1. Catalogs
  2. Fresenius Umwelttechnik GmbH
  3. BioBasic Bionix Biomethane Analyzer Data Sheet

BioBasic Bionix Biomethane Analyzer Data Sheet

BioBasic Bionix Biomethane Analyzer Data Sheet

BioBasic Bionix Biomethane Analyzer Data Sheet

Product catalog summary
Overview: The BioBasic Bionix is an advanced online gas analyzer designed for the continuous monitoring of biomethane quality at treatment and grid injection plants. It measures methane (CH4), carbon dioxide (CO2), oxygen (O2), and hydrogen sulfide (H2S) at one-second intervals, providing rapid feedback to ensure gas quality remains within network specifications.
Specifications:
  • Measurement Technology: Utilizes NDIR dual-beam method for CH4 and CO2, electrochemical method for O2, and NDUV with a reference detector for H2S.
  • Measurement Ranges: CH4: 90–100 vol.%, CO2: 0–10 vol.%, O2: 0–30 vol.%, H2S: 0–100 ppm.
  • Resolution: CH4: 0.1 vol.%, CO2: 0.01 vol.%, O2: 0.05 vol.%, H2S: 0.1 ppm.
  • Precision and Uncertainty: Varies by gas component, with expanded measurement uncertainty ranging from 0.2% to 0.5% of the full-scale value.
  • Measurement Interval: Continuous at 1-second intervals.
  • Temperature and Pressure Compensation: Operates effectively between 5–45 °C with pressure compensation from 0.7 to 1.1 bar.
Advantages:
  • Rapid response time with measurements every second.
  • High long-term stability and sensitive, wear-free H2S measurement.
  • Automatic calibration with error monitoring.
  • Plug & Play setup with a color touch display and plain-text fault messages.
  • Data logging capabilities and flexible data transmission options.
Applications: Suitable for biomethane feed-in systems, gas network transfer stations, biogas upgrading, and Power-to-Gas/SNG applications.
Technical Data:
  • Gas Outputs: Includes measurement gas outlet and bypass (bypass not applicable on pump-less version).
  • Calibration: Automatic with adjustable intervals, approximately 4 minutes per cycle.
  • Climatic Conditions: Operates best in ambient temperatures of 20–30 °C with relative humidity around 90%.
  • Operating Voltage: Standard 230 V AC / 50 Hz, with alternatives available.
  • Housing and Protection: Wall-mounted enclosure with IP 54 protection class.
  • Data Interfaces: Includes standard 4–20 mA analogue output, RS232, with optional Profibus DP, Profinet, Ethernet IP, Modbus TCP.
  • Safety: ATEX-certified flame arrester and UL/CSA listed.
Contact Information: Fresenius Umwelttechnik GmbH, Doncaster-Platz 5, 45699 Herten, Germany. Phone: +49 2366/ 93 96 1 – 10, Email: [email protected], Website: www.fresenius-ut.com
See more

Catalog excerpts

BioBasic Bionix Biomethane Analyzer Data Sheet-1

The BioBasic Bionix is a combined online gas analyser for the rapid, continuous monitoring of biomethane quality at treatment and grid injection plants. It measures CH4, CO2, O2 and H2S at one-second intervals — serving as a rapid, process-close complement to the process gas chromatograph. APPLICATION The BioBasic Bionix monitors the quality of the product gas following biogas processing — directly at the point of transfer to the gas network. Whilst the custody-transfer process gas chromatograph only provides a reading every few minutes, the BioBasic takes measurements every second and immediately highlights any fluctuations in quality — this allows the processing system to be quickly readjusted and ensures that the gas fed into the network remains within the network specifications. • Measurement in a matter of seconds (1 s) — rapid process response • NDIR dual-beam method (CH4/ CO2), high long-term stability • H2S measurement using NDUV with a reference detector — sensitive, wear-free • Automatic, error-monitored calibration • Plug & Play, colour touch display, faults in plain text • Data logging to USB stick, GasAnalyzer Connect app via Bluetooth • Flexible data transmission (4–20 mA, RS232, fieldbus) • Optional pump-free version • Biomethane feed-in systems • Gas network transfer stations • Biogas upgrading • Power-to-Gas / SNG Fresenius Umwelttechnik GmbH Doncaster-Platz 5 [email protected] www.fresenius-ut.com

 Open the catalog to page 1
BioBasic Bionix Biomethane Analyzer Data Sheet-2

Technical Data Model BioBasic Bionix Measurement technology CH4 / CO2: NDIR dual-beam method H2S: NDUV with reference detector Precision CH4: 0.1% of the full-scale value of the measuring range CO2: 0.05% of the full-scale value of the measuring range O2: 0.1% of the full-scale value of the measuring range H2S: 0.2% of the full-scale value of the measuring range Expanded measurement uncertainty CH4: 0.2% of the full-scale value of the measuring range CO2: 0.5% of the full-scale value of the measuring range O2: 0.3% of the full-scale value of the measuring range H2S: 0.5% of the full-scale value...

 Open the catalog to page 2
BioBasic Bionix Biomethane Analyzer Data Sheet-3

Technical Data Measurement gas outputs 1* measurement gas outlet, ix bypass (bypass not applicable on the pump less version) Connections for the sample gas line CK screw connections 8/6 Internal gas flow PTFE Calibration Automatic (test gas and zero gas), adjustable time interval; duration approx. 4 mins per calibration cycle (no readings available during this time) error monitoring (limits for calibration factor, standard deviation at zero point, offset) Climatic conditions Ambient temperature 20-30 °C relative humidity ~ 90 %, non-condensing air-conditioned indoor installation only (tight measurement...

 Open the catalog to page 3

All Fresenius Umwelttechnik GmbH catalogs and technical brochures

  1. Substance List

    10  Pages

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About DirectIndustry
DirectIndustry is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. DirectIndustry is an entity of VirtualExpo Group.