Pressure Compensation of a CO2 Sensor

Product catalog summary
Abstract
Non-dispersive infra-red (NDIR) sensors measure gas concentration by detecting IR light absorption. Measurement accuracy is influenced by environmental factors like temperature and pressure, which affect gas molecule density. This document discusses the impact of these factors on CO2 sensor measurements and compensation methods to enhance accuracy.
Basic Principles
NDIR sensors measure CO2 concentration based on molecule count in a fixed volume. Temperature and pressure variations affect this count, necessitating compensation for accurate measurements.
Environmental Factors
Temperature and pressure significantly influence gas molecule density, as per the ideal gas law. Higher temperatures cause molecules to occupy more space, reducing density, while increased pressure compresses molecules, increasing density.
Gas Vibrations
CO2 molecules exhibit four vibrational modes, with asymmetrical stretch and bending modes absorbing IR radiation. The asymmetric stretch mode, absorbing at 4.26µm, is commonly measured by NDIR sensors.
Absorption
Absorption at 4.26µm is mainly due to asymmetrical vibrations. Pressure and temperature changes affect absorption characteristics, with pressure broadening the absorption peak and temperature reducing absorption.
Compensation Options
Temperature and pressure affect sensor accuracy. Temperature compensation involves controlling sensor temperature or correcting based on characterized behavior. Pressure compensation is crucial at higher gas concentrations, where changes significantly impact accuracy.
Applying Pressure Compensation
Pressure changes alter absorption band shape due to spectral broadening. Compensation uses gas laws, with corrections based on pressure deviations from a reference level. A correction factor maintains measurement accuracy.
Adding Effects of Spectral Broadening
Spectral broadening increases pressure impact. Correction factors from spectral data can be used, with simplifications for certain concentration ranges. Polynomial fits or lookup tables refine corrections.
Conclusion
Accurate CO2 measurement with NDIR sensors requires compensation for temperature and pressure effects. Proper compensation ensures reliable sensor performance.
Introduction
This application note from Gas Sensing Solutions Ltd. (GSS) focuses on pressure compensation for CO2 sensors, emphasizing the importance of correcting measurements for ambient pressure changes, especially above 1% concentration.
Correction Techniques
The document discusses pressure compensation techniques, including constant correction coefficients, lookup tables, and polynomial fits. A 6th order polynomial fit is highlighted for reducing measurement errors to less than 1% in most cases.
Figures and Data
Figures compare theoretical correction coefficients and polynomial fits, showing the impact on measurement errors. Polynomial fits significantly reduce errors compared to uncorrected measurements, which can exceed 43% at certain pressures.
Conclusion
Accurate pressure compensation is crucial for applications from ambient CO2 sensing to high concentration environments. GSS provides guidance and technical support for implementing these corrections.
Important Notice
The document includes a disclaimer about GSS product terms and conditions, emphasizing they are not for life support or critical systems. It clarifies the company's stance on intellectual property and liability.
Contact Information
Gas Sensing Solutions Ltd. is located at Grayshill Road, Cumbernauld, United Kingdom.
Revision History
The document is identified as the first revision, dated May 3, 2021.
See more

Catalog excerpts

Pressure Compensation of a CO2 Sensor-1

APPLICATION NOTE AN001: PRESSURE COMPENSATION OF A CO2 SENSOR ABSTRACT Non-dispersive infra-red (NDIR) sensors work by measuring the amount of IR light absorbed by the target gas in a fixed volume. Gas concentration is proportional to the amount of light absorbed as it passes through the gas. The amount of light absorbed by the target gas, and hence measurement accuracy, is influenced by external environmental factors. The absorption of IR light is dependent on the number of gas molecules present in the fixed volume. Three factors influence the number of molecules in this fixed space, the gas concentration, gas temperature and gas pressure. To accurately measure gas concentration, it is therefore important to understand the effects of temperature and pressure. The general behaviour of a gas, due to changes in temperature and pressure is based on well understood laws. However, the effects of temperature and pressure on the absorption of infra-red light by gas molecules and how they influence measurement accuracy are less well understood. This application note describes the physical effects on the gas molecules due to changes in temperature and pressure. Changes in ambient temperature and pressure will induce CO2 concentration measurement errors unless corrected. It also explains how CO2 sensor measurements are affected by temperature and pressure and how these effects can be compensated for to reduce their impact on sensor accuracy. Gas Sensing Solutions Ltd. Page | 1 For regular updates, sign up at https://gassensing.co.uk Revision 1.0, 3 May 2021 Copyright © 2021 Gas Sensing Solutions Ltd.

 Open the catalog to page 1
Pressure Compensation of a CO2 Sensor-2

Gas Sensing Solutions Ltd. Page | 2 For regular updates, sign up at https://gassensing.co.uk Revision 1.0, 3 May 2021 Copyright © 2021 Gas Sensing Solutions Ltd.

 Open the catalog to page 2
Pressure Compensation of a CO2 Sensor-3

APPLICATION NOTE AN001: PRESSURE COMPENSATION OF A CO2 SENSOR BASIC PRINCIPLES All NDIR sensors measure the concentration of CO2 in a fixed volume, which in turn is based on the number of molecules present. The more molecules there are in the fixed volume, the more IR radiation is absorbed. Both temperature and pressure influence the number of molecules present in the fixed volume and therefore, measurement accuracy will be affected unless steps are taken to compensate for these effects. ENVIRONMENTAL FACTORS The two environmental factors that decide the number of molecules present in a defined...

 Open the catalog to page 3
Pressure Compensation of a CO2 Sensor-4

APPLICATION NOTE AN001: PRESSURE COMPENSATION OF A CO2 SENSOR Figure 1: CO2 Vibration Modes The symmetric stretch does not create a dipole and consequently is not able to absorb any photons. It therefore does not absorb infra-red radiation. The other three vibration modes produce a dipole and absorb photons although at different frequencies. The asymmetric stretch absorbs radiation at 4.26µm, whereas the bending modes absorb at 15µm. Most NDIR based sensors measure the absorption of photons at 4.26um, and this is due to the asymmetric stretch vibration mode. The proportion of molecules vibrating...

 Open the catalog to page 4
Pressure Compensation of a CO2 Sensor-5

APPLICATION NOTE AN001: PRESSURE COMPENSATION OF A CO2 SENSOR ABSORPTION Absorption at 4.26μm due to the symmetric vibration mode is negligible. This is because the average position of the gas molecule is the same as symmetric bending mode case, with no dipole and hence little absorption. However, as the atoms vibrate asymmetrically and form different dipole strengths depending on the energy levels of the electrons, the gas molecules begin to absorb infrared radiation. The different energy levels of the molecules cause absorption at a series of distinct wavelengths as shown below in Figure 2....

 Open the catalog to page 5
Pressure Compensation of a CO2 Sensor-6

APPLICATION NOTE AN001: PRESSURE COMPENSATION OF A CO2 SENSOR Figure 3: Absorption at 4.28µm vs. Pressure Figure 4: Absorption at 4.28µm vs. Temperature As the gas temperature is increased, the gas molecules vibrate more intensely and increase in energy. However, the absorption of infra-red radiation actually decreases, as shown in Figure 4 by the drop in the peak and narrowing of the absorption band. This is due to a reduction in the number of gas molecules in the fixed volume that are available to absorb the IR light. Gas Sensing Solutions Ltd. Page | 6 For regular updates, sign up at https://gassensing.co.uk...

 Open the catalog to page 6
Pressure Compensation of a CO2 Sensor-7

APPLICATION NOTE AN001: PRESSURE COMPENSATION OF A CO2 SENSOR COMPENSATION OPTIONS It has been shown that IR radiation is affected by temperature and pressure. Depending on CO2 measurement accuracy requirements, it may be necessary to compensate for these effects. Temperature Compensation Techniques Changes in temperature effect sensor accuracy in several different ways. As well as those effects that can be described by simple gas laws, the LED and photo-diode, the electronics and the mechanical parts of the sensor will be all be affected by temperature. There are several ways to mitigate the...

 Open the catalog to page 7
Pressure Compensation of a CO2 Sensor-8

APPLICATION NOTE AN001: PRESSURE COMPENSATION OF A CO2 SENSOR APPLYING PRESSURE COMPENSATION As discussed earlier, the shape of the absorption band changes with pressure mainly caused by increased molecule collisions. This effect is known as spectral broadening. As pressure increases, so do the number of collisions due to the higher density of molecules present, and as CO2 has a higher density than air, there is also an effect driven by CO2 concentration. Ignoring the effect of spectral broadening, the measurements can be corrected for pressure based on basic gas laws. Boyle’s Law states that...

 Open the catalog to page 8
Pressure Compensation of a CO2 Sensor-9

APPLICATION NOTE AN001: PRESSURE COMPENSATION OF A CO2 SENSOR Figure 5 below shows how sensor measurements are affected using a constant Y correction factor. Without applying the correction, the sensor measurement would increase erroneously with increasing pressure. Including this correction factor maintains the accuracy of the measurement with changing pressure. Pressure (mbar) Uncorrected Sensor Meaurement (ppm) -Corrected Measurement (ppm) Figure 5: Effects of Pressure on Gas Concentration Page | 9 For regular updates, sign up at https://gassensing.co.uk Copyright © 2021 Gas Sensing Solutions...