Overview: The FOX12-XM-USB is an 11 W medical power supply designed by FRIWO, featuring a switchmode power supply with an interchangeable adapter system. It complies with IEC 60601-1 standards, ensuring safety and efficiency for medical applications.
Key Specifications:
  • High efficiency with low no-load power consumption of 50mW.
  • Meets DoE efficiency level VI and CoC version 5 Tier 2 standards.
  • Ultra-low leakage current of ≤10µA, ensuring patient safety.
  • Wide operating temperature range from 0 to 45°C.
  • Isolation of 2x MOPP for enhanced safety.
  • Available with fixed plugs as FOX12-FM-USB.
Performance and Safety:
  • Designed for sensitive applications and harsh conditions, maintaining functionality despite physical impacts.
  • Equipped with a 2 x MOPP safety system and double-sealed housings.
  • Operates at a universal input voltage range of 90Vac to 264Vac.
  • Complies with US DOE and European ErP efficiency standards.
Model Selection and Output Specifications:
  • Article no. 1898350: 5 V, 2200 mA, 11 W, Ripple voltage 150 mV pp, Efficiency ≥ 79.5%, No-Load Power <50mW.
Input Specifications:
  • Input voltage: 100-240 V ±10%, Frequency: 50-60 Hz, Input current: 300-150 mA, Inrush current (@240V): < 35 A.
General Specifications:
  • Operating temperature: 0 to +45°C, Humidity: 10-95%, Altitude: 3000 m.
  • Storage temperature: -40°C to +70°C, Humidity: 10-95%.
  • MTBF: 200,000 hours based on Mil217F calculations.
  • Housing material: PC-ABS, Dimensions: 77.5x31.5x41 mm, Weight: 65 g.
Safety and Compliance:
  • Safety standards: IEC/EN/UL60601-1 3.1 Edition, Approvals in Europe, USA, Australia, Japan.
  • Protection class II, Isolation: Input-Output 2xMOPP, Input-Ground 1xMOPP.
  • Flame class: UL 94 V0, IP code: IP40.
  • Overload, overvoltage, and short circuit protection included.
EMC Compliance:
  • Conducted and radiated emissions: EN55022, EN55011, FCC15, Class B, EN60601-1-2, IEC60601-1-2 4th Edition.
  • Immunity: EN55024, EN60601-1-2, IEC60601-1-2 4th Edition.
  • Harmonics: EN61000-3-2 Class A, Flicker noise: EN61000-3-3.
Mechanical and Environmental Testing:
  • Drop Test: IEC 60068-2-32, Drop height: 1 m, Number of drops: 3.
  • Vibration: IEC 60068-2-6, Frequency range: 10–500 Hz, Peak acceleration: 3 g.
See more

Catalog excerpts

FOX12-XM-USB-1

Switchmode power supply MEDICAL with interchangeable adapter system. All products conform to IEC 60601-1. Characteristics High efficiency Low no load power consumption 50mW Meets DoE efficiency level VI Meets CoC version 5 Tier 2 ready Ultra-low leakage current <10pA IEC 60601-1-2, 4th edition ready Wide operating temperature range 0 - 45°C Isolation 2x MOPP Double housing safety: snapped & welded Also available as version with fixed plugs: FOX12-FM-USB FOX12-XM-USB <? 11 W MEDICAL POWER SUPPLY FRIWO's medical products are designed for the most sensitive applications and for use under the harshest conditions. They must continue to work perfectly despite falls in the emergency room while protecting patients by offering the lowest possible levels of leakage current. FRIWO develops and manufactures power supply units you can trust. In order to ensure maximum patient safety, FRIWO medical technology devices are equipped with its 2 x MOPP (Means of Patient Protection) safety system and feature a minimal leakage current of < 10 pA as well as double-sealed housings. The power supply will operate at universal input voltage range of 90Vac to 264Vac in a wide temperature range of 0-45°C without derating. In addition to patient safety, efficiency plays a major role: with low standby losses of 50mW and high efficiency rates the medical FOX line complies with the demanding US efficiency standard DOE (Department of Energy) and its European counterpart ErP (Energy related Products). Article no. FW Type Voltage Current Power Ripple voltage* Efficiency (typ) No-Load Power * Ripple measured with 20MHz Bandwidth Oscilloscope and 0.1uF/50V ceramic capacitor and 10uF/50V aluminum electrolytic capacitor across the output terminal.

 Open the catalog to page 1
FOX12-XM-USB-2

11 W MEDICAL POWER SUPPLY Input Specifications Input voltage 100-240 V ±10% Frequency 50-60 Hz Input current 300-150 mA Inrush current (@240V) < 35 A General Specifications Operating temperature 0..+45°C Operating humidity 10..95 % Operating altitude 3000 m Storage temperature -40°C..+70°C Storage humidity 10..95 % Non-operating altitude 3000 m Atmospheric pressure 70-106kPa Output voltage tolerance ±5 % Line regulation ±1 % Load regulation ±4 % Turn-on delay ≤2 s Hold-up time >5 ms (120 V) >50 ms (230 V) MTBF calculation 200.000 h acc. Mil217F (based on calculations at 120Vac/60Hz & 230Vac/50Hz,...

 Open the catalog to page 2
FOX12-XM-USB-3

Drop Test Mechanical Drawings Standard IEC 60068-2-32 Surface Steel Drop height 1 m Numbers of Drops (total) 3 Vibration Standard IEC 60068-2-6 31,5 Frequency range 10 – 500 Hz (1.24) Cross-over-frequency 58 – 62 Hz Displacement amplitude 0.2 mm Peak acceleration 3 g Number of cycles 10 per axis Alle Abmessungen in Millimeter (Inch), Abweichung ± 0,5 (0.02) All Dimensions in Millimeter (Inch), deviation ± 0,5 (0.02) Output Characteristics

 Open the catalog to page 3
FOX12-XM-USB-4

FRIWO Geratebau GmbH Von-Liebig-StraBe 11 ■ 48346 Ostbevern ■ Germany Tel. +49 2532 81-0 • www.friwo.com Primary Adapters Mechanical Drawings Flyer FOX12-XM-USB, 11/2016. The content of this brochure is subject to technical modification.

 Open the catalog to page 4

All FRIWO Gerätebau GmbH catalogs and technical brochures

  1. FOX30-X

    4  Pages

Archived catalogs

  1. FOX6-XM

    4  Pages

  2. FOX12-XM

    4  Pages

  3. FOX18-XM

    4  Pages

  4. FOX6-XM-USB

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.