1. Catalogs
  2. Fresenius Umwelttechnik GmbH
  3. Gas Analyzer GA210 O2 Paramagnetic Data Sheet

Gas Analyzer GA210 O2 Paramagnetic Data Sheet

Gas Analyzer GA210 O2 Paramagnetic Data Sheet

Gas Analyzer GA210 O2 Paramagnetic Data Sheet

Product catalog summary
Overview: The Gas Analyzer 210 O2 Paramagnetic is designed for efficient online measurement of oxygen in process gases, utilizing a reliable paramagnetic measurement method. It offers high long-term stability, low calibration effort, and minimal cross-influences from other gas components.
Advantages: The analyzer features an open interface architecture for flexible data transmission, low service costs due to reduced maintenance requirements, and user-friendly operation via a touch panel with plain text menu navigation.
Technical Specifications:
  • Model: GA 210 O2 Paramagnetic
  • Sample Gas Inputs: 1 standard, expandable up to 6
  • Measurement Technology: Paramagnetic measuring cell
  • Gas Components: Oxygen (O2)
  • Measurement Range: 0-25 Vol.-%, 0-50 Vol.-%, 0-100 Vol.-%
  • Measurement Uncertainty: < 0.5 Vol.-% of the measured value per week
  • Detection Limit/Resolution: 0.1 Vol.-%
  • Measurement Interval: 1 second
  • Pressure Tolerance: Standard 0.7 to 1.1 bar
  • Flow Rate: Standard 4 l/min at sample gas input
  • Climatic Conditions: Ambient temperature 5° - 45° C, relative humidity ~ 90%, dew-free
  • Operating Voltage: Standard 230 V AC / 50 Hz, 0.9 A / 207 VA
  • Protection Class: IP 20
  • Weight: 13 kg
  • Display: Touch display 128 x 64 pixels black/white
Operational Readiness: Measurement readiness is reached after 5 minutes, with full specifications achieved after 30 minutes.
Contact Information: Fresenius Umwelttechnik GmbH, Doncaster-Platz 5, 45699 Herten, Phone: +49 2366/ 93 96 1 – 10, Email: [email protected], Website: www.fresenius-ut.com
See more

Catalog excerpts

Gas Analyzer GA210 O2 Paramagnetic Data Sheet-1

The Gas Analyzer 210 02 paramagnetic was developed for quick and easy online measurement of oxygen in process gases. • High long-term stability and low calibration effort due to reliable paramagnetic measurement method • Continuous measurements with short response time and minimized cross-influences from other gas components • Open interface architecture for flexible data transmission (standard RS232) • Low service costs due to reduced maintenance requirements and high operational reliability • Pharmaceutical industry • Direct measured value display and user-friendly operation via touch panel with menu navigation in plain text Fresenius Umwelttechnik GmbH Doncaster-Platz 5 45699 Herten % wwv.fresenius-ut.com #greentech Made in Germany % /freseniusumwelttechnik Q @fresenius_umwelttechnik /freseniusumwelttechnik @fresenius_UT

 Open the catalog to page 1
Gas Analyzer GA210 O2 Paramagnetic Data Sheet-2

Technical Data GAS ANALYZER GA 210 O2 PARAMAGNETIC Model Gas Analyzer 210 O2 paramagnetic Sample Gas Inputs 1 (standard), expandable up to 6 2 calibration inputs (nitrogen, oxygen) Measurement Technology Paramagnetic measuring cell Measurement Uncertainty < 0.5 Vol.-% of the measured value per week Detection limit / Resolution 0.1 Vol.-% Zero point deviation < ±0.1 Vol.-% O2/week (offset) Measured Value Correction None required Pressure Tolerance Standard 0.7 to 1.1 bar (process gas pressure) Internal Sample Gas Pump/ Flow Rate Internal: fixed pre-pressure, constant Yes Standard 4 l/min at sample...

 Open the catalog to page 2
Gas Analyzer GA210 O2 Paramagnetic Data Sheet-3

Technische Data GAS ANALYZER GA 210 O2 PARAMAGNETIC Operation Menu navigation via 2.5” resistive touch display Electrical Inputs 4 digital inputs 24 V AC/DC Electrical Outputs 1 digital output “ready” (relay) max 48 V / 500 mA 1 digital output “fault” (relay) max 48 V / 500 mA 4 digital outputs (relay) max 48 V / 500 mA, expandable up to 8 Data Interfaces Standard: RS232 Optional: Profibus DP, Ethernet Modbus TCP, Ethernet IP, Profinet, up to 8 analog outputs (galvanically isolated) 4-20 mA, freely configurable via display menu Operational Readiness Measurement readiness reached after 5 min Full...

 Open the catalog to page 3

All Fresenius Umwelttechnik GmbH catalogs and technical brochures

  1. Substance List

    10  Pages

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.