Operation Manual MICRO III/IV – Pump Pump for G200-Series GfG GESELLSCHAFT FÜR GERÄTEBAU MBH▐ KLÖNNESTRASSE 99▐ D-44143 DORTMUND▐ TELEPHONE +49 / (0) 2 31 / 5 64 00 –0 ▐ FAX +49 / (0) 2 31 / 51 63 13 INFO@GFG-MBH.COM ▐ WWW.GASDETECTION.BIZ▐ AMTSGERICHT DORTMUND HR B 2742▐ CERTIFIED AS PER DIN EN ISO 9001:2000▐
Open the catalog to page 1General description and Design 3 Test of battery capacity 5 Spare parts and accessories 6 EC-Type Examination Certificate 8
Open the catalog to page 2Introduction For your Safety According to § 3 of the law about technical working media, this manual points out the proper use of the product and serves to prevent dangers. All individuals who have or will have the responsibility for using and servicing this product must carefully read this manual. This product will do the job designed to do, only, if it is used, serviced, maintained and controlled in accordance with the manufacturer's instructions. If not used properly the warranty of the product is voided. Adjustments in the service mode must be done by experts only. Before operating the...
Open the catalog to page 3Operational Note General hints • Attention! Airstreams can cause measurement faults. The air that enters the detector by the diffusion inlet can distort the concentration of the sampled gas. To avoid such an inflow of air, please cover the diffusion inlet with your hand. Make sure that diffusion inlet is not completely closed to avoid a pressure increase inside the sensor. For sampling gases out of sewers, rooms or drains a hose (with or without telescopic probe) can be used that is plugged on the intake. As the response time heavily depends on the inside volume of the intake appliance the...
Open the catalog to page 4Battery capacity The MICRO Ill-pump is powered by two 1.5 V AA alkaline batteries (mignon). These batteries allow a continuous operation of up to 8 hours. The operational time may be reduced by increased load of the pump motor (e.g. through bended sample hose). The correct battery types are: DURACELL PROCELL MN 1500 LR6 AA or INDUSTRIAL BY DURACELL ID1500 AA (LR6). The pump automatically turns off, if the battery voltage falls below the minimum level needed for proper functioning. • Turn pump on (press button) Activation of battery test The battery capacity is indicated by a multi-signal...
Open the catalog to page 5The unit continuously checks its processor. If a fault is recognized a visual alarm is given. Solution: Change battery, if fault is not solved, send unit to the manufacturer. Processor fault Cleaning Give the MICRO III-pump a short sight check after use. Use a damp cloth to remove stains or soiling from the casing. Never use solvents or cleaning agents! Maintenance and repair A function test should be carried out at least once a year and contain: • the charge status of the battery • sight check • check of sampling performance The test has to be carried out by an expert and a written...
Open the catalog to page 6Worldwide Supplier of Gas Detection Solutions 180-000.19_OM_Pump.doc Version 25th October 2017 from firmware version 2.01 We reserve the right of modification.
Open the catalog to page 7EG-Baumusterpriifbescheinigung - Richtlinie 94/9/EG -Cerate und Schutzsysteme zur bestimmungsgeinafien Verwendung in explosionsgefahrdeten Bereichen DMT 03 ATEX E 072 X Gerat: Pumpe Typ MICRO Ill-Pump Die Bauart dieses Gerates sowie die verschiedenen zulassigen Ausfilhrungen sind ii Baumusterpriifbescheinigung festgelegt. (8) Die Zertifizierungsstelle der Deutsche Montan Technologie GmbH, benannte Stelle Nr. 0158 gemafi Artikel 9 der Richtlinie 94/9/EG des Europaischen Parlaments und des Rates vom 23. Marz 1994, bescheinigt, dass das Gerat die grundlegenden Sicherheits- und...
Open the catalog to page 8EU Declaration of Conformity GfG Gesellschaft fur KlonnestraBe 99 TTT _ 44143 Dortmund MICRO Ill-Pump Tel: +49(231) 56400-0 E-Mail: info@gfg-mbh.com Edited: 03.11.2005 Amended: 31.07.2017 www.afg.biz GfG Gesellschaft fur Geratebau mbH develops produces and sells gas sensors and gas warning devices which are subject to a quality management system as per DIN EN ISO 9001. Subject to supervision by means of a quality system, surveilled by the notified body, DEKRA EXAM GmbH (0158), is the production of electrical apparatus of instrumentation Group I and II, categories Ml,...
Open the catalog to page 9All GfG - Gesellschaft für Gerätebau catalogs and technical brochures
-
MiniCal III
4 Pages
-
The leak detector G300 II
1 Pages
-
Smart Pump G400-MP2
12 Pages
-
Microtector II G450
4 Pages
-
Drop-in charger
15 Pages
-
Dockingstation DS404
17 Pages
-
Smart Pump G400-MP2
2 Pages
-
EC22
2 Pages
-
ZD22
2 Pages
-
IR22
2 Pages
-
CS22
2 Pages
-
CC28DA
2 Pages
-
CC28D
2 Pages
-
EC28i
2 Pages
-
EC28Di
2 Pages
-
EC28DB
2 Pages
-
EC28DAR
2 Pages
-
EC28DAB
2 Pages
-
EC28DA
2 Pages
-
EC28D
2 Pages
-
EC28B
2 Pages
-
CI21
2 Pages
-
CS21_Flammable
2 Pages
-
CS21_Refrigerant
2 Pages
-
GMA200 MGSS
4 Pages
-
MiniCall III
2 Pages
-
MiniCal Refrigerant System
4 Pages
-
Transmitter EC24
2 Pages
-
Transmitter CC28
2 Pages
-
Transmitter IR29
4 Pages
-
9025-XPS
2 Pages
-
GMA200-MGSS
2 Pages
-
G300 III
2 Pages
-
EC28
2 Pages
-
CS21
2 Pages
-
Micro III
2 Pages
-
GMA81/ GMA81 A
1 Pages
-
Controller GMA200-MT
4 Pages
-
Controller GMA200-MW
4 Pages
-
GMA313
2 Pages
-
Transmitter CC28 DA
2 Pages
-
Transmitter CC28 D
2 Pages
-
Transmitter ZD21
2 Pages
-
Transmitter IR24
4 Pages
-
Dockingstation DS400
2 Pages
-
GfG Products & services Brochure
15 Pages
-
Microtector II G450 Brochure
4 Pages
-
Transmitter CC28 Brochure
2 Pages
-
GMA36 Pro Brochure
2 Pages
-
GMA 40 Series Broschure
2 Pages
-
Transmitter EC24
2 Pages
-
GMA80/81/84/88 Brochure
4 Pages
Archived catalogs
-
Test gases
2 Pages
-
MICRO Motorised pump
2 Pages
-
GMA101/103/104
4 Pages