1. Catalogs
  2. Furness Controls
  3. FCO730 - Pressure Decay Leak Detector

FCO730 - Pressure Decay Leak Detector

FCO730 - Pressure Decay Leak Detector

FCO730 - Pressure Decay Leak Detector

Product catalog summary
Overview: The FCO730 is an entry-level pressure decay leak detector designed for applications requiring multiple test-heads, such as rotary tables. It can function as a standalone instrument or be integrated with PLCs or PCs for enhanced control and data logging via RS232 or RS485 communications.
Key Features:
  • Clear display of leak and pressure measurements.
  • Supports up to 50 test settings.
  • Pressure ranges from vacuum to 30 bar.
  • Leak resolution of 0.1 Pa or 0.01 cc/min.
  • Front panel calibration port and start/stop controls.
  • Compact, robust steel bench-top case.
Leak Measurement:
  • Leak ranges: ±200.0 Pa, ±2.000 kPa, ±20.00 kPa.
  • Accuracy at 20°C: 10% to 100% range: < ± (1% reading + 1 digit); 0 to 10% range: < ± (0.1% range + 1 digit).
  • Resolution: 4-digit display.
  • Temperature Coefficients: Zero - Automatic; Span - < 0.15% per °C.
  • Long Term Drift (span): < ± 1% per year.
Pressure Measurement:
  • Pressure ranges: ±99.99 mbar to -1 bar to +30.00 bar.
  • Accuracy at 20°C: 10% to 100% range: < ± (1% reading + 1 digit); 0 to 10% range: < ± (0.1% range + 1 digit).
  • Resolution: 4-digit display.
  • Temperature Coefficients: Zero - < 0.05% per °C; Span - < 0.1% per °C.
  • Long Term Drift (span): < ± 1% per year.
Electrical Specifications:
  • Supply Voltage: 24 VDC ± 10%, < 500 mA.
  • Electrical connections: Power, Outputs, Inputs, RS232, RS485.
  • Control Inputs: Up to 12 Opto-isolated, 5 VDC to 24 VDC.
  • Control Outputs: Up to 16 Active High transistor output, 12 VDC to 45 VDC, 120 mA per channel.
Pneumatic Specifications:
  • Media Compatibility: Clean dry air or non-corrosive gas.
  • Air Supply Pressure: Maximum 10 bar gauge, Minimum 5 bar gauge.
  • Regulator Supply Pressure: Maximum 16 bar gauge or 35 bar for 30 bar option.
  • Pneumatic Connections: Various push-in tube connectors.
  • Leak Tightness: < 0.2 cc/Hour.
Construction:
  • Enclosure: Steel construction with paint finish, suitable for 19” 3U rack mounting.
  • Dimensions: Rack Case - 267 x 133 x 296 mm; Bench Case - 232 x 147 x 296 mm.
  • Weight: 5 kg.
See more

Catalog excerpts

FCO730 - Pressure Decay Leak Detector-1

Leak detector Clear display of leak and pressure Up to 50 test settings Pressure ranges from vacuum to 30bar 0.1Pa or 0.01cc/min leak resolution Front panel calibration port Front panel or remote start/stop Compact robust steel bench-top case Communications via RS232 or RS485 The FCO730 has much to offer as an entry level pressure decay Leak Detector. It is ideal for applications where multiple test-heads are needed, such as on a rotary table. It can be easily interfaced to PLCs or PCs where integration is required, or it may be used as a standalone instrument. The communications facility may be used for configuration, control and data logging. Furness Controls Pressure|Flow|Leak Detection Pressure Decay

 Open the catalog to page 1
FCO730 - Pressure Decay Leak Detector-2

Leak Measurement Leak ranges Accuracy @ 20°C Resolution Temperature Coefficients Long Term Drift (span) ±200.0 Pa ±2.000 kPa ±20.00 kPa 200 Pa and 2 kPa models: 10% to 100% range: < ± (1% reading + 1 digit) 0 to 10% range: < ± (0.1% range + 1 digit) 20 kPa model: < ± (1% range + 1 digit) 4 digit display Zero: Automatic Span: < 0.15% per °C < ± 1% per year Pressure Measurement Pressure Ranges Accuracy @ 20°C Resolution Temperature Coefficients Long Term Drift (span) ± 99.99 mbar -1 bar to +4.000 bar -1 bar to +14.00 bar ± 200 mbar -1 bar to +8.000 bar -1 bar to +30.00 bar ± 999.9 mbar -1 bar to...