© Guarantees gas analysis without any water residue © Robust industrial manufacturing © Low noise level © Quick installation © Stainless steel cooling circuit © Up to 3 circuits managed per device ■ Hermetic piston compressor ■ Electronic thermostat control ■ Automatic expansion valve ■ Low-carbon 316L stainless steel cooling circuit ■ Filtered, anti-rejection power supply for protecting the electronics inside the analyzers ■ Electro-galvanized sheet metal casing ■ Main models available: SGA400-1, SGA400-2, SGA400-3 ■ 18-month warranty (unless otherwise agreed)
Open the catalog to page 1(CAPTION) ICompressor 2Air condenser 3Fan 4Dryer 5Variable speed drive 6Expansion valve 7Water separator 8Hot gas solenoid valve 9Gas inlet lKBas outlet ICondensate drain lZemperature controller 1PT100 probe
Open the catalog to page 3EURODIFROID gas dryers ensure the absence of water in analyzers. They operate on the principle of “direct expansion”: the gas to be cooled enters a heat exchanger-separator cooled by the evaporating refrigerant. The water (or condensate) is then trapped in the lower part of the separator and can be discharged by gravity or a peristaltic pump. The gas outlet temperature is regulated and maintained constant by a hot gas injection valve, ensuring a dew point of +1°C. Each dryer can be configured with 1, 2, or 3 independent circuits. Delivered ready for use, they are charged with CFC-free gas. The...