Operating Specifications
The 7F cold cathode sensor operates within a range of 1 x 10-11 to 1 x 10-2 Torr. It is controller-dependent for analog output and supply voltage, and can be mounted in any orientation. The sensor is calibrated using dry air or nitrogen and can withstand overpressure up to 150 PSI. It operates effectively between 0° to +50° C and can be stored between -20° to +70° C. The bakeout temperature is 400° C without magnets and 150° C with magnets. The response time is ≤ 10 ms, with an accuracy of ±0.2 decade from 1 x 10-9 to 1 x 10-2 Torr.
Physical Characteristics
The sensor uses SHV5 electrical connections, weighs 572 g (1.26 lbs), and is made of stainless steel and ceramic materials exposed to vacuum.
Benefits
The 7F sensor is competitively priced, compact, resistant to gas inrushes, and supports high bakeout temperatures. It is interchangeable with models 7E, 7FC, and 7FCS, and is backed by excellent customer support. It is designed and manufactured in the USA.
Description
The 7F sensor measures absolute pressure by ionizing gas in a magnetic field, offering resistance to gas inrushes and high bakeout temperatures. It is compatible with Televac controllers like MM200 and MC300, providing a full range vacuum measurement solution when used with rough vacuum sensors.
Ratings and Compliance
The sensor is RoHS compliant.
Ordering Information
The 7F Cold Cathode sensor is available with various cable options and configurations, including standard, high temperature, triaxial, and radiation-resistant cables. Controllers like MM200 and MC300 are available for different configurations.
Related Products
Other products include the 7E sensor, MP7FR Active Gauge, and MM200 Vacuum Sensor Controller, each offering unique features for vacuum measurement.
Contact Information
Televac, The Fredericks Company, 2400 Philmont Avenue, Huntingdon Valley, PA 19006. Tel: +1 215 947 2500, Fax: +1 215 947 7464, Email: [email protected], Web: www.televac.com
See more

Catalog excerpts

Televac 7F Double Inverted Magnetron Cold Cathode Sensor-1

Supply Voltage Mounting Orientation Calibration Medium Overpressure Analog Output Resolution Operating Temperature Storage Temperature Bakeout Temperature Response Time Accuracy 1 x 10-9 to 1 x 10-2 Torr Controller-dependent Any Dry air or nitrogen 150 PSI Controller-dependent 0° to +50° C -20° to +70° C 400° C (without magnets) 150° C (with magnets) < 10 ms 7F Double Inverted Magnetron Cold Cathode Sensor Part Number: 2-2144-052Operating Specifications1 Operating Range 1 x 10" to 1 x 10" Torr Analog Output Controller-dependent The 7F cold cathode sensor has a measurement range from 1 x 10-11 up to 1 x 10-2 Torr. It indirectly measures absolute pressure by ionizing gas in the presence of a magnetic field. Unlike hot filament sensors, the 7F is resistant to inrushes of gas. The design of this sensor allows for an extended range of measurement and high bakeout temperatures up to 400° C. The 7F can be operated by several Televac controllers, including the MM200 or MC300 rack mountable controllers and the MP7FR. The MM200 and MC300 controllers automatically operate the cold cathode to provide a full range vacuum measurement solution when the 7F is used in conjunction with rough vacuum sensors. Dimensions See dimensional drawings Stainless Steel Ceramic • Competitively priced • Compact design • Minimal stray magnetic field • Resistant to inrushes of gas • High bakeout temperature • Interchangeable with 7E, 7FC, 7FCS • Excellent customer support • Designed and manufactured in the United States of America • RoHS compliant 1 See the Televac website at www.televac.com for a list of definitions for terms used in the operating specifications. All test data was acquired at 23° C.

 Open the catalog to page 1
Televac 7F Double Inverted Magnetron Cold Cathode Sensor-2

7F Double Inverted Magnetron Cold Cathode Sensor Part Number: 2-2144-052 Ordering Information Dimensional Drawings 7F Cold Cathode 2.75" CF (DN40 ISO-CF-F) Standard Cables High Temperature Cables (200 C) Triaxial Cables (MM200 only) Radiation Resistant Cables Cable Length 10 ft (3 m) 20 ft (6.1 m) 35 ft (10.6 m) 50 ft (15.2 m)_ Related Products 7E Double Inverted Magnetron Cold Cathode Sensor - This sensor has a measurement range from 1 x 10-8 up to 1 x 10-2 Torr. It indirectly measures absolute pressure by ionizing gas in the presence of a magnetic field. Unlike hot filament sensors, the 7E...

 Open the catalog to page 2

All The Fredericks Company catalogs and technical brochures

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.