1. Catalogs
  2. ESPEC
  3. Walk-in Type Temperature (& Humidity) Chambers E series

Walk-in Type Temperature (& Humidity) Chambers E series

Walk-in Type Temperature (& Humidity) Chambers E series
1 / 20 PagesView full catalog

Walk-in Type Temperature (& Humidity) Chambers E series

Product catalog summary
Overview: The document provides a comprehensive overview of the E Series Walk-In Type Temperature (& Humidity) Chamber by ESPEC CORP, emphasizing its energy efficiency, advanced control systems, and networking capabilities.
Specifications: The E Series chambers offer temperature ranges from -40°C to +80°C and humidity levels from 10% to 95% RH. They are available in sizes from 4.2m³ to 52.7m³ and feature both air-cooled and water-cooled systems.
Energy Efficiency: These chambers are designed to significantly reduce energy consumption, utilizing a wide-range refrigeration control system and patented cross-output control, achieving up to 50% energy savings. The Eco-Mode further enhances efficiency.
Control and Networking: Equipped with N-instrumentation, the chambers provide a user-friendly interface and precise control. They support remote monitoring and operation via Ethernet and USB, allowing for program pattern editing, data management, and email alerts. A backtrace function aids troubleshooting.
Customization and Options: The chambers can be customized with various mechanical compartment variations and options to meet specific testing needs, including different power capacities to match the test area and specimen heat load.
Additional Features: Features include a high-speed processing N-instrumentation with a 10.4-inch color touch panel, multi-lingual display options, and a variety of program settings. The system supports constant and program operation modes, with alarm functions, calendar timers, and sampling functions.
Operating Conditions: Temperature and humidity control ranges are specified for various conditions, with energy savings rates up to 60% under certain conditions.
Power Consumption and Energy Savings: Detailed power consumption comparisons show significant energy savings in newer models, such as a 60% reduction in annual power usage for the EBL-4 model.
Characteristics and Features: The chamber includes features like automatic fog prevention, an automatic backup function, and a boiler humidification system. Custom-made chambers can be tailored to specific testing requirements.
Safety Devices: Safety features include overcurrent protection, short-circuit protection, and temperature and humidity limit alarms.
Options and Accessories: Additional features include enlarged viewing windows, double swing doors, entrance curtains, and preparation rooms. Options for remote control and data recording are also available.
Specifications and Features: Key features include a full-ceiling air duct, insertion ramp, frost-free expansion, airflow adjuster, low humidity equipment, refrigerator for heat load, auxiliary humidifier, air-cooled inverter chiller system, water purifier, flow switch, and exhaust air duct.
Safety Precautions: Users are advised to avoid using explosive or inflammable specimens, not to place corrosive materials in the chamber, and to read the operation manual before use.
Certifications: ESPEC CORP. is ISO 9001:2008 certified for its Quality Management System.
Contact Information: The document includes contact details for ESPEC's head office and various international branches.
See more

Catalog excerpts

Walk-in Type Temperature (& Humidity) Chambers E series-1

Walk-In Type Temperature (& Humidity) Chamber

 Open the catalog to page 1
Walk-in Type Temperature (& Humidity) Chambers E series-2

New N-instrumentation produces a smart chamber Leading eco model better than ever The E Series walk-in type chamber is already renowned for its epoch-making power savings. Power savings achieved thanks to a wide-range refrigeration control system, an active map system, and other original control devices. Energy reduction accounts in some cases for half of the previous model consumption, due to the development of a cross output control system (patented) lowering electrical requirements of temperature & humidity operation and many other features. Energy consumption is, more than ever, an important...

 Open the catalog to page 2
Walk-in Type Temperature (& Humidity) Chambers E series-3

* Equipped with options.

 Open the catalog to page 3
Walk-in Type Temperature (& Humidity) Chambers E series-4

SERIES Model Temperature & Humidity Chambers Temperature & Humidity Chamber EBE Inside capacity (Floor Area) Temperature Chambers Temperature Chamber EBUU Temperature & Humidity Chamber EBR Temperature & Humidity Chamber EBL Refrigerator System Air-cooled Water-cooled ● System configuration Test chamber variations Mechanical compartment variations Program control operation 2.2kW Custom configuration

 Open the catalog to page 4
Walk-in Type Temperature (& Humidity) Chambers E series-5

Characteristics Networking communications and USB data transfer for optimum test operations Networking with new N-instrumentation Supports chamber control from a variety of terminals T h e t a b b e d u s e r i n t e r f a c e of t h e chamber instrumentation is designed to m a ke o p e r at io n s fo r a c c e s si ng t o specific screens easier than ever. An Ethernet (LAN) port is equipped as standard, so it is possible to connect to and monitor the chamber operation conditions from a device such as a PC or t ablet PC. A s a n opt ion , a Web browser can be used to communicate with chambers...

 Open the catalog to page 5
Walk-in Type Temperature (& Humidity) Chambers E series-6

Characteristics ● Instrumentation panel Program pattern edit and copy A prog ram pat ter n created on a computer can be written to a chamber, a nd p rog r a m p at t e r n s st or e d on a chamber can be copied to a computer via the USB interface. Downloading test programs ESPEC's web-based Tech site "Test Navi" introduces various test standards related to environmental testing. Test standards can be perfor med by dow n loa d i ng a test prog r a m to t he computer connected to the chamber. A program pattern editing function enables modification of test profiles. In addition, accurate testing...

 Open the catalog to page 6
Walk-in Type Temperature (& Humidity) Chambers E series-7

When the chamber stops because of trouble, the operation state just before the chamber stops is automatically recorded and saved. Saved data can be Attach this data file to an email to ESPEC, and we will perform troubleshooting. i Online diagnostics services Diagnostics service is available using the backtrace data from the time of Send the backtrace data to ESPEC via email; we will analyze the cause of the trouble and report the diagnosis back to This service ensures accurately- performed diagnosis so that, in the case that repair work is required, appropriate troubleshooting will be prescribed...

 Open the catalog to page 7
Walk-in Type Temperature (& Humidity) Chambers E series-8

Characteristics High-speed processing N-instrumentation and 10.4 inch color touch panel feature improved operability and legibility Tabbed user interface Controller’s new layout includes tabs at t he bot tom of t he sc reen to ea sily activate any section. Calculating and processing performances have been improved, and the screen layout optimized. A variety of program settings The program memory has a capacity for 40 patterns (1 pattern 99 steps). The time for each step is set in 1-minute increments up to a total of 9999 hours and 59 minutes. Steps can be inserted, copied, and deleted. Created...

 Open the catalog to page 8
Walk-in Type Temperature (& Humidity) Chambers E series-9

Characteristics N-Instrumentation ● Constant operation registration Operating mode Constant operation, program operation, stop Temperature & humidity control function Control system: PID control (sample temperature control support), energy-saving refrigeration capacity automatic control Temperature input: Test chamber temperature (Thermocouple type T) Humidity input: Test chamber wet-and-drybulb temperature input system (Thermocouple type T) Input functions: 100 ms high-speed sampling, wire break detection Adjustment function: Temperature offset setting Setting resolution • Constant setup Setting...

 Open the catalog to page 9
Walk-in Type Temperature (& Humidity) Chambers E series-10

Characteristics Allowable heat generation load comparison Model Previous series An automatic control system to quickly respond to load changes (Patented) Simply by setting the temperature and humidity conditions, the automatic cont rol system works at maximu m capacity up to the setting, and then after reaching it, can maintain it at minimum capacity (Japanese patent No. 2933269). It can quick ly respond to door operat ions a nd cha nges i n heatgeneration loads during testing to create a constantly stable testing environment. Improving temperature & humidity distribution performance/allowable...

 Open the catalog to page 10
Walk-in Type Temperature (& Humidity) Chambers E series-11

Characteristics Saving energy –major reduction in power consumption and power equipment capacity Featuring a new high precision cooling system with greatly reduced power consumption (Patent pending) We h ave d e velo p e d a w id e - r a n ge cooling control system composed of a D C i n ve r t e r t h a t c a n w i d e l y a n d continuously vary refrigeration capacity an electronic expansion valve, as well as unique control methods such as the a c t ive m a p s y s t e m t o c o nt r ol t h e operation of multiple refrigerators at minimum power consumption. This cooling system controls out put...

 Open the catalog to page 11
Walk-in Type Temperature (& Humidity) Chambers E series-12

Characteristics The viewing window automatically prevents fogging A heater in the viewing window door section automatically turns on according to the temperature setting and glass incorporating a heat generator is used to p r eve nt fog g i ng. A la rge v iew i ng window option is also provided. Continue operation with the automatic backup function when trouble occurs Viewing window (W180 mm x H289 mm ) Option: Large viewing window (W440 mm x H295 mm) If a problem occurs with a portion of the chamber, the remaining devices can be f lexibly r un to continue operation, making it unnecessary to...

 Open the catalog to page 12
Walk-in Type Temperature (& Humidity) Chambers E series-13

Characteristics New features –Enhanced testing accuracy All-weather LED lights (Patent pending) These LED lights now illuminates in the entire temperature and humidity control range. Allying energy savings, long life, and excellent responsiveness when turned on, LED lights bring brightness (approx. 2.5x), and power consumption is reduced of approx. 1/5th compared to conventional incandescent lights illuminating the chamber. Reduced odors/siloxanes A new seal material used in the panel con nections has been developed and generates no low-molecular siloxanes. This not only prevents any effect on...

 Open the catalog to page 13
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.