1. Catalogs
  2. ERICHSEN GmbH & Co. KG
  3. Technical Description Model GalvanoTest 2000 / 3000

Technical Description Model GalvanoTest 2000 / 3000

Technical Description Model GalvanoTest 2000 / 3000

Technical Description Model GalvanoTest 2000 / 3000

Product catalog summary
Overview: The document provides a detailed description of the GalvanoTest, a device used for measuring the thickness of electroplated coatings on various substrates using the coulometric or anodic de-plating technique. This method is applicable to a wide range of materials including steel, non-ferrous metals, and insulating materials.
Application: The GalvanoTest is used for measuring coatings such as nickel on steel, zinc on steel, tin on copper, silver on copper, and copper on epoxy. It is particularly useful for multi-layer systems like chrome plus nickel plus copper on steel.
Measuring Principle: The device operates based on Faraday’s law, performing electrolytic dissolution of a precisely delimited area of the sample. The process involves placing a measuring cell on the test sample, which is then filled with a specially adapted electrolyte. The current through the electrolyte dissolves the coating, and the thickness is displayed digitally once the process is complete.
Features: The GalvanoTest is available in two models, Type 2000 and Type 3000, both supporting over 70 coating/substrate combinations. They offer features like adjustable deplating speeds, calibration settings, data storage, and statistical evaluation. The Type 3000 model includes additional conveniences such as an integrated circulator pump and the ability to measure very small areas.
Components: The GalvanoTest consists of three main components: a measuring cell, a stand for holding the test sample, and a processing unit with an interactive LCD. It can connect to peripheral devices like a PC or printer for data documentation.
Accessories and Options: Various accessories are available, including replacement gaskets, masks, electrolyte solutions, and data printers. The device supports multiple languages and measurement units.
Usage and Maintenance: The solutions used in the GalvanoTest do not harm the coating until current is applied, and the accurate operation of the gauge is ensured by using original solutions. The document also provides supply schedules for both GalvanoTest models, detailing included components and accessories.
See more

Catalog excerpts

Technical Description Model GalvanoTest 2000 / 3000-1

testing equipment for quality management Versatile Coating Thickness Gauge: • measures almost any electroplated coating/base combination • suitable for single and multi-layer coatings • utilizes the coulometric principle, a de-plating method conforming to DIN 50 955 and ISO 2177 Also suitable for extremely thin coatings from 0.05 microns!

 Open the catalog to page 1
Technical Description Model GalvanoTest 2000 / 3000-2

Coating thickness gauge GalvanoTest Application The coulometric or anodic de-plating technique is used for measuring the thickness of electroplated coatings on virtually all substrates such as steel, non-ferrous metals or insulating material bases. Typical applications include: nickel on steel, zinc on steel, tin on copper, silver on copper or copper on epoxy. metric method ensures reliable and exact results. Designed for complete ease of use, GalvanoTest requires no specific operator skills and only minimal training before actual measurements can take place. The coulometric principle is the...

 Open the catalog to page 2
Technical Description Model GalvanoTest 2000 / 3000-3

GalvanoTest Features Coating/ - more than 70 coating/substrate combinations (standard version) • • substrate combinations - coatings on flat and curved surfaces • • - coatings on small components and wires • • - 10 preset types of metal: Cr, Ni, Cu, brass, Zn, Ag, Sn, Pb, Cd, Au • - 9 preset types of metal: Cr, Ni, Cu, brass, Zn, Ag, Sn, Pb, Cd • - 8 further metal coatings for special applications • - 1 further metal coating for special application • Measuring cell - with circulation pump • Measuring surface - gasket 8 mm2/12.4 x 10'3 inch2 • - mask 0.25 mm2/0.388 x 10'3 inch2 (deplated area...

 Open the catalog to page 3
Technical Description Model GalvanoTest 2000 / 3000-4

The GalvanoTest gauge consists of three major components: 1. Measuring cell For measuring on even or curved surfaces up to 3 mm curvature radius, two different measuring cells are available. A measuring cell with circulating pump is connected to GalvanoTest 3000, a cell with air pulsator is connected to the 2000 model. The electrolyte is constantly moved around the measuring cell to ensure an even de-plating of the measuring area as well as on optimum use of the electrolyte. For measuring wires or small parts, an electrolyte cup is available to dip the samples into the The stand serves to hold...

 Open the catalog to page 4
Technical Description Model GalvanoTest 2000 / 3000-5

Two GalvanoTest models are available: The GalvanoTest 2000 is the basic model and measures a variety of coating/base combinations without any accessories being required. Designed for easy use, this model is mostly recommended for measuring multi-layer applications. With the GalvanoTest 3000 additional measuring conveniences are provided. Its measuring stand with integrated electrolyte pump and easy to change measuring cell for automatic electrolyte removal and refilling enable electrolyte savings as one filling may be used for a large series of de-plating processes. Additional measuring cells...

 Open the catalog to page 5
Technical Description Model GalvanoTest 2000 / 3000-6

The print-out of the characteristic measuring cell voltage curve graphically represents the measuring process with its typical voltage jump at the end of measurement. The voltage jump terminates measurement and thickness is automatically displayed in microns or mils. The size of de-plating area can be precisely delimited using the supplied gaskets or masks. The illustrations below represent the actual deplating area size along with a microscopic view of the measuring area delimited when using the D-mask (0,25 mm2). Microscopic view of measuring area created by the D-mask thickness gauges ■ Ultrasonic...