Product catalog summary
Overview: The AQMS-600 Nitrogen Oxides Analyzer is designed for monitoring nitrogen oxides using chemiluminescence technology, recognized by the US EPA as a federal reference method. It measures various forms of nitrogen oxides in ambient air.
Key Features:
  • Utilizes chemiluminescence technology for accurate measurement.
  • Includes a permeation dryer and catalytic ozone scrubber for durability and safety.
  • Independent measurement ranges for NO, NO2, and NOX.
  • Complies with US EPA reference methods.
  • Offers multiple outputs including Ethernet and RS232.
  • User-friendly interface with a large screen and continuous system diagnostics with alarms.
  • Multi-tasking software for real-time monitoring of test variables.
  • Temperature and pressure compensation for measurement stability.
  • Internal data logging capabilities ranging from 1 minute to 365 days.
Measurement Principle: The analyzer measures nitrogen oxides indirectly by photometrically measuring the light intensity from the chemiluminescent reaction of nitric oxide (NO) with ozone (O3). NO2 is reduced to NO, and the total NOX concentration is determined.
Technical Specifications:
  • Standard range: 0~100 ppb to 0~20 ppm (selectable).
  • Zero noise: < 0.4 ppb (RMS).
  • Span noise: < 2.5 ppb.
  • Lower detectable limit: < 0.4 ppb.
  • Zero drift (24 hours): < 0.4 ppb.
  • Span drift: < 1% F.S.
  • Response time: < 40s (T90).
  • Precision: < 1%.
  • Linearity: < 1% F.S.
  • Sample flow rate: 500 cc/min ±10%.
  • Operating temperature: 20~30°C (per US EPA guidelines), safe operation from 5~40°C.
  • Power requirement: 100~240 VAC.
  • Dimensions and weight: 178mm(H) x 432mm(W) x 609mm(D), 22kg.
System Components: The system includes a pressure filter, NO2 converter, gas chamber, ozone purifier, pressure and flow controls, vacuum manifold, ozone scrubber, and sample filters.
Contact Information: Focused Photonics Inc. has offices in the US, China, India, Japan, and Singapore.
See more

Catalog excerpts

FPI AQMS-600 NOx, NO, NO2, nitrogen dioxide analyzer Ambient air quality monitoring system-1

AQMS-600 Nitrogen Oxides Analyzer FPI AQMS-600 nitrogen oxides (NO-NO2-NOX) analyzer utilizes chemiluminescence technology indicated by US EPA as federal reference method for monitoring on multiple forms of nitrogen oxides. v Permeation dryer on Ozone generator v Catalytic Ozone scrubber v Independent ranges for NO, NO and NO2 v Compliance with US EPA reference method v Various outputs include ethernet and RS232 v User friendly interface with large screen v Continuous system diagnosis with alarm v Multi-tasking software allows viewing test variables while operating v Temperature and pressure compensation v Internal data logging with 1 min to 365 day multiple averages Nitrogen oxides in ambient are measured indirectly by photometrically measuring the light intensity, resulting from the chemiluminescent reaction of nitric oxide (NO) with ozone (O3). NO is first quantitatively reduced to NO by means of a converter. NO, which commonly exists in ambient air together with NO, passes through the converter unchanged causing a resultant total NO concentration equal to NO + NO2. PMT The optical filter(660nm) attached on PMT provides reliable measurement, and temperature are controlled between 6~8OC with only 0.1OC fluctuation by a thermoelectric cooler, ensure the measurement stability. Ozone generation Permeation dryer are introduced in ozone generation to provide long system durability without any replacement. An catalytic ozone scrubber is standard for maximum safety and pump life before exhaust.

 Open the catalog to page 1
FPI AQMS-600 NOx, NO, NO2, nitrogen dioxide analyzer Ambient air quality monitoring system-2

Standard Range Zero Noise Span Noise Lower Detectable limit Zero Drift (24 hours) Span Drift Response Time Sample Flow Rate Operating Temperature Power Requirement 20~30°C range (per US EPA guidelines). Instrument may be safely operated over the range of 5~40°C 100~240 VAC, Converter applicable 178mm(H) x 432mm(W) x 609mm(D), 22kg System Drawing Exhaust —. Sample Filter Ozone Purifier

 Open the catalog to page 2
FPI AQMS-600 NOx, NO, NO2, nitrogen dioxide analyzer Ambient air quality monitoring system-3

© Focused Photonics (Hangzhou) Inc. All rights reserved. All trademarks are the property of Focused Photonics (Hangzhou) Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Focused Photonics Inc. US Focused Photonics Inc. China 3272 Countryside Drive 760 Binan Road, Hangzhou San Mateo, Ca. Zhejiang, China #317 Manjeera Majestic 403 Shimomaruya, 2-chome 7 Gul link, 629377 JNTU Road, Hyderabad, India kawaramachidori nijo sagaru, Singapore