Thermal Shock Chamber (TSD-100)
16Pages

{{requestButtons}}

Catalog excerpts

Thermal Shock Chamber (TSD-100) - 1

Quality is more than a word

Open the catalog to page 1
Thermal Shock Chamber (TSD-100) - 2

Large-capacity two-zone chamber capable of exposing specimens to a uniform thermal stress. Two-zone thermal shock chamber ideally suited for Japanese and international test standards such as MIL, IEC or JASO . With its 100L test area, and its outstanding thermal uniformity characteristics, the TSD Thermal Shock Chamber from ESPEC can test specimens under a uniform thermal stress, and is ideally suitable for a wide range of applications, from research and development to inspection or production.

Open the catalog to page 2
Thermal Shock Chamber (TSD-100) - 3

*The viewing window, paperless recorder and additional overheat protector are optional.

Open the catalog to page 3
Thermal Shock Chamber (TSD-100) - 4

Performance Meeting the high performance required by today's test standards Specimen temperature recovery (example) (based on MIL-STD-883 condition C) Test conditions Temperature uniformity measurement method High temp. exposure: 155 for 30 min. Thermocouples were embedded in 10 ICs Low temp. exposure: 68 for 30 min. placed on two levels in each of the corners Specimens: ICs, 10 kg and in the center of a specimen basket. (Specimens with embedded thermocouples were placed beneath other ICs.) Temperature recovery time shortened with the two-zone system During testing of 10 kg of plastic...

Open the catalog to page 4
Thermal Shock Chamber (TSD-100) - 5

Utility Boast of a 100L capacity Featuring a 100L test area capacity, the chamber dramatically increases the processing capabilities and even allows to test A4-size printed circuit boards laid flat at once. Smooth specimen transfer The “Soft-move mode” is used to reduce vibration shocks when specimens are movi ng bet ween the high and low temperature chambers. Test area anti-drop mechanism to protect specimens A braking system fitted to the drive mechanism prevents specimens from falling into the test area when the chamber stops operation. Test areas top: high temperature chamber bottom:...

Open the catalog to page 5
Thermal Shock Chamber (TSD-100) - 6

Control operation Color LCD interactive touch-screen system Operation and settings simplified by the use of a touch-screen LCD display (instructions displayed on-screen). At-aglance confirmation of test patterns, test area temperatures, temperature cycles, upstream/ downstream control, and trend graph displays. Door-mounted instrumentation Instrumentation including the touchscreen controller is incorporated into the door. It reduces the overall footprint and frees up both sides of the chamber for easy access. Setting Instrumentation Interactive key input by touch panel Test area: exposure...

Open the catalog to page 6
Thermal Shock Chamber (TSD-100) - 7

NSTRUMENTATION PANEL •Test detail monitor Test details are displayed while the Product sensor settings Possibility to enable or disable the specimen temperature sensor Same procedure for additional optional specimen temperature •Test setting Displays the conditions to define Service guide If an inspection item is selected. maintenance steps appears. This is useful before proceeding with tests or to assist during periodic Detailed description Details the regular inspection ■f tl.cf niH±lld! Hf E.HL'lt HTLl

Open the catalog to page 7
Thermal Shock Chamber (TSD-100) - 8

Eco-friendly A whole range of environmentally-friendly features Reduced power consumption Reducing power consumption is an important issue for our customers. The TSD-100 employs number of measures, such as the use of electronic expansion valves to regulate the refrigeration capacity control, and thus reduce overall power consumption by 30% (compared to previous ESPEC models). Material labeling for easy recycling Built-in paperless recorder (option) *Rear side, on the right Plastic molded components are labeled and easily detachable to make recycling easier duning future disposal of the...

Open the catalog to page 8
Thermal Shock Chamber (TSD-100) - 11

TEST STANDARD (TSD-100 COMPATIBILITY) Temperature setting Test standard High temp. ( Specimen less than 15 min. 10 min. or longer after transition start 15g and below: 15 to 150g: 150 to 1,500g: More than 1,500g: Within 5 min. after solder joint temp. reaches ±2℃ of preset temp. Or, 0.2kg and below: 0.5 hours 0.2 to 0.8kg: 1 hour 0.8 to 1.5kg: 1.5 hours More than 1.5kg: 2 hours preset temp Depends on parties involved 0.2kg and below: 0.2 to 0.8kg: 0.8 to 1.5kg: More than 1.5kg: Depends on parties involved Max. storage temp. Min. storage temp. Air 5 min. or 10 of soak time, whichever is...

Open the catalog to page 11
Thermal Shock Chamber (TSD-100) - 12

SPECIFICATIONS Model 2-zone transition by vertical transfer of specimens High temp. exposure range Test area Low temp. exposure range Temp. uctuation Pre-heat upper limit Cold chamber Pre-cool lower limit Temp. pull down time *2 2-zone High temp. exposure: Low temp. exposure: Power supply voltage: Sensor position: Specimen: Recovery conditions 150 *3 30 min. 65 *3 30 min. rated voltage downstream Plastic molded ICs, 10kg Specimen IC temp. within 15 min. High temp. exposure: Ambient temp.: Power supply voltage: Specimen: Recovery conditions Ambient temp. recovery time 150 to max. 55 23 rated...

Open the catalog to page 12
Thermal Shock Chamber (TSD-100) - 13

SAFETY DEVICES • Leakage breaker (200, 220, 380V AC) • Specimen basket • Circuit breaker (208, 400/415V AC) (18-8 Cr-Ni stainless steel: 5 mesh metal basket) • Electrical compartment door switch W700 x H40 x 0410 mm/ load capacity 5kg 2 • Hot chamber overheat protection switch * Shelf brackets 2 sets • Cold chamber overheat protection switch * Cartridge fuse (1A, 7A, 10A, 15A) 4 • Hot chamber overheat protectors * Cable Port rubber Plu9 2 (Built into temperature controller) • Perforated cable port cap 1 • Cold chamber overheat/ overcool protectors • Wire fisher (specimen wiring tool) 1...

Open the catalog to page 13
Thermal Shock Chamber (TSD-100) - 14

OPTIONS Paperless recorder Records temperature of each section such as the temperature inside the chamber. Select either built-in or portable type. Built-in type Number of inputs (Initial setting): PL1S: 1 (5 more channels can be turned ON) Data saving cycle: 1 sec PL3S: 3 (3 more channels can be turned ON) Data saving cycle: 1 sec PL3L: 3 (3 more channels can be turned ON) Data saving cycle: 5 sec PL5S: 5 (1 more channels can be turned ON) Data saving cycle: 1 sec PL5L: 5 (1 more channels can be turned ON) Data saving cycle: 5 sec PL6S: 6 Data saving cycle: 1 sec PL6L: 6 Data saving cycle:...

Open the catalog to page 14
Thermal Shock Chamber (TSD-100) - 15

OPTIONS Additional cable port Additional overheat protector Provided in addition to the standard cable port. (right side) Location: Left side of the main unit Internal diameter: 100 mm Additional preventive measures can be taken for excessive temperature rise in the chamber, in addition to the standard equipped double overheat protector. RS-232C GPIB * Select one, instead of standard RS-485. Communication cables Cable port rubber plug External alarm terminal Prevents air leakage from the cable port. If the safety device of the chamber is activated, the external alarm terminal will notify :t...