Overview
The FCS522 Flow Checker by Furness Controls is a portable device designed for dual range flow measurement, capable of displaying volumetric flow at working pressure and temperature. It is powered by a 9-volt DC supply or a 9-volt battery (PP3) and features USB communication for configuration.

Key Features
  • Dual range flow measurement
  • Display zero switch for calibration
  • USB-B configuration ports compatible with FCO301 software
  • Portable design with integrated carry handle

Flow Measurement Specifications
  • Flow ranges: 2cc/min to 200L/min
  • Accuracy at 20°C: < ± (1% reading +1 digit)
  • Resolution: Better than 0.033% Span
  • Temperature Coefficients: Zero - Manual switch, Span - approx. -0.3% per °C
  • Display: 8 digit starburst LCD
  • Long Term Drift (span): Typically 0.2% per annum

Electrical Specifications
  • Power Requirements: 9 volt DC supply or 9 volt battery (PP3)
  • Battery Life: Up to 250 hours
  • Operating Temperature: 0 to 50°C

Pneumatic Specifications
  • Media Compatibility: Clean dry air or non-corrosive gas, viscosity corrected for air only
  • Overload: 20 x flow range
  • Static Pressure: ±1 bar Gauge

Construction
  • Enclosure: ABS Case
  • Dimensions: 222 x 205 x 80mm (W x H x D) excluding Flow Element
  • Weight: 1.2kg (Excluding Flow Element)
  • Media Compatibility: Air and non-corrosive gases, max 95% humidity non-condensing

Additional Information
All information in this document is provisional and subject to change without notice. Furness Controls offers pressure calibration from 0 to 40 kPa and flow calibration from 0.1 ml/min to 2000 litres/min through its UKAS accredited laboratory.
See more

Catalog excerpts

FCS522 - Dual Range Flow Checker-1

9Volt DC supply or 9Volt battery (PP3) Display zero switch Dual range USB communication Portable The FCS522 Flow Checker provides two ranges of flow measurement. The FCS522 will display volumetric flow at working pressure and temperature. The FCS522 is housed in a plastic enclosure with integrated carry handle mounted on a Laminar Flow Element. The right hand side gives access to the USB-B configuration ports, the On/Off switch, PP3 battery and DC power and the right hand side has the display zero switch. The USB-B configuration ports are for use with the optional FCO301 software. Using the FCO301, the instruments engineering units can be altered. Furness Controls Pressure|Flow|Leak Detection Dual Range Flow

 Open the catalog to page 1
FCS522 - Dual Range Flow Checker-2

Flow Measurement All information in this document is provisional and is subject to change without notice. Furness Controls Pressure |Flow|Leak Detection Beeching Road • Bexhill On Sea East Sussex • UK • TN39 3LG Tel: +44 1424 819980 Email: [email protected] Web: www.furness-controls.com 08/08/2016

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.