Catalog excerpts
Recently, we introduced Metrology Wells, which provide the very best performance in portable temperature calibration. The new 914X Series Field Metrology Wells extend high performance to the industrial process environment by maximizing portability, speed, and functionality with little compromise to metrology performance. Field Metrology Wells are packed with functionality and are remarkably easy to use. They are lightweight, small, and quick to reach temperature set points, yet they are stable, uniform, and accurate. These industrial temperature loop calibrators are perfect for performing transmitter loop calibrations, comparison calibrations, or simple checks of thermocouple sensors. With the “process” option, there is no need to carry additional tools into the field. This optional built-in two-channel readout reads resistance, voltage, and 4–20 mA current with 24 volt loop power. It also has on-board documentation. Combined, the three models (9142, 9143, and 9144—each with a “process” option) cover the wide range of –25 °C to 660 °C. High performance for the industrial environment Field Metrology Wells are designed for the industrial process environment. They weigh less than 8.2 kg (18 lb) and have a small footprint, which makes them easy to transport. Optimized for speed, Field Metrology Wells cool to –25 °C in 15 minutes and heat to 660 °C in 15 minutes. Field environment conditions are typically unstable, having wide temperature variations. Each Field Metrology Well has a built-in gradient-temperature compensation (patent pending) that adjusts control characteristics to ensure stable performance in unstable environments. In fact, all specifications are guaranteed over the environmental range of 13 °C to 33 °C. Field Metrology Wells Technical Data Lightweight, portable, and fastœœ Cool to –25 °C in 15 minutes and œœheat to 660 °C in 15 minutes Built-in two-channel readout for œœPRT, RTD, thermocouple, 4-20 mA current True reference thermometry with œœaccuracy to ± 0.01 °C On-board automation and œœdocumentation Metrology performance in accuracy, œœstability, uniformity, and loading
Open the catalog to page 1Built-in features to address large workloads and common applications Whether you need to calibrate 4-20 mA transmitters or a simple thermostatic switch, a Field Metrology Well is the right tool for the job. With three models covering the range of –25 °C to 660 °C, this family of Metrology Wells calibrates a wide range of sensor types. The optional process version (models 914X-X-P) provides a built-in two-channel thermometer readout that measures PRTs, RTDs, thermocouples, and 4-20 mA transmitters which includes the 24 V loop supply to power the transmitter. Each process version accepts an...
Open the catalog to page 2turnkey solution that will automate and document the results? The process versions of Field Metrology Wells have on-board non-volatile memory for documentation of up to 20 tests. Each test can be given a unique alphanumeric ID and will record block temperature, reference temperature, UUT values, error, date, and time. Each test can be easily viewed via the front panel or exported using Model 9930 Interface-it software, which is included with each shipment. Interface-it allows you to pull the raw data into a calibration report or an ASCII file. Operation is as easy as 1-2-3 You’ll find Field...
Open the catalog to page 3-P Specifications Built-in Reference Thermometer Readout Accuracy (4-Wire Reference Probe)† ± 0.013 °C at -25 °C ± 0.015 °C at 0 °C ± 0.020 °C at 50 °C ± 0.025 °C at 150 °C ± 0.030 °C at 200 °C ± 0.040 °C at 350 °C ± 0.050 °C at 420 °C ± 0.070 °C at 660 °C Reference Resistance Range 0 ohms to 400 ohms Reference Resistance Accuracy‡ 0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading Reference Characterizations ITS-90, CVD, IEC-751, Resistance Reference Measurement Capability 4-wire Reference Probe Connection 6-pin Din with Infocon Technology Built-in RTD Thermometer...
Open the catalog to page 4Ordering Information Ordering Information for 9142 9142-X Field Metrology Well, –25 °C to 150 °C, w/9142-INSX 9142-X-P Field Metrology Well, –25 °C to 150 °C, w/9142-INSX, w/Process Electronics X in the above model numbers to be replaced with A, B, C, D, E, or F as appropriate for the desired insert. See the inserts illustration and listing below. 9142-INSA Insert “A” 9142, Imperial Misc Holes 9142-INSB Insert “B” 9142, Imperial Comparison Holes 9142-INSC Insert “C” 9142, 0.25-inch Holes 9142-INSD Insert “D” 9142, Metric Comparison Holes 9142-INSE Insert “E” 9142, Metric Misc Holes...
Open the catalog to page 5Fluke. Keeping your world up and running.® Fluke Corporation Hart Scientific Division 799 E Utah Valley Drive American Fork, UT 84003 Tel: 801.763.1600 Fax: 801.763.1010 E-Mail: info@hartscientific.com www.hartscientific.com Fluke Eurpoe B. V., Hart Scientific Division PO Box 1186, 5602 BD Eindhoven The Netherlands Tel: +31 (0)40 2675 403 Fax: +31 (0)40 2675 404 E-mail: Hart.Logistics@Fluke.NL All other countries: Tel: +1 801.763.1600 Fax: +1 801.763.1010 ©2007 Fluke Corporation. All rights reserved. Specifications subject to change without notice. 7/2007 3082057 D-EN-Rev C Pub_ID:...
Open the catalog to page 6All Fluke Calibration catalogs and technical brochures
-
9500C
9 Pages
-
6135A/PMUCAL
8 Pages
-
P5514B Hydraulic
3 Pages
-
Temperature Calibration
40 Pages
-
molbox1+™ Flow Terminal
6 Pages
-
Fluke Precison Measurement
24 Pages
-
Triple Point of Water Cells
4 Pages
-
Micro-Baths
2 Pages
-
Really Cold Baths
2 Pages
-
The “DewK” Thermo-Hygrometer
4 Pages