1. Catalogs
  2. Fresenius Umwelttechnik GmbH
  3. Smart Control 300 Data Sheet

Smart Control 300 Data Sheet

Smart Control 300 Data Sheet

Smart Control 300 Data Sheet

Product catalog summary
Application
The Smart Control 300 is a stationary NDIR measuring device designed for monitoring threshold concentrations of solvent-containing gas mixtures. It is cost-effective and specifically tailored for high concentration ranges.

Advantages
  • Equipped with display and LED indicators
  • Supports up to two measuring points
  • Easy integration into systems
  • High long-term stability
  • Capable of measurements in vacuum conditions

Areas of Application
  • Solvent monitoring
  • Surface treatment
  • Various industrial applications

Technical Data
  • Model: Smart Control 300
  • Sample Gas Inputs: 1 standard, expandable to 2
  • Measurement Technology: NDIR dual-pressure reference method, with a heated cuvette at 50°C
  • Gas Components: Perchloroethene, hydrocarbon solvents, refrigerants, others on request
  • Measurement Range: Dependent on substance
  • Measurement Uncertainty: Less than 5% of the measured value
  • Detection Limit/Resolution: Dependent on substance
  • Measurement Interval: 20 seconds
  • Measured Value Correction: Includes pressure compensation and cross-sensitivity correction
  • Pressure Tolerance: Standard 0.7 to 1.1 bar
  • Internal Sample Gas Pump/Flow Rate: Standard 4 l/min
  • Internal Gas Routing: PTFE
  • Internal Filter: Stainless steel sintered filter
  • Connections for Sample Gas Line: CK screw connections 6/4, alternative cutting ring fittings 6/4
  • Climatic Conditions: Ambient temperature 5° - 45°C, relative humidity ~90%, dew-free
  • Operating Voltage: Standard 230 V AC / 50 Hz, 0.9 A / 207 VA, alternative 115 V AC / 60 Hz or 230 V / 60 Hz
  • Housing: Dimensions W - 310 mm x H - 133 mm x D - 221 mm, Protection Class IP 30
  • Weight: 4.7 kg
  • Display: Touch display 128 x 64 pixels black/white, 4 display LEDs
  • Operation: Menu navigation via 2.5” resistive touch display

Electrical Inputs and Outputs
  • 4 digital inputs 24V AC/DC
  • 1 digital output “ready” (relay) max 48 V / 500 mA
  • 1 digital output “fault” (relay) max 48 V / 500 mA
  • 4 digital outputs (relay) max 48 V / 500 mA, expandable up to 8

Data Interfaces
  • Standard: RS232
  • Optional: Profibus DP, Ethernet Modbus TCP, Ethernet IP, Profinet
  • Up to 8 analog outputs (galvanically isolated) 4-20 mA, configurable via display menu

Operational Readiness
  • Readiness to measure is substance-dependent
  • Full specifications reached after 30 minutes

Contact Information
Fresenius Umwelttechnik GmbH, Doncaster-Platz 5, 45699 Herten, +49 2366/ 93 96 1 – 10, [email protected], www.fresenius-ut.com
See more

Catalog excerpts

Smart Control 300 Data Sheet-1

The SmartControl 300 is a cost-effective stationary NDIR measuring device. Its main area of application is the monitoring of threshold concentrations in various gas mixtures. It is designed exclusively for measuring high concentration ranges. The Smart Control 300 is used to monitor threshold concentrations of solvent-containing gas mixtures. Display and LED indicators Up to two measuring points Easy to integrate High long-term stability Measurements in vacuum possible • Solvent monitoring • Surface treatment • Industrial applications Fresenius Umwelttechnik GmbH Doncaster-Platz 5 45699 Herten % wwv.fresenius-ut.com #greentech Made in Germany

 Open the catalog to page 1
Smart Control 300 Data Sheet-2

Technical Data SMART CONTROL 300 Model Smart Control 300 Sample Gas Inputs 1 (standard), expandable to 2 Measurement Technology IR: NDIR dual-pressure reference method, cuvette heated to 50°C Gas Components NDIR 1-3 (perchloroethene, hydrocarbon solvents, refrigerants) Other substances on request Measurement Range Dependent on substance Measurement Uncertainty < 5 % of the measured value Detection limit / Resolution Dependent on substance Measured Value Correction Pressure compensation, cross-sensitivity correction Pressure Tolerance Standard 0.7 to 1.1 bar (process gas pressure) Internal Sample...

 Open the catalog to page 2
Smart Control 300 Data Sheet-3

Technical Data SMART CONTROL 300 Electrical Inputs 4 digital inputs 24V AC/DC Electrical Outputs 1 digital output “ready” (relay) max 48 V / 500 mA 1 digital output “fault” (relay) max 48 V / 500 mA 4 digital outputs (relay) max 48 V / 500 mA, expandable up to 8 Data Interfaces Standard: RS232 Optional: Profibus DP, Ethernet Modbus TCP, Ethernet IP, Profinet, up to 8 analog outputs (galvanically isolated) 4-20 mA, freely configurable via display menu Operational Readiness Readiness to measure substance-dependent Full specifications reached after 30 min #greentech Made in Germany fresenius Umwelttechnik...

 Open the catalog to page 3

All Fresenius Umwelttechnik GmbH catalogs and technical brochures

  1. Substance List

    10  Pages

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About DirectIndustry
DirectIndustry is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. DirectIndustry is an entity of VirtualExpo Group.