1. Catalogs
  2. FISO Technologies
  3. VELOCE 100, Fiber Optic Temperature Monitor

VELOCE 100, Fiber Optic Temperature Monitor

VELOCE 100, Fiber Optic Temperature Monitor

VELOCE 100, Fiber Optic Temperature Monitor

Product catalog summary
Overview: The VELOCE 100 Fiber Optic Temperature Monitor is a multi-channel, upgradeable fiber optic signal conditioner designed for electromagnetic interference (EMI) characterization on electronic explosive devices (EEDs). It offers simultaneous readings from multiple measuring points with fast sampling rates, specifically catering to bridge wire type EEDs.
Design and Functionality: The monitor is used with the FOT-HERO fiber optic temperature sensor, featuring a miniature probe for precise positioning and minimal heat sink effect. It provides temperature resolution down to 0.025°C at a 200 KHz sampling rate. The system's patented technology ensures accurate and reliable measurements with a linear response, allowing calibration with minimal reference points.
Physical Specifications: The VELOCE 100 is housed in a 19-inch industrial rack chassis, accommodating up to eight optical modules. Each module offers BNC ±5 volts analog output and can be independently controlled and upgraded via RS-232 or FISOCommander BUS/VELOCE Edition software.
Key Features:
  • Temperature resolution: 0.025°C
  • 1 to 8 simultaneous channels
  • 200 KHz sampling rate
  • Voltage output
  • Upgradeable channels
Applications: The monitor is suitable for EED applications, HERO testing, point measurement applications, airbag igniters, and EMI compatibility testing of sensitive parts.
Technical Specifications:
  • Number of channels: 1 to 8
  • Sampling rate: 200 KHz
  • Averaging: 0.005 to 100 ms
  • Precision: 0.3°C
  • Resolution: 0.03°C
  • Dynamic range: 150°C (302°F)
  • Analog outputs: ±5 V with BNC connector
  • Communication: RS-232
  • Upgradeability: Flash ROM firmware
  • Weight: Chassis 4.8 kg, Module 0.5 kg
  • Dimensions: 448.8 × 326 × 132.5 mm
  • Power requirements: Chassis 29 W, Module 4 W
  • Operating temperature: 15°C to 35°C
Contact Information: FISO Technologies Inc., 500 St-Jean-Baptiste Ave., suite 195, Québec, QC, Canada G2E 5R9. Phone: (418) 688-8065, Fax: (418) 688-8067, Email: [email protected], Web: www.fiso.com
See more

Catalog excerpts

VELOCE 100, Fiber Optic Temperature Monitor-1

VELOCE 100 Fiber Optic Temperature Monitor Temperature resolution down to 0.025°C Voltage output Upgradeable number of channels The VELOCE 100 Fiber Optic Temperature Monitor is an upgradeable, multichannel, ber optic signal conditioner especially designed for electromagnetic interference (EMI) characterization on electronic explosive devices (EEDs). It provides simultaneous reading of a large number of measuring points with fast sampling rates. The VELOCE 100 is specically designed to meet the requirements needed for characterizing the EMI susceptibility of most bridge wire type EEDs. The VELOCE 100 Fiber Optic Temperature Monitor is designed to be used in conjunction with the FOT-HERO ber optic temperature sensor. The miniature probe size facilitates the positioning of the probe over the center position of the bridge wire while its low thermal mass minimizes the heat sink effect, allowing a temperature resolution down to 0.025°C at a sampling rate of 200 KHz. Thanks to its unique patented technology, the VELOCE 100 Fiber Optic Temperature Monitor is a relative measurement instrument that provides highly accurate and reliable measurements. The system’s linear response is so reliable that it allows sensor calibration with only one or two known current reference points. The VELOCE 100 Fiber Optic Temperature Monitor comes in a 19-inch industrial rack chassis that incorporates guide rails, connectors and a power supply. The chassis can accommodate from one to eight sliding independent optical modules. All optical input channels are easily accessible on the modules’ front panels. The system scans all the channels simultaneously at a 200 KHz sampling rate. Each module has its own BNC ±5 volts analog output. Temperature changes, expressed in volts accordingly to user’s own scale factor entry, are then collected by optional external analog acquisition systems. Each module can be independently controlled, upgraded, and optimized through its own RS-232 communication port or, even more easily, thanks to its accompanying software, FISOCommander BUS/ VELOCE Edition. EED applications HERO Testing Point measurement applications Airbag igniters EMI compatibility of sensitive parts

 Open the catalog to page 1
VELOCE 100, Fiber Optic Temperature Monitor-2

Specications Number of channels Sampling rate Dynamic range Analog outputs Weight Chassis Module Power requirements Chassis Module Operating temperature 500 St-Jean-Baptiste Ave., suite 195 Québec, QC Canada G2E 5R9 Email [email protected] Web www.so.com FISO Technologies Inc. reserves the right to make any changes in the specications of their products without prior notice. © 2006 FISO Technologies Inc. Imprimé au Canada / Printed in Canada DOC: MC-00012R9

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.