Application
The Smart O2 gas analyzer is designed for continuous and precise oxygen measurement in industrial processes. It can be integrated into most plant processes, except those involving gas mixtures with silicone gases, fluorocarbons, sulfur oxides, hydrogen, or explosive gases. It offers reliable and economical online oxygen measurement.

Advantages
  • Direct measured value display
  • High cost-effectiveness
  • Utilizes a zirconium oxide sensor
  • Continuous and long-term stable measurements
  • Low calibration and service costs
  • Compact housing

Areas of Application
  • Chemical industry
  • Air quality monitoring
  • Food industry
  • Biogas plants
  • Pharmaceutical industry
  • Recycling industry
  • Environmental monitoring

Technical Specifications
  • Sample Gas Inputs: 1 (standard), expandable up to 4
  • Sample Gas Outputs: 1
  • Measurement Technology: Zirconium oxide ion conductor principle
  • Gas Components: Oxygen (O2)
  • Measurement Range: 0-25 Vol.-%, 0-50 Vol.-%, 0-100 Vol.-%
  • Measurement Uncertainty: +/- 0.5% per year of the measuring range end value
  • Detection Limit/Resolution: 0.1 Vol.-%
  • Measurement Interval: 1 second
  • Pressure Tolerance: Standard 0.7 to 1.1 bar
  • Internal Sample Gas Pump/Flow Rate: Yes, Standard 4 l/min
  • Internal Filter: PE filter 5 μm
  • Climatic Conditions: Ambient temperature 5° - 45° C, relative humidity ~ 90%, dew-free
  • Operating Voltage: Standard 230 V AC / 50 Hz, 0.9 A / 207 VA
  • Housing Dimensions: B - 310 mm x H - 133 mm x T - 221 mm
  • Weight: 4.7 kg
  • Display: Touch display 128 x 64 pixels black/white, 4 display LEDs

Electrical and Data Interfaces
  • Electrical Inputs: 4 digital inputs 24 V AC/DC
  • Electrical Outputs: 1 digital output “ready” (relay), 1 digital output “fault” (relay), 4 digital outputs (relay), expandable up to 8
  • Data Interfaces: Standard RS232, optional Profibus DP, Ethernet Modbus TCP, Ethernet IP, Profinet, up to 8 analog outputs (galvanically isolated) 4-20 mA
  • Operational Readiness: Measurement readiness reached after 60 seconds, full specifications reached after 30 minutes

Contact Information
Fresenius Umwelttechnik GmbH, Doncaster-Platz 5, 45699 Herten, +49 2366/ 93 96 1 – 10, [email protected], www.fresenius-ut.com
See more

Catalog excerpts

Smart O2 Zirconium Oxide Data Sheet-1

ADVANTAGES AREAS OF APPLICATION The Smart O2 is used for continuous and precise oxygen measurement in industrial processes. The analyzer can be integrated into almost any plant process. Exceptions are Gas mixtures containing silicone gases, fluorocarbons, sulphur oxides or hydrogen, explosive gas mixtures Direct measured value display High cost-effectiveness Zirconium oxide sensor Continuous Long-term stable Low calibration effort Low service costs Compact housing • Chemical industry • Food industry • Biogas plants • Pharmaceutical industry • Recycling industry Fresenius Umwelttechnik GmbH Doncaster-Platz 5 45699 Herten % vww.fresenius-ut.com #greentech Made in Germany

 Open the catalog to page 1
Smart O2 Zirconium Oxide Data Sheet-2

Technical Data GAS ANALYZER SMART O2 Model Smart O2 Sample Gas Inputs 1 (standard), expandable up to 4 Measurement Technology ZrO2: Zirconium oxide ion conductor principle Measurement Uncertainty + / - 0.5% per year of the measuring range end value Detection limit / Resolution 0.1 Vol.-% Measured Value Correction / Pressure Tolerance Standard 0.7 to 1.1 bar (process gas pressure) Internal Sample Gas Pump/ Yes Flow Rate Standard 4 l/min Internal Gas Routing PTFE Internal Filter PE filter 5 pm Connections for Sample Gas Line CK screw connections 6/4 Alternative: cutting ring fittings 6/4 Sample...

 Open the catalog to page 2
Smart O2 Zirconium Oxide Data Sheet-3

Technical Data GAS ANALYZER SMART O2 Electrical Inputs 4 digital inputs 24 V AC/DC Electrical Outputs 1 digital output “ready” (relay) max 48 V / 500 mA 1 digital output “fault” (relay) max 48 V / 500 mA 4 digital outputs (relay) max 48 V / 500 mA, expandable up to 8 Data Interfaces Standard: RS232 Optional: Profibus DP, Ethernet Modbus TCP, Ethernet IP, Profinet, up to 8 analog outputs (galvanically isolated) 4-20 mA, freely configurable via display menu Operational Readiness Measurement readiness reached after 60 s Full specifications reached after 30 min #greentech Made in Germany fresenius...

 Open the catalog to page 3

All Fresenius Umwelttechnik GmbH catalogs and technical brochures

  1. Substance List

    10  Pages

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About DirectIndustry
DirectIndustry is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. DirectIndustry is an entity of VirtualExpo Group.