The Gas Analyzer 150 was developed for quick and easy online measurement of solvent concentrations in the gas phase where low detection limits are not required. High long-term stability and low calibration effort thanks to patented NDIR measurement method with proven reference pressure measurement Direct measured value display and user-friendly operation via touch panel with menu navigation in plain Open interface architecture for flexible data transmission (standard RS232) Low service costs due to reduced maintenance requirements and high operational reliability • Chemical cleaning • Metal degreasing • Parts cleaning Fresenius Umwelttechnik GmbH Doncaster-Platz 5 45699 Herten % wwv.fresenius-ut.com #greentech Made in Germany
Open the catalog to page 1Technical Data GAS ANALYZER GA 150 Model Gas Analyzer 150 Sample Gas Inputs 1 (standard), expandable up to 8 Measurement Technology IR: NDIR single-beam method, cuvette heated to 50°C Gas Components 1 x NDIR (Perchloroethene, trichloroethene, dichloromethane or hydrocarbon Measurement Uncertainty < 3 % of the measured value Detection limit / Resolution 2 mg/m3 Measured Value Correction Pressure compensation Pressure Tolerance Standard 0.7 to 1.1 bar (process gas pressure) Internal Sample Gas Pump/ Yes Flow Rate Standard 4 l/min Internal Gas Routing Standard PTFE Alternative stainless steel Internal...