GA315 Formaldehyde Data Sheet-1

The Gas Analyzer 315 enables simple online measure ment of IR-active substances in gas streams with tem peratures of up to 300° C, e.g. formaldehyde concen trations in combustion processes. • Enables simple and continuous monitoring of formaldehyde • Direct measured value display and user-friendly operation via touch panel with menu navigation in plain text on extra-large display • Open interface architecture for flexible data transmission (standard RS232) • Low service costs due to reduced maintenance requirements and high operational reliability • High long-term stability and low calibration effort thanks to patented NDIR measuring method with proven reference pressure measurement • Integrated PC for fast data processing and analysis • Waste disposal companies • Power generation • Engine manufacturers We are your competent partner for gas measurement applications. Our analyzers are developed and produced exclusively in Germany. Fresenius Umwelttechnik GmbH Doncaster-Platz 5 45699 Herten

 Open the catalog to page 1
GA315 Formaldehyde Data Sheet-2

Technical Data GAS ANALYZER GA 315 FORMALDEHYDE Model Gas Analyzer 315 Formaldehyde Sample Gas Inputs 1, Heated at the front Measurement Technology IR: NDIR single-beam method with vacuum reference technology, cuvette Gas Components 1 x NDIR (Formaldehyd) Measurement Uncertainty < 5 % of the measured value. Detection limit / Resolution 4 mg/m3 Measured Value Correction Pressure compensation, humidity compensation, methane compensation Pressure Tolerance Standard 0.7 to 1.1 bar (process gas pressure) Internal Sample Gas Pump/ Yes Flow Rate Standard 4 l/min Internal Gas Routing Standard stainless...

 Open the catalog to page 2
GA315 Formaldehyde Data Sheet-3

Technical Data GAS ANALYZER GA 315 FORMALDEHYDE Electrical Inputs 4 digital inputs 24 V AC/DC Electrical Outputs 1 digital output “ready” (relay) max 48 V / 500 mA 4 digital outputs (relay) max 48 V / 500 mA, expandable up to 8 Data Interfaces Standard: RS232 Optional: Profibus DP, Ethernet Modbus TCP, Ethernet IP, Profinet, up to 8 analog outputs (galvanically isolated) 4-20 mA, freely configurable via display menu Operational Readiness Measurement readiness reached after 5 min Full specifications reached after 30 min #greentech Made in Germany fresenius Umwelttechnik

 Open the catalog to page 3

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.