Overview: The Smart O2
gas analyzer is designed for continuous and precise oxygen measurement in industrial processes. It is versatile and can be integrated into most processes, except those involving silicones, fluorocarbons, sulfur oxides, hydrogen, or explosive gas mixtures.
Advantages: The Smart O2 offers direct measured value display, high efficiency, continuous operation, long-term stability, low calibration effort, low maintenance costs, and a compact design.
Fields of Application: The analyzer is suitable for use in the chemistry industry, air quality monitoring, food industry, biogas, pharmaceutical industry, recycling, and environmental applications.
Technical Specifications:- Model: Smart O2
- Measuring Technology: Zirconium oxide sensor
- Measuring Range: 0–5% vol / 0–50,000 ppm, 0–25% vol / 0–250,000 ppm, 5–95% vol / 50,000–950,000 ppm
- Error of Measurement: +/- 0.5% per year from upper range value
- Pressure Compatibility: 300 mbar – 1.5 bar
- Gas Flow: High flow: 4.0 l/min
- Climatic Conditions: Ambient temperature 5°C – 45°C / 41°F - 113°F, relative humidity ~90%, dew free
- Supply Voltage: ~230 V / 50 Hz
- Case Dimensions: 310 mm x 133 mm x 221 mm, 3 kg
- Display: 128 x 64 touch panel
- Operation: Resistive touch panel
- Electrical Inputs/Outputs: 4 digital inputs 24 V AC/DC, 1 digital output “ready” (relay) max. 48 V / 500 mA, 1 digital output “fault” (relay) max. 48 V / 500 mA, 4 digital outputs (relay) max 48 V / 500 mA, optional 4 analog outputs (galv. separated) 4-20 mA, individually programmable
- Interfaces: Standard RS 232, optional: Analog 4-20 mA, Profibus DP, Ethernet Modbus TCP, Ethernet IP, Profinet
- Operational Availability: Heat-up 120 seconds (at 20°C)
Contact Information: Fresenius Umwelttechnik GmbH, Doncaster-Platz 5, 45699 Herten, Tel.: +49 (0) 2366 / 93 96 1 -10, Fax: +49 (0) 2366 / 93 96 1 -16, Email:
[email protected], Web: www.fresenius-ut.com