The CEMS-2000 B FT is a Full Parameter Monitoring System utilizing FTIR (Fourier Transform Infrared) gas analysis technology. It is designed for direct thermal wet extraction and is particularly suitable for waste incineration applications. The system comprises three main components: a sampling pre-processing system, an FTIR gas analyzer, and a data acquisition and processing unit.
Features and Benefits
  • Simultaneous monitoring of multiple gas components such as SO2, NO, NO2, NOx, NH3, HCL, HF, O2, CO, CO2, CH4, and H2O.
  • Low minimum detection limits and high signal-to-noise ratio.
  • High-temperature measurement cell with a White chamber, capable of measuring up to a range of 5 meters.
  • Unique laminated filter with up to 180°C high-temperature heat tracing for a stable and reliable pre-treatment system.
  • Patented ROCKSOLID interferometer resistant to shear motion, frictional heat, and vibration interference, ensuring reliable operation in industrial environments.
Technical Specifications
  • Temperature Range: 18~35°C
  • Measurement Ranges: SO2 (0~75~2000 mg/m³), NO (0~100~2000 mg/m³), NO2 (0~50~1000 mg/m³), NOx (0~100~2000 mg/m³), NH3 (0~10~50 mg/m³), HCL (0~15~1500 mg/m³), HF (0~10~50 mg/m³), CO (0~75~1000 mg/m³), CO2 (0~20%), H2O (0~40%), O2 (0~25%)
  • Detection Limits: SO2 (0.45 mg/m³), NO (0.8 mg/m³), NO2 (0.4 mg/m³), NH3 (0.1 mg/m³), HCL (0.35 mg/m³), HF (0.25 mg/m³), CO (0.2 mg/m³), CO2 (0.025%), H2O (0.1%)
  • Linearity: ≤±2% Full Scale
  • Span Drift: ≤±2% Full Scale over 7 days
  • Zero Drift: ≤±2% Full Scale over 7 days
  • Response Time: SO2, NOx, CO, CO2 ≤180s (50m); HCL, HF, NH3 ≤300s
  • Power Supply: 220VAC/50Hz, 2KW (excluding sampling line and compressor)
  • Output Options: Analog, Ethernet connection, RS485, GPRS, System on-off, DP (optional)
Principle of Operation
FTIR spectroscopy involves using an interferometer to generate an interference pattern, which is then converted into infrared spectroscopy. This technique measures the energy changes between vibrational and rotational energy states caused by the absorption of infrared light by substances, allowing for both qualitative and quantitative analysis of samples.
See more

Catalog excerpts

CEMS-2000B FT Full Parameter Monitoring System-1

For more information, visit our website at www.fpi-inc.com/en © Focused Photonics Inc. All right reserved. All trademarks are the property of Focused Photonics Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Focused Photonics (Hangzhou) Inc. No. 459, Qianmo Road, Binjiang District, Hangzhou, China Tel: +86 571 8501 2188 CEMS-2000 B FT Full Parameter Monitoring System (FTIR)

 Open the catalog to page 1
CEMS-2000B FT Full Parameter Monitoring System-2

CEMS-2000 B FT system is a direct thermal wet extraction method based on FTIR gas analysis technology, the system mainly has three major components: sampling pre-processing system, FTIR gas analyzer, and data acquisition and processing unit. FTIR gas analysis technology uses the characteristic absorption of material molecules on the infrared spectrum to invert their concentrations, and through the single-channel spectrogram, the concentration value of each gas parameter is Principles Components calculated by an algorithm model. The system can simultaneously monitor multiple groups of components...