LGA-3500

LGA-3500

LGA-3500

Product catalog summary
Overview: The LGA-3500 is a laser gas analyzer developed by Focused Photonics Inc. using DLAS technology for detecting industrial process gases. It is suitable for high temperature, pressure, and corrosive environments, and is applicable in industries like chemical, cement, and environmental protection.
Specifications:
  • Linear error: ≤±1% F.S.
  • Range drifting: ≤±1% F.S. /month
  • Repetitiveness error: ≤±1% F.S.
  • Sample gas flow: 1L/min
  • Response time: T90<21s (long optical length), T90<11s (short optical length)
  • Analog output: 2 channels of (4~20)mA current with isolation, max load 500Ω
  • Relay output: 3 channels relay of 24V DC/1A
  • Digital communication: RS485/RS232
  • Power: (100~240)V AC/(48-63)Hz, consumption <12W
  • Electrical safety: IEC 1010-1
  • EMC: IEC 61000-4-2, IEC 61000-4-4, IEC 61000-4-5
  • Enclosure rating: IP66
  • Anti-explosion class: Ex d px IIC T5 Gb
Gas Input Conditions:
  • Pressure: (0.5~3) bar (absolute pressure)
  • Temperature: (-30~+100)°C
  • Sampling flow rate: ≥1L/min
  • Positive pressure gas: (0.3~0.8)MPa purified instrument air or industrial N2
Working Conditions:
  • Temperature: (-20°C~+50)°C
  • Storage temperature: (-40~+80)°C
Features:
  • Explosion-proof technology for safety
  • Flexible installation options (wall-mounted and rack-mounted)
  • High anti-interference capability and sensitivity
  • Compact size with IP66 protection and temperature compensation
Frequently Measured Gas Types and Indexes:
  • O2: 0.01% Vol, Range: 0~5%
  • CO: 10 ppm, Range: 0~5000 ppm
  • CO2: 10 ppm, Range: 0~5000 ppm
  • H2O: 0.3 ppm, Range: 0~500 ppm
  • CH4: 10 ppm, Range: 0~10%

Note: Detection limits are specified for a 1m optical path at 20°C, 1 bar abs. Dual Gas CO+CO2 is available for specific applications.

See more

Catalog excerpts

LGA-3500-1

Specification Linear error Range drifting Repetitiveness error Technical Indexes Anti-explosion class (wall-mounted) Enclosure rating (wall-mounted) Response Time Sample gas flow: 1L/min, long optical length: T90<21s, short optical length: T90<11s 2 channels of (4~20)mA current with Analog output Electrical I/O isolation and max load is 500Ω 3 channels relay of 24V DC/1A Digital communication Power consumption Electrical Features Relay output Electrical safety Fuse Sampling flow rate Working temperature Storage temperature Working Conditions Pressure of sample gas Gas Input Conditions For more information, visit our website at www.fpi-inc.com/en © Focused Photonics Inc. All right reserved. All trademarks are the property of Focused Photonics Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Focused Photonics (Hangzhou) Inc. No. 459, Qianmo Road, Binjiang District, Hangzhou, China Tel: +86 571 8501 2188

 Open the catalog to page 1
LGA-3500-2

3.Leading DLAS Technology Adopting the high reliable and low noisy diode-laser optical source and DLAS technology to eliminate The LGA-3500 series laser gas analyzer is a cost-effective laser product developed by Focused Photonics (Hangzhou) Inc. using Semiconductor Laser Absorption Spectroscopy (DLAS) technology. It performs bypass processing of process environments with high temperature, high pressure, and high corrosion process environments to realize online industrial process gases detection. It can be widely used in chemical, cement, environmental protection and other markets. the influence...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.