Pernicka 700H

Pernicka 700H

Pernicka 700H

Product catalog summary
Overview: The Cumulative Helium Leak Detector (CHLD) Pernicka 700H is a sophisticated leak testing system designed for sealed packages. It integrates mass spectrometer technology with cryogenic ultra-high vacuum capabilities, offering superior hermetic testing compared to traditional GROSS and FINE leak methods. It is applicable to any hermetically sealed device containing gases like Nitrogen, Helium, Argon, Krypton, Xenon, or those that can be 'bombed' with Helium.
Key Features:
  • High sensitivity with the smallest detectable leak rates of 4 x 10-14 atm-cc/s.
  • Combines Fine and Gross tests into a single operation for fast and effective testing.
  • Simultaneous detection of various gases including FCs, Nitrogen, Argon, and Xenon.
  • Environmentally friendly with no toxic or hazardous materials required.
  • Integrated LCD monitor for simple operation and an onboard computer for real-time data reading and recording.
  • Industrial, user-friendly design.
  • Complies with leak test methods according to MIL-STD-750, method 1071, procedure CH1-CH2, and MIL-STD-883H, method 1014.
Specifications:
  • Minimum detectable leak rate for Helium in FINE mode: >4E-14 mbar•L/s.
  • Maximum detectable leak rate for Helium in GROSS mode: >E-4 mbar•L/s.
  • Detectable masses range from 2-200.
  • Calibrated built-in test leak: E-10 mbar•L/s.
  • Test port: DN 16 CF.
  • Vacuum pump system includes a turbomolecular pump, oil-sealed roughing pump, and cryo pump.
  • Supply voltage: 110/120 V (ac) at 50/60Hz, 15A for the unit; 208-240 V (ac) at 50/60 Hz, 10A for the cryo compressor.
  • Operational temperature range: 15-28°C (60-80°F).
  • Weight: 245 kg (540 lb).
  • Dimensions: 660x1390x870 mm (26x54.5x34.25 in).
Accessories:
  • High Purity Gas Regulator customized 551-700S.
  • Large Double O-ring Chamber 551-710.
  • Medium Double O-ring Chamber 551-711.
  • Small Double O-ring Chamber 551-712.
  • Small Metal Seal Chamber 551-715.
Additional Information: The Pernicka 700H is designed for indoor use only and is part of a continuing program of product improvements, with specifications subject to change without notice. RateWatcher is a trademark of INFICON, and all other trademarks are the property of their respective owners.
See more

Catalog excerpts

Pernicka 700H-1

Cumulative Helium Leak Detector Pernicka 700H Cumulative Helium Leak Detector (CHLD), a single leak testing system for sealed packages, combines mass spectrometer expertise with cryogenic ultra-high vacuum. The Pernicka 700H offers hermetic testing superior to conventional GROSS Leak and FINE leak methods. This technique can be applied to any hermetically sealed device which contains a gas such as Nitrogen, Helium, Argon, Krypton, Xenon, etc. or can be "bombed" with Helium. The Pernicka 700H is controlled by an onboard computer that makes it easy to analyze sealed objects and archive test data simultaneously. The database provides a historical record and facilitiates the tracking of production yields. BENEFITS ■ High sensitivity for smallest detectable leak rates of 4 x 10E-14 atm-cc/s ■ Combining Fine and Gross tests into a single operation results in fast and effective testing procedures ■ Simultaneous detection of FCs, Nitrogen, Argon, Xenon, etc. ■ Environmentally friendly - no toxic or hazardous materials required ■ Integrated LCD monitor provides simple operation ■ Onboard computer provides real-time data reading and recording ■ Industrial, user-friendly design ■ Suitable for leak test methods according to MIL-STD-750, method 1071, procedure CH1-CH2, MIL-STD-883H, method 1014 Cumulative Helium Leak Detector Pernicka 700H jIllMFlCON

 Open the catalog to page 1
Pernicka 700H-2

Pernicka 700H CHLD Cumulative Helium Leak Type Detector Min. detectable leak rate for mbar^L/s Helium (FINE mode) Max. detectable leak rate for mbar^L/s Helium (GROSS mode) Detectable masses Calibrated built-in test leak mbar^L/s Test port Vacuum pump system Pernicka 700H CHLD Cumulative Helium Leak Detector >4E-14 Supply voltage unit V (ac) Supply voltage cryo compressor V (ac) 2-200 E-10 DN 16 CF turbomolecular pump, oil sealed roughing pump, cryo pump 110/120 ( 50/60Hz) 15A 208-240 (50/60 Hz)10A (air cooled ) Ambient conditions Operational temperature °C Cumulative Helium Leak Detector Pernicka...

 Open the catalog to page 2
Pernicka 700H-3

Pernicka 700H CHLD Cumulative Helium Leak Detector High Purity Gas Regulator customized 551-700S Large Double O-ring Chamber 551-710 Medium Double O-ring Chamber 551-711 Small Double O-ring Chamber 551-712 Small Metal Seal Chamber 551-715 www.inficon.com [email protected] Due to our continuing program of product improvements, specifications are subject to change without notice. RateWatcher is a trademark of INFICON. All other trademarks are the property of their respective owners. (2013-05)

 Open the catalog to page 3
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.