1. Catalogs
  2. ERICHSEN GmbH & Co. KG
  3. Technical Description Model 602

Technical Description Model 602

Technical Description Model 602

Technical Description Model 602

Product catalog summary
Technical Description
The Cathodic Delaminator Model 602 is designed for rapid corrosion testing and quality control of coated metals. It consists of a control unit and a temperature control vessel with eight test receptacles, each with its own constant current source, allowing independent testing of up to eight specimens.

Purpose and Application
This equipment is used to assess the adhesion of coatings on metals, particularly after mechanical damage such as stone impacts or transport damage. It evaluates the effectiveness of corrosion protection systems, focusing on substrate preparation and coating composition.

Principle of the Test
The test involves creating a defined scratch on the coating using a scratching tool, exposing the metal. The specimen is then immersed in an electrolyte solution (0.5 N KOH) and secured with an alligator clip. The test runs for 30 minutes, after which the infiltration of the coating is assessed. The test helps identify pretreatment errors and coating infiltration in damaged areas.

Technical Data
  • Dimensions: Immersion test container - approx. 700 x 350 x 352 mm; Supply unit - approx. 260 x 275 x 110 mm
  • Power Supply: 230 V AC, 50 Hz, with a residual current circuit breaker 30mA
  • Weight: Supply unit - approx. 3 kg; Immersion container with test receptacles - approx. 12.5 kg; with temperature control vessel - approx. 16 kg
  • Heating Period: Test receptacles - approx. 20 min; Immersion container - approx. 1.5 hours
  • Environment Temperature: 23°C; Water Temperature: 20°C

Order Information
Order No. 0311.01.31 for the Cathodic Delaminator Model 602, which includes the control unit, temperature control vessel, 8 test receptacles, circulator, mains cable, and manual. Technical modifications are reserved.
See more

Catalog excerpts

Technical Description Model 602-1

pre-treatment faults Cathodic Delaminator Model 602 testing equipment for quality management Technical Description

 Open the catalog to page 1
Technical Description Model 602-2

Mechanical injury of organic coatings on metal cause the loss of corrosion protection. Known examples of the mechanical destruction of coatings are: stone impact on motor vehicles, transport and installation damage. Corrosion will be affected by corrosion protection system. Particularly the preparation of the substrate surface before coating and the composition of the coating are important for corrosion protection systems. The Cathodic Delaminator, Model 602, is used for rapid checking of product quality and frequently quality control of coated metals. The apparatus consists of a control unit...

 Open the catalog to page 2

All ERICHSEN GmbH & Co. KG catalogs and technical brochures

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.