1. Catalogs
  2. Fresenius Umwelttechnik GmbH
  3. Gas Analyzer GA320 N2O CO2 CH4 Data Sheet

Gas Analyzer GA320 N2O CO2 CH4 Data Sheet

Gas Analyzer GA320 N2O CO2 CH4 Data Sheet

Gas Analyzer GA320 N2O CO2 CH4 Data Sheet

Product catalog summary
Overview: The Gas Analyzer GA 320 is designed for measuring up to six IR-active gases with high precision and low detection limits. It is suitable for applications in emissions monitoring, wastewater treatment plants, recycling, environmental monitoring, and inspections.
Key Features:
  • Direct display of measured values and user-friendly operation via a large touch panel with plain text menu navigation.
  • Precise measurement results even at low detection limits.
  • Integrated PC for fast data processing and analysis.
  • High long-term stability and low calibration effort due to patented NDIR measurement method with reference pressure measurement.
  • Open interface architecture for flexible data transmission, standard RS232.
  • Low service costs due to reduced maintenance requirements and high operational reliability.
Technical Specifications:
  • Model: Gas Analyzer 320
  • Sample Gas Inputs/Outputs: 1 standard, expandable up to 6 inputs; 1 output
  • Measurement Technology: NDIR dual-pressure reference technology
  • Gas Components: N2O, CO2, CH4
  • Measurement Range: N2O: 0-2000 ppmV, CO2: 0-10000 ppmV, CH4: 0-10000 ppmV
  • Measurement Uncertainty: N2O: ≤ 3%, CO2: ≤ 2%, CH4: ≤ 2%
  • Detection Limit/Resolution: N2O: 0.5 ppm, CO2: 1 ppmV, CH4: 0.1 ppmV
  • Measurement Interval: 20 seconds
  • Pressure Tolerance: 0.7 to 1.1 bar
  • Internal Sample Gas Pump/Flow Rate: 4 l/min
  • Climatic Conditions: Ambient temperature 5°C - 45°C, Relative humidity ~90%, non-condensing
  • Operating Voltage: Standard 230 V AC / 50 Hz, Alternative 115 V AC / 60 Hz or 230 V / 60 Hz
  • Housing: 19" rack-mounted unit, IP 20 protection class
  • Weight: 15.5 kg
Display and Operation:
  • Touch display with 800 x 480 pixels color, 7” resistive touch display for menu navigation.
  • Electrical Inputs: 4 digital inputs 24 V AC/DC.
  • Electrical Outputs: 1 digital output “ready” (relay) max 48 V / 500 mA, 4 digital outputs (relay) max 48 V / 500 mA, expandable up to 8.
  • Data Interfaces: Standard RS232, optional Profibus DP, Ethernet Modbus TCP, Ethernet IP, Profinet, up to 8 analog outputs (galvanically isolated) 4-20 mA.
  • Operational Readiness: Full specifications reached after 30 minutes.
Contact Information: Fresenius Umwelttechnik GmbH, Doncaster-Platz 5, 45699 Herten, +49 2366/ 93 96 1 – 10, [email protected], www.fresenius-ut.com
See more

Catalog excerpts

Gas Analyzer GA320 N2O CO2 CH4 Data Sheet-1

The Gas Analyzer 320 enables the measurement of up to 6 IR-active gases for which very low detection limits and a sophisticated cross-sensitivity compensation are required to the highest degree. • High long-term stability and low calibration effort thanks to patented NDIR measurement method with proven reference pressure measurement • Open interface architecture for flexible data transmission (standard RS232) • Low service costs due to reduced maintenance requirements and high operational reliability • Direct measured value display and user-friendly operation via touch panel with menu navigation in plain text on extra-large display • Precise measurement results even when measuring low detection limits • Integrated PC for fast data processing and analysis Fresenius Umwelttechnik GmbH Doncaster-Platz 5 45699 Herten % vww.fresenius-ut.com #greentech Made in Germany

 Open the catalog to page 1
Gas Analyzer GA320 N2O CO2 CH4 Data Sheet-2

Technical Data Sample Gas Inputs 1 (standard), expandable up to 6 Measurement Technology NDIR dual-pressure reference technology, self-referencing, zero-point correction with every measurement, long-term stable Gas Components 1 - 6 x NDIR Measurement Range N2O: 0 - 2000 ppmV Measurement Uncertainty N2O: < 3% of measured value CO2: < 2% of measured value CH4: < 2% of measured value Detection limit / Resolution N2O: 0.5 ppm Measurement Interval 20 seconds (based on one sample gas inlet) Measured Value Correction Pressure compensation Pressure Tolerance Standard: 0.7 to 1.1 bar (process gas pressure)...

 Open the catalog to page 2
Gas Analyzer GA320 N2O CO2 CH4 Data Sheet-3

Technical Data Display Touch display 800 x 480 pixels color Operation Menu navigation via 7” resistive touch display Electrical Inputs 4 digital inputs 24 V AC/DC Electrical Outputs 1 digital output “ready” (relay) max 48 V / 500 mA 4 digital outputs (relay) max 48 V / 500 mA, expandable up to 8 Data Interfaces Standard: RS232 Optional: Profibus DP, Ethernet Modbus TCP, Ethernet IP, Profinet, up to 8 analog outputs (galvanically isolated) 4-20 mA, freely configurable via display menu Operational Readiness Measurement readiness substance-dependent Full specifications reached after 30 min ^www.fresenius-ut.com...

 Open the catalog to page 3