1. Catalogs
  2. Fresenius Umwelttechnik GmbH
  3. Gas Analyzer GA220 Biomethane Data Sheet

Gas Analyzer GA220 Biomethane Data Sheet

Gas Analyzer GA220 Biomethane Data Sheet

Gas Analyzer GA220 Biomethane Data Sheet

Product catalog summary
Overview: The Gas Analyzer GA 220 is designed for precise and rapid online measurement of biomethane concentrations, specifically CH4 and CO2, without interference from other gas components. It is suitable for various applications including biomethane plants, CHP, power-to-gas plants, biogas plants, fuel cells, and gas heat pumps.
Advantages: The device offers high long-term stability and low calibration effort due to its patented NDIR measurement method. It features a user-friendly touch panel for direct value display and flexible data transmission through an open interface architecture (standard RS232). The analyzer also boasts low service costs and high operational reliability.
Technical Specifications:
  • Measurement Technology: NDIR dual-beam method with a heated cuvette at 50°C.
  • Gas Components: CH4 (90-100 Vol.-%), CO2 (0-10 Vol.-%).
  • Measurement Range and Uncertainty: CH4: 0.2% of full scale, CO2: 0.5% of full scale.
  • Detection Limit/Resolution: CH4: 0.1% of full scale, CO2: 0.05% of full scale.
  • Measurement Interval: 1 second.
  • Pressure Tolerance: Standard 0.7 to 1.1 bar.
  • Internal Sample Gas Pump/Flow Rate: Standard 4 l/min.
  • Climatic Conditions: Ambient temperature 5° - 45°C, relative humidity ~90%, dew-free.
  • Operating Voltage: Standard 230 V AC / 50 Hz, alternative options available.
  • Housing and Protection: 19” slide-in unit, IP 20 protection class.
  • Weight: 12 kg.
  • Display and Operation: Touch display 128 x 64 pixels, menu navigation via 2.5” resistive touch display.
  • Data Interfaces: Standard RS232, optional Profibus DP, Ethernet Modbus TCP, Ethernet IP, Profinet.
  • Operational Readiness: Measurement readiness in 5 minutes, full specifications in 30 minutes.
Contact Information: Fresenius Umwelttechnik GmbH, Doncaster-Platz 5, 45699 Herten, Phone: +49 2366/ 93 96 1 – 10, Email: [email protected], Website: www.fresenius-ut.com
See more

Catalog excerpts

Gas Analyzer GA220 Biomethane Data Sheet-1

The Gas Analyzer 220 for biomethane was developed for fast and simple online measurements of biogas concentrations where no interference from the various sample gas components needs to be compensated for but low detection limits are required. The device measures the biogas concentration with high accuracy. High long-term stability and low calibration effort thanks to patented NDIR measurement method with proven reference pressure measurement Direct measured value display and user-friendly operation via touch panel with menu navigation in plain Open interface architecture for flexible data transmission (standard RS232) Low service costs due to reduced maintenance requirements and high operational reliability • Biomethane plants • Power-to-gas plants • Biogas plants • Fuel cells Fresenius Umwelttechnik GmbH Doncaster-Platz 5 45699 Herten © wwv.fresenius-ut.com #greentech Made in Germany

 Open the catalog to page 1
Gas Analyzer GA220 Biomethane Data Sheet-2

Technical Data Model Gas Analyzer 220 - Biomethan Measurement Technology IR: NDIR dual-beam method, cuvette heated to 50°C Extended Measurement CH4: 0.2 % of full scale Uncertainty CO2: 0.5 % of full scale Detection limit / Resolution CH4: 0.1 % of full scale CO2: 0.05 % of full scale Measurement Interval 1 s Measured Value Correction Pressure compensation Pressure Tolerance Standard 0.7 to 1.1 bar (process gas pressure) Internal Sample Gas Pump/ Optional Flow Rate Standard 4 l/min Internal Gas Routing Standard PTFE Alternative stainless steel Internal Filter Stainless steel sintered filter Connections...

 Open the catalog to page 2
Gas Analyzer GA220 Biomethane Data Sheet-3

Technical Data Data Interfaces Standard: RS232 Optional: Profibus DP, Ethernet Modbus TCP, Ethernet IP, Profinet, up to 8 analog outputs (galvanically isolated) 4-20 mA, freely configurable via display menu Operational Readiness Measurement readiness reached after 5 min Full specifications reached after 30 min ^www.fresenius-ut.com #greentech Made in Germany

 Open the catalog to page 3