Technical Data 452R Radiation Survey Meter Built to Withstand and Designed to Perform The Fluke Biomedical 452R combines the trusted precision of the original 452 with a rugged new design that's built to meet the toughest industrial demands. The 1 meter drop-tested 452R, with its enhanced durability and upgraded IP65 rating, is ready to tackle a broad range of applications, from non-destructive testing and contamination measurements to x-ray scanning and process control. Whether you're helping to ensure workplace safety, conducting critical inspections, or monitoring radiation, the Fluke Biomedical 452R delivers reliable, fast, and accurate results—no matter where the job takes you. One device for every situation. Endless possibilities. Technology The measurement technology of the Fluke Biomedical 452R is based on a combination of a silicon sensor cluster and a Geiger-Muller pancake. The instrument has two interchangeable lids (depending on model) to switch between air kerma, ambient dose equivalent and counts. This design makes it a versatile instrument with a wide and flat energy response along with high sensitivity and quick response time. Models The 452R comes in three different models.
Open the catalog to page 1Typical applications • X-ray tube leakage • X-ray wall leakage • Contamination measurements • Environmental radiation • Non-destructive testing • Scattered room radiation FLUK Biomedical • Broad application range • IP65 (dust tight and watertight against low pressure water spray) • Shock resistance: Designed to typically withstand a 1-meter drop onto concrete, as per the IEC 60068-2-31:2008 standard, with the lid mounted • Automatic data storage • Measures alpha, beta, gamma, X-ray radiation types • Alarm threshold setting • Built for indoor and outdoor applications 2 Fluke Biomedical 452R Radiation...
Open the catalog to page 21. The instrument uses a Geiger-Muller pancake at low rates and a cluster of solid-state sensors at higher rates. The rate where the solid-state sensors engage gradually increase with temperature, for temperatures above 30 °C (86 °F). 2. Limit where the response is within ±20 % of the response at continuous radiation. Above 30 °C (86 °F) the instrument's ability to handle low linac pulse rates and short X-ray pulses gradually declines with increasing temperature. 3. Refers to the microwave pulse repetition frequency of typical medical linear accelerators. Each pulse has a typical duration of...
Open the catalog to page 31. Above 30 °C (86 °F) the minimum dose rate gradually increases with increasing temperature. 2. Measured at 3 mm distance between instrument housing (without lid) and wide area class 2 sources according to ISO 8769:2010. 4 Fluke Biomedical 452R Radiation Survey Meter
Open the catalog to page 4Solid-state sensors (> 20 pSv/h (2 mrem/h), T< 30 ”C Geiger-Muller pancake (< 6 pSv/h (0.6 mrem/h)) Linoc. Nucleor power Industrial X-ray 5 Fluke Biomedical 452R Radiation Survey Meter
Open the catalog to page 5Ordering Information The system includes Instrument with mounted lids (depending on model). Power supply + plugs, 5 m USB cable, Printed user manual and quick guide, calibration certificate, cardboard box with fitted foam. Optional accessories Heavy duty case with fitted foam Protect+ Service Program The Protect+ Service Program helps ensure a predictable, annual expense to keep your instrument performing and looking new. This optional service program will keep your Fluke Biomedical 452R Survey Meter working accurately and efficiently through annual checks and calibrations and extends the instrument...
Open the catalog to page 6