Reference Frequency Input Frequency Deviation Recommended Input Level Voltage Range Reference Tuning Lock/Level Indicator (CW Mode Only) > -5 dBm into 50 Ω for 10 MHz > +5 dBm to < +8 dBm into 50 Ω for 100 MHz 0 to 10V 2 ppm/V nominal 0.2 ppm/V nominal with option 28 Frequency Bands Band After 30 days...

Open the catalog to page 6
Microwave Signal Generator 2500B Series - 7

Maximum Leveled Output Power in dBm Specification applies into 50 O load over 0 °C to 35 °C range and degrades < 2 dB from 35 °C to 55 °C Number in () is for instruments with mechanical step attenuator option 26 Number in [ ] is for instruments with electronic step attenuator option 27 Giga-tronics 2540 Maximum Unleveled Output Power with Step Attenuator (Nominal) Giga-tronics 2550B Standard Maximum Unleveled Output Power with Step Attenuator (Nominal) 4 Specification is typical below 10 MHz 5 Specification for model 2502B applies to its maximum frequency of 2.5 GHz 5 Model 2550B frequency...

Open the catalog to page 7
Microwave Signal Generator 2500B Series - 8

Option 20 Maximum Leveled Output Power in dBm Specification applies into 50 O load over 0 °C to 35 °C range and degrades < 2 dB from 35 °C to 55 °C Number in () is for instruments with mechanical step attenuator option 26 Number in [ ] is for instruments with electronic step attenuator option 27 X-Band Power Boost is a special feature included in 2520B with Option 20, and when enabled, increases the maximum unleveled output power to X-Band Power Boost On (Nominal] 4 Specification is typical below 10 MHz 5 Specification for model 2502B applies to its maximum frequency of 2.5 GHz 5 Model...

Open the catalog to page 8
Microwave Signal Generator 2500B Series - 9

RF Power Level Accuracy (dB) Specifications apply over 15 °C to 35 °C range and degrades < 0.1 dB/°C outside that range Standard performance

Open the catalog to page 9
Microwave Signal Generator 2500B Series - 10

RF Power Level Accuracy (dB) Specifications apply over 15 °C to 35 °C range and degrades < 0.1 dB/°C outside that range Performance with mechanical step attenuator option 26: Frequency Range Performance with electronic step attenuator option 27: Frequency Range Minimum Leveled Output Power in dBm Specification applies over 0 °C to 35 °C range and degrades < 2 dB from 35 °C to 55 °C Number in ( ) is for instruments with mechanical step attenuator option 26 Number in [ ] is for instruments with electronic step attenuator option 27 Model Additional Output Power Specifications Power Offset (CW...

Open the catalog to page 10
Microwave Signal Generator 2500B Series - 11

List Mode Number of Points Frequency Settling Amplitude Settling Trigger Modes Continuous, Single Step, Single Sweep Step Time External, GPIB GET, Software Sweep Modes Digital Sweep Sync Out Delay Resolution Frequency Change, ∆F = | (F stop x N stop) – (F start x N start) | where N is the value in the Frequency Band Table Time for frequency to settle within 50 kHz of final value after a frequency switch Settling time not specified with FM turned on ∆F = | (F stop x N stop) – (F start x N start) | where N is the value in the Frequency Band Table 14 Time for amplitude to settle within 0.1 dB...

Open the catalog to page 11
Microwave Signal Generator 2500B Series - 12

Frequency and Power Sweep Frequency Sweep Modes Frequency Sweep Range Full Frequency Range Frequency Range Resolution Ramp Frequency Sweep Resolution Analog Sweep, 401, 801 or 1601 points Analog Sweep Mode Provides very fine resolution sweep, for use with Scalar Network Analyzers Ramp Frequency Sweep Time Frequency Sweep Time Resolution Step Sweep Step Time Step Sweep Time Resolution Ramp Power Sweep Ramp Power Sweep Steps Ramp Power Sweep Resolution Ramp Power Sweep Time16 Ramp Power Time Resolution Power Slope (CW Mode, List Mode) Ramp Output Z-Axis Blanking +5V (Positive polarity only) 5...

Open the catalog to page 12
Microwave Signal Generator 2500B Series - 13

Spectral Purity Maximum leveled output power or +10 dBm, whichever is lower. Specification for harmonics above instrument frequency range are typical 100 kHz to 10 MHz Maximum leveled output power or +10 dBm, whichever is lower. Specification for sub-harmonics above instrument frequency range are typical 100 kHz to 2.0 GHz Specification is for offsets > 300 Hz Specification is -45 dBc + 20 log(1/N) dBc typical for offsets < 300 Hz 100 kHz to 10.1 GHz Offset > 5 MHz at maximum leveled power. Applies in CW only Specification is -35 dBc for frequencies < 50 MHz on 2550B model only...

Open the catalog to page 13
Microwave Signal Generator 2500B Series - 14

Phase Noise SSB Phase Noise - Standard Carrier Offset from Carrier (dBc/Hz)

Open the catalog to page 14
Microwave Signal Generator 2500B Series - 15

Phase Noise SSB Phase Noise - Option 28 2500B Option 28 Absolute SSB Phase Noise Performance (Nominal) Offset from Carrier (Hz) Measured phase noise plotted without spurs

Open the catalog to page 15
Microwave Signal Generator 2500B Series - 16

Frequency Modulation Table (Specification applies for frequencies above 10 MHz) Rate (3 dB bandwidth) Accuracy 5 kHz rate 1 MHz rate ± 5% at 5 kHz rate with 1 Vpeak input, 12.024 kHz/V sensitivity ± 5% at 1 MHz rate with 1 Vpeak input, 2.4048 MHz/V sensitivity Sensitivity Range Sensitivity Resolution Input Range Input Impedance Phase Modulation (Specification applies for frequencies above 10 MHz) Rate (3 dB Bandwidth) Peak Deviation ± 5% at 1 kHz rate with 1 Vpeak input, 3.83 rad/V sensitivity Sensitivity Range Sensitivity Resolution Input Range Input Impedance Amplitude Modulation22...