Fluke 120B Series Industrial ScopeMeter
11Pages

{{requestButtons}}

Catalog excerpts

Fluke 120B Series Industrial ScopeMeter - 1

TECHNICAL DATA Fluke 120B Series Industrial ScopeMeter® Hand-Held Oscilloscopes Simplified testing, more insight and faster electro-mechanical troubleshooting The compact ScopeMeter® 120B Series, is the rugged oscilloscope solution for industrial electrical and electro-mechanical equipment troubleshooting and maintenance applications. It’s a truly integrated test tool, with oscilloscope, multimeter and high-speed recorder in one easy-to-use instrument. The ScopeMeter 120B Series also integrates with Fluke Connect® mobile app and FlukeView® for ScopeMeter software to enable further collaboration, data analysis and archiving of critical test information. The 120B Series Industrial ScopeMeter Test Tools include innovative functions designed to help technicians troubleshoot faster and get the answers they need to keep their systems up and running. Display waveforms with Connect and View ™ trigger and setup technology and automatically view related numerical measurements using Fluke IntellaSet™ technology, all without making manual measurement adjustments. With Recorder Event Detect capabilities, elusive intermittent events are captured and logged for easy viewing and analysis. • • • • • KEY MEASUREMENTS Voltage, current and power waveforms with numerical values including harmonics, resistance, diode, continuity and capacitance measurements. AUTOMATICALLY CAPTURE, VIEW AND ANALYZE COMPLEX WAVEFORMS Fluke Connect and View™ triggering automatically displays waveforms without having to adjust amplitude, timebase and trigger settings, while Intellaset™ technology analyzes the signal and automatically displays critical numerical readings, making troubleshooting faster than ever. ® FLUKE CONNECT  COMPATIBLE* View data locally on the instrument, or via Fluke Connect mobile app. *Not all models are available in all countries. Check with your local Fluke representative. Dual-input digital oscilloscope and multimeter 40 MHz or 20 MHz oscilloscope bandwidth Two 5,000-count true-rms digital multimeters Connect-and-View ™ trigger simplicity for hands-off operation IntellaSet™ technology automatically and intelligently adjusts numerical readout based on the measured signal • Dual-input waveform and meter reading recorder for trending data over extended periods • Recorder Event Detect captures elusive intermittent signals on

Open the catalog to page 1
Fluke 120B Series Industrial ScopeMeter - 2

• Shielded test leads for oscilloscope, resistance and continuity measurements • Resistance, continuity, diode and capacitance meter measurements • Power measurements (W, VA, VAR, PF, DPF, Hz) • Voltage, current and power harmonics • Check Industrial networks with BusHealth physical layer tests against defined reference levels • Save or recall data and instrument setups • Store instrument setups defined by a test sequence for routine *Not all models are available in all countries. Check with your local Fluke representative. maintenance or most often used test procedures. • External...

Open the catalog to page 2
Fluke 120B Series Industrial ScopeMeter - 3

Industrial equipment needs a reliable power supply to operate properly, use the dual input to obtain key power measurements. For single phase or 3-phase balanced systems, the dual inputs of the Industrial ScopeMeter® 120B Series can measure ac+dc rms voltage on channel A and ac+dc rms current on channel B. The Fluke 125B can then calculate; frequency, phase angle, active power (kW), reactive power (VA or var), power factor (PF) or displacement power factor (DPF) and can also calculate the power values for a 3-phase system where all phases have equal voltage and currents. This applies to...

Open the catalog to page 3
Fluke 120B Series Industrial ScopeMeter - 4

Quickly step through recorded events to identify and troubleshoot intermittent faults. Fluke Connect mobile app compatibility Automated industrial machinery is harder than ever to trouble shoot. It’s not enough to just know where you have to test, you also have to know what to look for—and that can be hard without baseline measurement data or access to subject matter experts. The Fluke Connect® Assets wireless system of software and wireless test tools enables technicians to reduce maintenance costs and increase uptime with accurate equipment records and maintenance data that is easy to...

Open the catalog to page 4
Fluke 120B Series Industrial ScopeMeter - 5

Specifications Oscilloscope mode Vertical Frequency response - dc coupled with STL120-IV 1:1 shielded test leads without probes and test leads with STL120-IV 1:1 shielded test leads Frequency response - ac coupled (lf roll off) without probes and test leads (with BB120) Rise time, excluding probes, test leads Input impedance without probes and test leads with STL120-IV 1:1 shielded test leads Analog bandwidth limiter Display modes direct, with test leads, or with VP41 Probe 600 Vrms CAT IV, 750 Vrms maximum voltage. Max. floating voltage, from any terminal to ground Horizontal Scope modes...

Open the catalog to page 5
Fluke 120B Series Industrial ScopeMeter - 6

Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C. For voltage measurements with 10:1 probe, add probe uncertainty +1 %. More than one waveform period must be visible on the screen. Input A and input B DC voltage (VDC) Ranges Common mode rejection (CMRR) Full scale reading True-rms voltages (V ac and V ac+dc) Ranges Accuracy for 5 % to 100 % of range (DC coupled) Accuracy for 5 % to 100 % of range (AC or dc coupled) Common mode rejection (CMRR) Full scale reading 5000 counts, reading is independent of any signal crest factor. Peak Modes Max peak, Min peak, or pk-to-pk...

Open the catalog to page 6
Fluke 120B Series Industrial ScopeMeter - 7

Temperature (TEMP) with optional temperature probe Range Scale factor as V dc (add temp. probe uncertainty) Full scale reading Crest factor (CREST) Range Full scale reading Phase Modes Power (125B only) Configurations 1 phase / 3 phase 3 conductor balanced loads (3 phase: fundamental component only, AUTOSET mode only) Ratio between watts and VA range - 0.00 to 1.00 RMS reading of multiplying corresponding samples of input A (volts) and input B (amperes) Full scale reading Vrms x Arms Full scale reading Full scale reading Vpwm Purpose to measure on pulse width modulated signals, like motor...

Open the catalog to page 7