1. Catalogs
  2. Fresenius Umwelttechnik GmbH
  3. Gas Analyzer GA210 Data Sheet

Gas Analyzer GA210 Data Sheet

Gas Analyzer GA210 Data Sheet

Gas Analyzer GA210 Data Sheet

Product catalog summary
Overview: The Gas Analyzer GA 210 is designed for fast and simple online measurement of IR-active process gases with low detection limits and no interfering components. It is suitable for applications such as chemical cleaning, metal degreasing, parts cleaning, monitoring, and inspection.
Key Features:
  • High long-term stability and low calibration effort due to patented NDIR measurement method.
  • User-friendly operation with a touch panel and menu navigation.
  • Open interface architecture for flexible data transmission (standard RS232).
  • Low service costs due to reduced maintenance requirements and high operational reliability.
Technical Specifications:
  • Sample Gas Inputs/Outputs: 1 standard input, expandable up to 8; 1 output.
  • Measurement Technology: NDIR single-beam method, electrochemical measurement, and zirconium oxide ion conductor principle.
  • Gas Components: NDIR for perchloroethene, trichloroethene, dichloromethane, or hydrocarbon solvents; EC for O2, H2S, H2; ZrO2 for O2.
  • Measurement Range/Uncertainty/Detection Limit: Dependent on substance.
  • Measurement Interval: Continuous at 1 second, quasi-continuous at 20 seconds.
  • Pressure Tolerance: Standard 0.7 to 1.1 bar.
  • Internal Sample Gas Pump/Flow Rate: Standard 4 l/min.
  • Climatic Conditions: Ambient temperature 5° - 45°C, relative humidity ~90%, dew-free.
  • Operating Voltage: Standard 230 V AC / 50 Hz, alternative options available.
  • Housing: 19” slide-in unit, IP 20 protection class, weight 12 kg.
  • Display: Touch display 128 x 64 pixels black/white.
Operation and Interfaces:
  • Menu navigation via 2.5” resistive touch display.
  • 4 digital inputs 24 V AC/DC, 1 digital output “ready” and “fault” (relay) max 48 V / 500 mA.
  • Data Interfaces: Standard RS232, optional Profibus DP, Ethernet Modbus TCP, Ethernet IP, Profinet, up to 8 analog outputs (4-20 mA).
  • Measurement readiness is substance-dependent, with full specifications reached after 30 minutes.
Contact Information: Fresenius Umwelttechnik GmbH, Doncaster-Platz 5, 45699 Herten, +49 2366/ 93 96 1 – 10, [email protected], www.fresenius-ut.com
See more

Catalog excerpts

Gas Analyzer GA210 Data Sheet-1

The Gas Analyzer 210 was developed for fast and simple online measurement of IR-active process gases where no other interfering components are expected and low detection limits are required. High long-term stability and low calibration effort thanks to patented NDIR measurement method with proven reference pressure measurement Direct measured value display and user-friendly operation via touch panel with menu navigation in plain Open interface architecture for flexible data transmission (standard RS232) Low service costs due to reduced maintenance requirements and high operational reliability • Chemical cleaning • Metal degreasing • Parts cleaning Fresenius Umwelttechnik GmbH Doncaster-Platz 5 45699 Herten % wwv.fresenius-ut.com #greentech Made in Germany % /freseniusumwelttechnik Q @fresenius_umwelttechnik /freseniusumwelttechnik @fresenius_UT

 Open the catalog to page 1
Gas Analyzer GA210 Data Sheet-2

Technical Data GAS ANALYZER GA 210 Model Gas Analyzer 210 Sample Gas Inputs 1 (standard), expandable up to 8 Measurement Technology IR: NDIR single-beam method, cuvette heated to 50°C EC: electrochemical measurement method ZrO2: Zirconium oxide ion conductor principle Gas Components 1 x NDIR (Perchloroethene, trichloroethene, dichloromethane or hydrocarbon Other substances on request Measurement Range Dependent on substance Measurement Uncertainty Dependent on substance Detection limit / Resolution Dependent on substance Measurement Interval Dependent on substance Continuous: 1 s Quasi-continuous:...

 Open the catalog to page 2
Gas Analyzer GA210 Data Sheet-3

Technical Data GAS ANALYZER GA 210 Operation Menu navigation via 2.5” resistive touch display Electrical Inputs 4 digital inputs 24 V AC/DC Electrical Outputs 1 digital output “ready” (relay) max 48 V / 500 mA 1 digital output “fault” (relay) max 48 V / 500 mA 4 digital outputs (relay) max 48 V / 500 mA, expandable up to 8 Data Interfaces Standard: RS232 Optional: Profibus DP, Ethernet Modbus TCP, Ethernet IP, Profinet, up to 8 analog outputs (galvanically isolated) 4-20 mA, freely configurable via display menu Operational Readiness Measurement readiness substance-dependent Full specifications...

 Open the catalog to page 3

All Fresenius Umwelttechnik GmbH catalogs and technical brochures

  1. Substance List

    10  Pages

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About DirectIndustry
DirectIndustry is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. DirectIndustry is an entity of VirtualExpo Group.