1. Catalogs
  2. GeoSIG Ltd
  3. VE-5x-DH Seismometer

VE-5x-DH Seismometer

VE-5x-DH Seismometer

VE-5x-DH Seismometer

Product catalog summary
Features
  • Sensitivity: 1000 V/m/s differential
  • Built-in compass, tilt, temperature, and humidity sensors
  • Robust downhole housing suitable for borehole diameters of 100 mm and larger
  • Proprietary iSensor™ interface
  • Dynamic Range: > 125 dB (0.9 - 15 Hz), > 120 dB (0.9 - 30 Hz)
  • Full Scale: 20 mm/s (± 10 mm/s)
  • Bandwidth: 1 to 80 Hz (-3 dB) or 8.0 s (0.125 Hz) to 160 Hz (BB version)
Outline

The VE-53-DH is a downhole, triaxial short period seismometer designed for seismic monitoring. It features a geophone mass-spring system with electronic feedback, housed in robust austenitic stainless steel. It offers stability against temperature fluctuations and aging, requiring no mass locking. The electronic offset adjustment simplifies installation, and the iSensor™ interface facilitates deployment with built-in sensors. It is compatible with all GeoSIG solutions.

Specifications
  • Configurations: Triaxial, Biaxial, Uniaxial
  • Axes Alignment: X – Y – Z H – H – V
  • Sensitivity: 1000 V/m/s differential
  • Full Scale Range: 20 mm/s (± 10 mm/s)
  • Dynamic Range: > 125 dB (1 - 15 Hz), > 120 dB (1 - 30 Hz)
  • Non-linearity: ± 0.05% of full scale
  • Cross Axis Sensitivity: ± 1% typical, ± 3% maximum
  • Bandwidth: 1.1 s (0.9 Hz) to 89 Hz, 8.0 s (0.125 Hz) to 160 Hz (BB version)
  • Damping: 0.7 critical
  • Full Scale Output: 0 ± 10 V differential
iSensor™ Interface

The iSensor™ interface is a proprietary hardware and software interface for operation, control, logging, and data export, featuring a compass, tilt sensor, temperature sensor, and humidity sensor.

Power
  • Supply Voltage: 9 to 18 VDC
  • Consumption: 60 mA typical, 100 mA max. @15 VDC
  • Overvoltage Protection: All pins protected
Testing

Test input activated by applying a 12 VDC voltage to generate a pulse output at 50% of full scale.

Environment/Housing
  • Housing Type: Austenitic stainless steel
  • Housing Size: φ 89 mm x 502 mm
  • Weight: 7.5 kg
  • Index of Protection: Watertight up to 10 bar (100 m)
  • Temperature Range: -20 to 70 °C (operating), -40 to 75 °C (non-operating)
Options

Cable & connector options and surface control unit available.

Ordering Information

Specify type of VE-53-DH, full scale range, depth of deployment, cable length, and other options.

See more

Catalog excerpts

VE-5x-DH Seismometer-1

□ Built-in compass as well as tilt, temperature and humidity sensors □ Extremely robust downhole housing □ Suitable for borehole diameters of 100 mm and larger □ Proprietary iSensor™ interface □ Dynamic Range > 125 dB (0.9 - 15 Hz) > 120 dB (0.9 - 30 Hz) □ Full Scale 20 mm/s (± 10 mm/s) The VE-53-DH is a downhole, triaxial short period seismometer designed for seismic monitoring applications. Based on a state of the art geophone mass-spring system with electronic feedback, the VE-53-DH seismometer is ideally suited for installations with low to moderate noise. The VE-53-DH is housed in an extremely robust downhole housing of austenitic stainless steel. Offering a remarkable stability under temperature fluctuations or against aging, the VE-53-DH also requires no mass locking due to the innovative design of the unit. The VE-53-DH is equipped with electronic offset adjustment features that make its installation very user friendly. This powerful feature allows the users to install the VE-53-DH without mechanical offset adjustment and fine levelling. The advanced iSensor™ interface allows easy deployment using built-in hardware like compass as well as tilt, temperature and humidity sensors. The VE-53-DH seismometer is directly compatible with all GeoSIG solutions. Gain Factor (K) Poles (P) Gain Factor (K) Poles (P) GeoSIG svw'ss made to measure Link to VE-53 sensor response files in the IRIS NRL library

 Open the catalog to page 1
VE-5x-DH Seismometer-2

Sensor Element Type: Dynamic Range: Non-linearity: Cross Axis Sensitivity: Bandwidth: Damping: Full Scale Output: Measuring Range (see plot) Power Supply Voltage: Consumption: Overvoltage Protection: Testing Test input: Environment/Housing Housing Type: Housing Size: Weight: Index of Protection: Temperature Range: General Characteristics Configurations: ** H: Horizontal, V: Vertical Sensitivity: 1000 V/m/s differential ( 2 x 500 V/m/s ) Full Scale Range: 20 mm/s (± 10 mm/s) nominal output Over damped geophones > 125 dB (1 - 15 Hz) > 120 dB (1 - 30 Hz) ± 0.05 % of full scale maximum ± 1 % typical...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.