Overview: The FOX18-XM is an 18W medical power supply designed by FRIWO, featuring a switchmode power supply with an interchangeable adapter system. It complies with IEC 60601-1 standards, ensuring safety and efficiency for medical applications.
Key Specifications:
  • High efficiency up to 86.9% and low no-load power consumption of less than 90mW.
  • Meets DoE efficiency level VI and CoC version 5 ready.
  • Ultra-low leakage current of ≤10µA and operates within a temperature range of 0-50°C.
  • Isolation of 2x MOPP and double housing safety with snapped and welded design.
  • IP42 approved interchangeable plug system available as an accessory.
Safety and Compliance:
  • Complies with IEC/EN/UL60601-1 3.1 Edition and approved for use in Europe, USA, Australia, and Japan.
  • Protection class II with overload, overvoltage, and short circuit protection.
  • EMC compliance with EN55022, EN55011, FCC15, Class B, and EN60601-1-2 standards.
Performance and Reliability:
  • Operates at a universal input voltage range of 90Vac to 264Vac.
  • MTBF of 200,000 hours based on Mil217F calculations.
  • Features low standby losses of 50mW and high efficiency rates.
Mechanical and Environmental Specifications:
  • Dimensions: 77x50x34 mm, Weight: 157 g.
  • Operating humidity range of 10-90% and storage temperature range of -40°C to +70°C.
  • Drop test compliant with IEC 60068-2-32 and vibration test compliant with IEC 60068-2-6.
Adapter and Connector Options:
  • Interchangeable mains plug system and universal output plug system available.
  • Various straight and angled coaxial connectors and jack connectors offered.
See more

Catalog excerpts

FOX18-XM-1

Switchmode power supply MEDICAL with interchangeable adapter system. All products conform to IEC 60601-1. FOX18-XM ? 18 W MEDICAL POWER SUPPLY ft Characteristics High efficiency: up to 86.9% Low no load power consumption <90mW Meets DoE efficiency level VI CoC version 5 ready Ultra-low leakage current <10pA IEC 60601-1-2, 4th edition Ultra-wide operating temperature range 0 - 50°C Isolation 2x MOPP Double housing safety: snapped & welded IP42 approved interchangeable plug system as additional accessories Also available as version with fixed plugs: FOX18-FM FRIWO's medical products are designed for the most sensitive applications and for use under the harshest conditions. They must continue to work perfectly despite falls in the emergency room while protecting patients by offering the lowest possible levels of leakage current. FRIWO develops and manufactures power supply units you can trust. In order to ensure maximum patient safety, FRIWO medical technology devices are equipped with its 2 x MOPP (Means of Patient Protection) safety system and feature a minimal leakage current of < 10 |jA as well as double-sealed housings. The power supply will operate at universal input voltage range of 90Vac to 264Vac in a wide temperature range of 0-50°C without derating. In addition to patient safety, efficiency plays a major role: with low standby losses of 50mW and high efficiency rates the medical FOX line complies with the demanding US efficiency standard DOE (Department of Energy) and its European counterpart ErP (Energy related Products) Another particular highlight of the FOX product family is its interchangeable adapter system with IP42 protection (against drops of water and dampness), which is available as an optional accessory. In practice, this means that the devices can be cleaned using wet cloths - an invaluable advantage in sterile environments. FRIWO's additional features make its products easier to use, day in, day out. Article no. FW Type Voltage Current Power Ripple voltage* Efficiency (typ) No-Load Power * Ripple measured with 20MHz Bandwidth Oscilloscope and 0.1uF/50V ceramic capacitor and 10uF/50V aluminum electrolytic capacitor across the output terminal.

 Open the catalog to page 1
FOX18-XM-2

18 W MEDICAL POWER SUPPLY Input Specifications Input voltage 100-240 V ±10% Frequency 50-60 Hz Input current 400-200 mA Inrush current (@240V) < 50 A General Specifications Operating temperature 0..+50°C Operating humidity 10..90 % Operating altitude 3000 m Storage temperature -40°C..+70°C Storage humidity 10..95 % Non-operating altitude 3000 m Atmospheric pressure 70-106kPa Output voltage tolerance ±5 % Line regulation ±1 % Load regulation ±4 % Turn-on delay ≤2 s Hold-up time >20 ms (120 V) >50 ms (230 V) MTBF calculation 200.000 h acc. Mil217F (based on calculations at 120Vac/60Hz & 230Vac/50Hz,...

 Open the catalog to page 2
FOX18-XM-3

Drop Test Mechanical Drawings Standard IEC 60068-2-32 Surface Steel Drop height 1 m Numbers of Drops (total) 3 Vibration Standard IEC 60068-2-6 50 (1.97) Frequency range 10 – 500 Hz Cross-over-frequency 58 – 62 Hz Displacement amplitude 0.2 mm Peak acceleration 3 g Number of cycles 10 per axis Alle Abmessungen in Millimeter (Inch), Abweichung ± 0,5 (0.02) All Dimensions in Millimeter (Inch), Deviation ± 0,5 (0.02) Output Characteristics V + 5% Vout - 5%

 Open the catalog to page 3
FOX18-XM-4

Primary Adapters FRIWO Geratebau GmbH Von-Liebig-StraBe 11 ■ 48346 Ostbevern ■ Germany Tel. +49 2532 81-0 • www.friwo.com Mechanical Drawings Flyer FOX18-XM, 11/2016. The content of this brochure is subject to technical modification.

 Open the catalog to page 4

All FRIWO Gerätebau GmbH catalogs and technical brochures

  1. FOX30-X

    4  Pages

Archived catalogs

  1. FOX12-XM-USB

    4  Pages

  2. FOX6-XM

    4  Pages

  3. FOX12-XM

    4  Pages

  4. FOX6-XM-USB

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.