Operating Specifications:
The sensor has a maximum operating range of 30 arc minutes and a linear range of ±15 arc minutes. It features a null voltage of ≤0.030 volts and a maximum null current of 5.0 mA. The nominal null impedance is 3500 Ohms at 25°C, measured between the left and right electrodes. The sensor offers a null repeatability of less than 0.5 arc seconds and a resolution of less than 0.05 arc seconds. It operates symmetrically with a typical symmetry of ≤20%. The operating temperature range is from -20°C to +100°C, while the storage temperature range is from -50°C to +100°C. The time constant is less than 500 milliseconds, and the materials used are non-magnetic. The temperature coefficient at null is 0.05%/°C when properly mounted.
Physical Dimensions:
The vial length is 1.06 inches (26.9 mm), and the diameter is 0.320 inches (8.13 mm). The lead length is 15.0 inches (381 mm).
Applications:
The sensor is suitable for geophysical monitoring, construction laser instruments and transits, aircraft avionics, machine tool leveling, and medical positioning and monitoring.
Performance Characteristics:
The sensor is noted for its ultra-high accuracy, excellent null repeatability, and linear output. It maintains long-term stability over its angle and temperature range. The hermetic glass-to-metal construction and platinum electrodes ensure a long operating life and stable characteristics.
Test Circuitry:
Tests were conducted using an AC signal of 400 Hz to excite the electrodes, with output readings taken between the center electrode and the center of balanced resistors R1 and R2. The tests were performed at +25°C. It is crucial to ensure that all test and operating circuits are free of direct current to prevent damage or instability.
Figures:
Figure 1 illustrates the output sensitivity as a function of tilt angle, while Figure 2 shows impedance as a function of temperature.
See more

Catalog excerpts

Glass Narrow Angle 0737-0604-99-1

DATA SHEET Operating Specifications Description The 0737-0604-99 Sensor has been used successfully in applications that demand ultra high accuracy, excellent null repeatability, and a linear output. Long term stability over its angle and temperature range is a distinctive characteristic of this sensor. The hermetic glass to metal construction and platinum (platinized) electrodes guarantee a long operating life and stable operating characteristics. Angle Range Resolution Null Repeat Applications Include Geophysical Monitoring Construction Laser Instruments and Transits Aircraft Avionics Machine Tool Leveling Medical Positioning and Monitoring Physical Dimensions Vial length Vial Diameter Lead Length Impedance of the electrode may be changed to limit null current. 2 Viscosity of the electrolyte may be modified to meet individual requirements for time constant or vibration. Sensor Test Circuitry Tests were conducted by exciting the left and right electrodes with an AC signal of 400 Hz and an RMS voltage to produce the maximum current at null as per operating specifications. Output readings are taken between the center electrode and the center of the balanced resistors R1 and R2. Tests were conducted at a temperature of +25 C. See sensor test circuitry in figure 3. Output curve is shown in figure 1. OUTPUT SENSITIVITY Figure 1 100 80 PERCENT OUTPUT 30’ Operating Range (max.) Linear Range ± 15’ Null Voltage ≤0.030 Volts Null Current(max.) 5.0 mA (continuous) Null Impedance (nom)1 3500 Ohms(25°C) (measured left to right electrode) see fig 2 Null Repeatability < 0.5 arc seconds Resolution < 0.05 arc seconds Symmetry (typ) ≤20% -20 C to +100 C Operating Temperature -50 C to +100 C Storage Temperature Time Constant (1)2 < 500 msec Materials non-magnetic Temperature Coefficient 0.05%/°C max. at null when properly mounted Descrition of Test Values IMPEDANCE / TEMPERATURE Figure 2 AC input voltage = Null Current (max) times Null Impedance (nom) R1 =R2 = ½ Null Impedance (nom) Caution!-Ensure that all test and operating circuits are entirely free of direct current. Direct current will cause level damage and/or instability. IMPEDANCE (Ohms) Eout = Angle of tilt from null (Direction of tilt determined by phase of Eout) The Fredericks Company l 2400 Philmont Ave ● HuntingdonValley, PA 19006-0067 Tel: 215.947.2500 l Fax: 215.947.7464 l Email: [email protected] l www.frederickscom.com 0737-0604-99data sheet Revision: A

 Open the catalog to page 1

All The Fredericks Company catalogs and technical brochures

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.