PANEL+

PANEL+

PANEL+

Product catalog summary
Specifications
The keyboard features 109 keys in a US layout, utilizing carbon contact keyswitch technology with an approximate lifespan of 2 million cycles per key. It is constructed with silicone housing available in white and black, and includes a blue backlight with five adjustable brightness levels. The interface is USB 2.0 with a cable length of 15cm. It operates within a temperature range of 0°C to +55°C and can be stored between -30°C and +70°C. The dimensions are 365mm (L) x 134mm (W) x 14.2mm (T) with a weight of approximately 415g. The product comes with a 2-year warranty and certifications including EN60601-1-2 Ed 4, CE, FCC, RoHS, and REACH.
Special Features
The keyboard includes a num pad and track pad, is designed to match surgical panels for wall mounting, and offers excellent tactile feedback. It features an integrated Clean Mode for cleaning, indicated by a breathing backlight.
Certifications
The keyboard is pre-certified according to EMC standards, specifically IEC 60601-1-2: 2014, which addresses electromagnetic disturbances. Additional standards include IEC 61000-4-2, IEC 61000-4-3, and IEC 61000-4-8.
Installation Recommendations
Two installation types are outlined: insulated installation and installation in electrically conductive system parts. For insulated installations, no additional measures are required. For conductive installations, it is recommended to separate the shield connection to avoid ground loops. A potential equalization line is advised to prevent excessive potential differences. The USB cable should not be stripped more than 25mm to maintain stability.
Cleaning Instructions
To enable Clean Mode, press and hold the lock key for 5 seconds. The backlight will indicate activation. Further cleaning details are available on the manufacturer's website.
Contact Information
GETT Gerätetechnik GmbH and GETT Asia Ltd. provide contact details for their locations in Germany, Hong Kong, and China for further inquiries.
See more

Catalog excerpts

PANEL+-1

PRIME SERIES PANEL+ (WITH BACKLIGHT) PANEL SPECIAL FEATURES Adjustable brightness of backlight (PANEL+) 109 keys incl. num pad and track pad Match with surgical panel (wall mount) Excellent tactile feeling Integrated Clean Mode for cleaning EN60601-1-2 Ed 4 certified SPECIFICATION Number of keys: Keyswitch technology: Switching cycles: Housing material: Housing color: Backlight: Interfaces: 109 (US-Layout) Carbon contact Approx. 2 mill. (per key) Silicone White & black Blue color, 5-level adjustable brightness USB 2.0 Cable length: Operating temperature: Storage temperature: Dimensions: Weight: Warranty: Certification: 15cm 0°C ~ +55°C -30°C ~ +70°C 365 (L) x 134 (W) x 14.2 (T) mm Approx. 415g 2 years EN60601-1-2 Ed 4, CE, FCC, RoHS, REACH Product Description CLEANTYPE® PRIME PANEL+ CLEANTYPE® PRIME PANEL CLEANTYPE® PRIME PANEL+ CLEANTYPE® PRIME PANEL White White Black Black US/DE/JP/KR/THAl US/DE/JP/KR/THAI US/DE/JP/KR/THAI US/DE/JP/KR/THAI Other layouts, configurations and interfaces on request. Customized key function on request (Firmware update via USB) GETT Gerätetechnik GmbH GETT Assembly (Asia) Co., Ltd. Mittlerer Ring 1 08233 Treuen Germany Room 806-7, Kowloon Plaza No. 485 Castle Peak Road Kowloon, Hong Kong, China 401 & 501, Building 1, No.97, Fuyong Area 107 Road Baishixia Community, Fuyong Street Baoan District, Shenzhen, China gett-asia.com gett-gro

 Open the catalog to page 1
PANEL+-2

EN60601-1-2 ED 4 CERTIFIED EN 60601-1-2Ed 4 applies to the basic safety and essential performance of Medical Equipment (ME) equipment and ME systems in the presence of electromagnetic disturbances and to electromagnetic disturbances emitted by ME equipment and ME systems. SHORT OPERATING INSTRUCTIONS Press and hold lock key for 5s to enable Clean Mode. Clean Mode is indicated by breathing backlight. Adjust backlight level with this button. Five level can be adjusted including off. CLEANING INSTRUCTION For further details please refer to below link: GETT Gerätetechnik GmbH GETT Assembly (Asia)...

 Open the catalog to page 2
PANEL+-3

INSTALLATION RECOMMENDATION FOR GCQ PANEL KEYBOARDS The keyboard is pre-certified according to the relevant EMC standards: IEC 60601-1-2: 2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests GETT Gerätetechnik GmbH GETT Assembly (Asia) Co., Ltd. Mittlerer Ring 1 08233 Treuen Germany Room 806-7, Kowloon Plaza No. 485 Castle Peak Road Kowloon, Hong Kong, China 401 & 501, Building 1, No.97, Fuyong Area 107 Road Baishixia Community, Fuyong Street Baoan District, Shenzhen,...

 Open the catalog to page 3
PANEL+-4

The Prime panel is a built-in keyboard, it was tested "alone" reference to above EMC standard without taking the customer system into account. The keyboard integrator is responsible for complying with their required standards for the overall system and cannot be guaranteed by the keyboard manufacturer. For reasons of interference immunity, the reference potential of the installation / system must be considered when integrating the keyboard. To ensure that the keyboard works properly, we recommend that you observe the following installation recommendation. A distinction is made between two types...

 Open the catalog to page 4
PANEL+-5

Distances when installing in electrically conductive systems: When installing in metal panels (machines, systems, control cabinets, process control systems, etc.), ensure that there is sufficient distance between the conductive parts and the keyboard circuit board. The sufficient distance is defined by the system integrator following the necessary ESD test voltages requirement and the safety distances to electrically conductive components of the system. If the distances are too short, the capacitive touchpad and the keyboard can be influenced and damaged. USB cable: To ensure stable USB operation,...

 Open the catalog to page 5
PANEL+-6

These are general installation recommendations provided by keyboard manufacturer. It’s important to note that the necessary integration actions may differ depending on the system and must be ensured by the system integrator, considering the standards and regulations relevant for him. GETT Gerätetechnik GmbH GETT Assembly (Asia) Co., Ltd. Mittlerer Ring 1 08233 Treuen Germany Room 806-7, Kowloon Plaza No. 485 Castle Peak Road Kowloon, Hong Kong, China 401 & 501, Building 1, No.97, Fuyong Area 107 Road Baishixia Community, Fuyong Street Baoan District, Shenzhen, China gett-asia.com gett-group.com...

 Open the catalog to page 6

All GETT North America LLC catalogs and technical brochures

  1. TOUCH+

    2  Pages

  2. CK4W

    3  Pages

  3. XTRA SERIES

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About DirectIndustry
DirectIndustry is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. DirectIndustry is an entity of VirtualExpo Group.