1. Catalogs
  2. GEO Calibration Inc.
  3. Model 2000 SP Spec Sheet

Model 2000 SP Spec Sheet

Model 2000 SP Spec Sheet

Model 2000 SP Spec Sheet

Product catalog summary
Product Overview: The Model 2000 SP is a humidity and temperature calibrator designed for precise environmental control. It is traceable to NIST standards and offers a 17025 validation certificate for an additional fee.
General Specifications: The calibrator operates in ambient conditions of 18°C to 28°C and up to 80% RH. It can be stored at temperatures from -20°C to 50°C and humidity levels from 0% to 95% RH, non-condensing, at altitudes up to 2000 meters.
Calibration System Accuracy: Temperature accuracy is ±0.10°C or better, typically ±0.05°C, while humidity accuracy is ±1.00% RH or better, with specially tuned systems achieving ±0.60% RH.
Controller and Stability: The device uses a PID controller with temperature uniformity of 0.10°C or better and humidity uniformity of 0.30% RH. Stability is maintained at 0.05°C for temperature and 0.15% RH for humidity.
Resolution and Control Range: Display resolution is 0.01°C for temperature and 0.01% RH for humidity. The control range for temperature is 5°C to 55°C, with varying humidity ranges depending on the temperature.
Consumables: The water reservoir holds 200 ml of distilled water, with a refill period of 15 days to 1 month. The desiccant type is a molecular sieve, requiring replacement when 3/4 used.
Mechanical Specifications: The generator dimensions are 55.33 cm x 40.65 cm x 24.65 cm, with a chamber diameter of 14.50 cm and depth of 20.00 cm. The unit weighs 15.5 kg and operates on a 12 Volt DC power supply.
Accessories and Ordering Information: Various door shapes and port quantities are available, with customization options. Accessories include desiccant canisters, expansion chambers, and calibration documents.
Contact Information: For more details, visit www.GeoCalibration.com or contact sales at [email protected].
See more

Catalog excerpts

Model 2000 SP Spec Sheet-1

Model 2000 SP PRODUCT OVERVIEW HUMIDIT Y AND TEMPERATURE CALIBRATORS

 Open the catalog to page 1
Model 2000 SP Spec Sheet-2

CALIBRATION TO (17025 TRACEABLE TO NIST) OPERATING AMBIENT CONDITIONS STORAGE CONDITIONS ALTITUDE CALIBRATION SYSTEM ACCURACY UNIFORMITY STABILITY WARM-UP PERIOD RESOLUTION TRANSFER STANDARD CERTIFICATE TEMPERATURE No Charge 17025 Validation (Additional Fees Apply) 18 °C to 28 °C HUMIDITY TEMPERATURE HUMIDITY 2000 m TEMPERATURE HUMIDITY CONTROLLER TYPE TEMPERATURE Up to 80 % RH -20 °C to 50 °C 0 % to 95 % RH (non-condensing) ± 0.10 °C Or Better Typically ± 0.05 °C *Based on Probe Accuracy ± 1.00 % RH Or Better Specially tuned systems can be as good as ± 0.60 % *Based on Probe Accuracy PID Controller...

 Open the catalog to page 2
Model 2000 SP Spec Sheet-3

‘‘Information subject to change, please visit our website for updates at www.geocalibration.com